Literature DB >> 30124071

Effect of Screw Length and Geometry on Interfragmentary Compression in a Simulated Proximal Pole Scaphoid Fracture Model.

Samik Patel1, Juan M Giugale1, Richard E Debski1, John R Fowler1.   

Abstract

Background: The objective of this study was to determine interfragmentary compression forces based on screw length and geometry for simulated proximal scaphoid fractures.
Methods: Sixty-four foam model simulated fractures were stabilized with screws of various length (10 mm, 18 mm, 20 mm, or 24 mm) and geometry (central threadless or fully threaded) across a proximal fracture. Interfragmentary compression was measured at the simulated fracture site upon fixation. An independent sample t test and 1-way analysis of variance were performed to assess differences in interfragmentary compression.
Results: Fixation utilizing a 10-mm screw generated significantly less interfragmentary compression than fixation utilizing a 20-mm or 24-mm screw. When accounting for both screw length and geometry, an 18-mm central threadless screw generated greater interfragmentary compression than a 20-mm and 24-mm fully threaded screw; there was no significant difference in compression between an 18-mm and 24-mm central threadless screw. Conclusions: The design of headless compression screws allows for maximal interfragmentary compression at the screw midpoint; we questioned whether a short screw centered on the fracture site resulted in superior compression to a longer, noncentered screw. Our data suggest that centering a small screw (10 mm) along a proximal fracture generates significantly less interfragmentary compression than a longer, noncentered screw. Our results demonstrate that balance between maximizing screw length and centering the screw on the fracture is vital toward maximizing interfragmentary compression for the fixation of proximal third scaphoid fractures.

Entities:  

Keywords:  compression force; fixation; proximal pole; scaphoid fracture; screw

Year:  2018        PMID: 30124071      PMCID: PMC7225878          DOI: 10.1177/1558944718795281

Source DB:  PubMed          Journal:  Hand (N Y)        ISSN: 1558-9447


  23 in total

1.  Interfragmentary compression across a simulated scaphoid fracture--analysis of 3 screws.

Authors:  Gordon P Beadel; Louis Ferreira; James A Johnson; Graham J W King
Journal:  J Hand Surg Am       Date:  2004-03       Impact factor: 2.230

2.  A biomechanical study on variation of compressive force along the Acutrak 2 screw.

Authors:  Hari Kovilazhikathu Sugathan; Max Kilpatrick; Tom J Joyce; John W K Harrison
Journal:  Injury       Date:  2011-08-11       Impact factor: 2.586

3.  Compression forces generated by Mini bone screws--a comparative study done on bone model.

Authors:  Deepthi Nandan Adla; C Kitsis; A W Miles
Journal:  Injury       Date:  2005-01       Impact factor: 2.586

4.  A central threadless shaft screw is better than a fully threaded variable pitch screw for unstable scaphoid nonunion: a biomechanical study.

Authors:  Il-Hyun Koh; Ho-Jung Kang; Ji-Sup Kim; Seong-Jin Park; Yun-Rak Choi
Journal:  Injury       Date:  2015-01-20       Impact factor: 2.586

Review 5.  Union rates after proximal scaphoid fractures; meta-analyses and review of available evidence.

Authors:  N Eastley; H Singh; J J Dias; N Taub
Journal:  J Hand Surg Eur Vol       Date:  2012-06-26

6.  A comparison of two headless compression screws for operative treatment of scaphoid fractures.

Authors:  Ruby Grewal; Joseph Assini; David Sauder; Louis Ferreira; Jim Johnson; Kenneth Faber
Journal:  J Orthop Surg Res       Date:  2011-06-07       Impact factor: 2.359

7.  The vascularity of the scaphoid bone.

Authors:  R H Gelberman; J Menon
Journal:  J Hand Surg Am       Date:  1980-09       Impact factor: 2.230

Review 8.  Surgical compared with conservative treatment for acute nondisplaced or minimally displaced scaphoid fractures: a systematic review and meta-analysis of randomized controlled trials.

Authors:  Geert A Buijze; Job N Doornberg; John S Ham; David Ring; Mohit Bhandari; Rudolf W Poolman
Journal:  J Bone Joint Surg Am       Date:  2010-06       Impact factor: 5.284

9.  Percutaneous screw fixation versus conservative treatment for fractures of the waist of the scaphoid: a prospective randomised study.

Authors:  M M McQueen; M K Gelbke; A Wakefield; E M Will; C Gaebler
Journal:  J Bone Joint Surg Br       Date:  2008-01

10.  Screw fixation of scaphoid fractures: a biomechanical assessment of screw length and screw augmentation.

Authors:  Seth D Dodds; Manohar M Panjabi; Joseph F Slade
Journal:  J Hand Surg Am       Date:  2006-03       Impact factor: 2.230

View more