BACKGROUND: Emerging hypofractionated prostate radiotherapy regimens require solutions for accurate target tracking during beam delivery. The goal of this study is to evaluate the performance of the Clarity ultrasound monitoring system for prostate motion tracking.Entities:
Keywords: Image-guidance; Prostate cancer; Prostate motion; Radiotherapy; Tracking; Ultrasound
Mesh:
Year: 2018 PMID: 30126434 PMCID: PMC6102861 DOI: 10.1186/s13014-018-1097-8
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Fig. 1A representative example of a patient’s prostate displacements throughout a treatment session. Note that KV image pair acquisition also includes image registration and couch repositioning
Fig. 2a Experimental setup for phantom validation. b A sample MV image with indicated positions of the segmented fiducials (solid blue circles), predicted positions of the fiducials based on ultrasound tracking (hollow blue circles), and predicted positions of the fiducials assuming no motion (green asterisks). c Pixel coordinate (u) of fiducial 1 at different times during the MV arc beam delivery. d Pixel coordinate (v) of fiducial 1 at different times during the MV arc beam delivery. e Difference vectors (scaled at isocenter) between mean predicted and mean actual (MV segmented) positions of the 4 fiducials. f Relative cumulative distribution of the magnitudes of the difference vectors shown in (e)
Fig. 3RMS of the prostate displacements in each fraction for all five patients. Mean prostate displacements and standard deviations along IS (Inferior-Superior), LR (Left-Right) and AP (Anterior-Superior) directions are also given
Fig. 4a-e Relative cumulative distributions of the magnitudes of the difference vectors (scaled at isocenter) between mean predicted and mean actual (MV segmented) fiducial positions for individual patients. Predicted positions are calculated with (blue) and without (green) ultrasound tracking. f Same as in A-E with aggregate data for all patients
Summary of the magnitudes of the max and 95% position differences (scaled at isocenter) between predicted and actual (MV segmented) fiducial positions for individual patients
| Patient # | Magnitude of position differences @ iso (mm) | |||
|---|---|---|---|---|
| Maximum | At 95% relative cumulative occurrence | |||
| Without tracking | With tracking | Without tracking | With tracking | |
| 1 | 2.3 | 1.8 | 1.7 | 1.2 |
| 2 | 3.1 | 1.7 | 2.2 | 1.3 |
| 3 | 1.3 | 1.3 | 1.1 | 1.0 |
| 4 | 3.9 | 3.3 | 2.4 | 2.1 |
| 5 | 1.9 | 1.5 | 1.5 | 1.2 |
Position differences are calculated with and without ultrasound tracking