| Literature DB >> 30122893 |
Matthew Hartman1, Michael Rauser2, Matthew Brucks2, K V Chalam2.
Abstract
PURPOSE: The purpose of this study was to compare the incidence of anterior capsular contraction syndrome (ACCS) in cataract patients after implantation with one of two most commonly used hydrophobic acrylic lenses.Entities:
Keywords: acrylic resins; biocompatible materials; capsular phimosis; capsule opacification; intraocular lens implantation; treatment outcome
Year: 2018 PMID: 30122893 PMCID: PMC6087024 DOI: 10.2147/OPTH.S172251
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Eyes with ACCS risk factors
| Risk factors | ACCS (−)
| ACCS (+)
| ||
|---|---|---|---|---|
| SN60WF | ZCB00 | SN60WF | ZCB00 | |
| PXE syndrome | 5 | 16 | 0 | 0 |
| Uveitis | 1 | 4 | 1 | 0 |
| Retinitis pigmentosa | 0 | 0 | 0 | 0 |
| History of retinal surgery | 1 | 6 | 0 | 1 |
| Diabetic retinopathy | 28 | 68 | 1 | 1 |
| Total | 35 | 94 | 2 | 2 |
Notes: (−), absence of ACCS. (+), presence of ACCS.
Single eye with two risk factors.
Number of patients with at least one risk factor: 37/571 (SN60WF) and 74/476 (ZCB00).
In all, 17 eyes had two risk factors and two eyes had three risk factors.
Abbreviations: ACCS, anterior capsular contraction syndrome; PXE, pseudoexfoliation.
Patient demographics and ACCS incidence
| Study parameters | Lens
| |
|---|---|---|
| SN60WF | ZCB00 | |
| Eyes (n) | 571 | 476 |
| Average age (years), mean±SD | 74.2±9.4 | 71.3±11.4 |
| Male (n) | 205 | 203 |
| Female (n) | 366 | 272 |
| Average follow-up (months), median statistic | 4.2 (1–33.6) | 5.7 (1–33.1) |
| ACCS cases (n) | 8 | 1 |
| ACCS incidence (%) | 1.40 | 0.21 |
Abbreviation: ACCS, anterior capsular contraction syndrome.
Lens specifications
| Characteristic | AcrySof SN60WF | Tecnis ZCB00 |
|---|---|---|
| Overall length (mm) | 13 | 13 |
| Material | 1-piece hydrophobic acrylic | 1-piece hydrophobic acrylic |
| Optic | ||
| Diameter (mm) | 6 | 6 |
| Design | Biconvex | Biconvex |
| Edge | Square, interrupted | 360° square edge |
| Haptics | ||
| Design | Modified C | Modified C |
| Position with respect to optic | In-line | Offset, 3-point fixation |
| | 29 | 36.5 |
Note:
Force applied by haptics to the capsular bag.