Simethicone is retained in endoscopes despite reprocessing: impact of its use on working channel fluid retention and adenosine triphosphate bioluminescence values (with video).

Monique T Barakat1, Robert J Huang1, Subhas Banerjee1.   

Abstract

BACKGROUND AND AIMS: Studies from our group and others demonstrate residual fluid in 42% to 95% of endoscope working channels despite high-level disinfection and drying. Additionally, persistent simethicone has been reported in endoscope channels despite reprocessing.
METHODS: Endoscopy was performed by using water or varied simethicone concentrations (0.5%, 1%, 3%) for flushing. After high-level disinfection/drying, we inspected endoscope working channels for retained fluid by using the SteriCam borescope. Working channel rinsates were evaluated for adenosine triphosphate (ATP) bioluminescence. Fourier transform infrared spectroscopy was performed on fluid droplets gathered from a colonoscope in which low-concentration simethicone was used.
RESULTS: Use of medium/high concentrations of simethicone resulted in a higher mean number of fluid droplets (13.5/17.3 droplets, respectively) and ATP bioluminescence values (20.6/23 relative light units [RLUs], respectively) compared with that of procedures using only water (6.3 droplets/10.9 RLUs; P < .001). Two automated endoscope reprocessing cycles resulted in return of a fluid droplet and ATP bioluminescence values to ranges similar to that of procedures that used only water (P = .56). Low-concentration simethicone did not increase the mean residual fluid or ATP bioluminescence values compared with procedures that used only water (5.8 droplets/15.6 RLUs). Fourier transform infrared analysis revealed simethicone in the endoscope working channel after use of low-concentration simethicone.
CONCLUSIONS: Use of medium/high concentrations of simethicone is associated with retention of increased fluid droplets and higher ATP bioluminescence values in endoscope working channels, compared with endoscopes in which water or low concentration simethicone was used. However, simethicone is detectable in endoscopes despite reprocessing, even when it is utilized in low concentrations. Our data suggest that when simethicone is used, it should be used in the lowest concentration possible. Facilities may consider 2 automated endoscope reprocessor cycles for reprocessing of endoscopes when simethicone has been used.
Copyright © 2019 American Society for Gastrointestinal Endoscopy. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 30125574      PMCID: PMC6754731          DOI: 10.1016/j.gie.2018.08.012

Source DB:  PubMed          Journal:  Gastrointest Endosc        ISSN: 0016-5107            Impact factor:   9.427


  5 in total

1.  An in vitro and clinical dose-finding study of antifoaming effects of simethicone during colonoscopy.

Authors:  Marguerite J Kutyla; Ayesha Shah; Jerome Elson; Vera Meeusen; Sam O'Connor; Luke F Hourigan; Gerald J Holtmann
Journal:  Indian J Gastroenterol       Date:  2019-07-17

2.  Efficacy of a novel channel-cleaning ball brush for endoscope reprocessing: a randomized controlled trial.

Authors:  Kwang Hyun Chung; Jeong Don Chae; Wonho Choe; Hyo Young Lee; Il Hwan Oh; Byoung Kwan Son
Journal:  Clin Endosc       Date:  2022-08-02

Review 3.  The Effectiveness of Drying on Residual Droplets, Microorganisms, and Biofilms in Gastrointestinal Endoscope Reprocessing: A Systematic Review.

Authors:  Hefeng Tian; Jiao Sun; Shaoning Guo; Xuanrui Zhu; Han Feng; Yijin Zhuang; Xiu Wang
Journal:  Gastroenterol Res Pract       Date:  2021-04-08       Impact factor: 2.260

4.  A Prospective, Randomized Comparison of Duodenoscope Reprocessing Surveillance Methods.

Authors:  Travis J De Wolfe; Nasia Safdar; Megan Meller; John Marx; Patrick R Pfau; Eric M Nelsen; Mark E Benson; Anurag Soni; Mark Reichelderfer; Megan Duster; Deepak V Gopal
Journal:  Can J Gastroenterol Hepatol       Date:  2019-11-18

Review 5.  Recent Advances in Endoscope Disinfection: Where Do We Stand in the COVID era?

Authors:  Tiffany Chua; Nasir Halim; Sofiya Reicher
Journal:  Tech Innov Gastrointest Endosc       Date:  2020-10-19
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.