Comparison of the stability of 1-piece and 3-piece acrylic intraocular lenses in the lens capsule.

Ken Hayashi1, Hideyuki Hayashi.   

Abstract

PURPOSE: To compare the degrees of intraocular lens (IOL) decentration, tilt, and longitudinal movement; the refractive change; and anterior capsule contraction after cataract surgery between eyes with a 1-piece acrylic IOL with soft acrylic loops and eyes with a 3-piece acrylic IOL with rigid poly(methyl methacrylate) loops.
SETTING: Hayashi Eye Hospital, Fukuoka, Japan.
METHODS: Fifty-six patients had implantation of a 1-piece acrylic IOL in 1 eye and a 3-piece acrylic IOL in the fellow eye. The degree of IOL decentration and tilt, the anterior chamber depth (ACD), and the area of anterior capsule opening were measured using Scheimpflug videophotography 3 days and 1, 3, and 6 months postoperatively. The postoperative refractive status was also examined.
RESULTS: The mean degrees of decentration and tilt in the 1-piece IOL group were similar to those in the 3-piece IOL group throughout the follow-up. The ACD did not change after surgery in the 1-piece group but showed significant shallowing in the 3-piece group (P < .0001). The spherical equivalent (SE) did not change in the 1-piece group, while the SE had a significant myopic shift of approximately 0.4 diopter in the 3-piece group. The percentage of anterior capsule contraction was similar between the groups.
CONCLUSIONS: The degree of IOL decentration and tilt and percentage of anterior capsule contraction in eyes with a 1-piece acrylic IOL with soft acrylic loops were similar to those in eyes with a 3-piece acrylic IOL. The longitudinal movement of the 1-piece IOL was less than the movement of the 3-piece IOL, resulting in less postoperative myopic shift.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15767155     DOI: 10.1016/j.jcrs.2004.06.042

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  22 in total

1.  [Intraocular lenses for microincisional cataract surgery].

Authors:  T Kohnen; O K Klaproth
Journal:  Ophthalmologe       Date:  2010-02       Impact factor: 1.059

2.  Temporary changes of visual outcomes and anterior chamber parameters after phacoemulsification and low-add-power segmented intraocular lens implantation for primary angle closure disease.

Authors:  Yohei Chikaraishi; Yoshikuni Arakaki; Hideki Koizumi
Journal:  Int Ophthalmol       Date:  2021-03-22       Impact factor: 2.031

3.  Influence of tilt and decentration of scleral-sutured intraocular lens on ocular higher-order wavefront aberration.

Authors:  T Oshika; G Sugita; K Miyata; T Tokunaga; T Samejima; C Okamoto; Y Ishii
Journal:  Br J Ophthalmol       Date:  2006-08-16       Impact factor: 4.638

4.  Investigation of the stability of one-piece acrylic intraocular lenses in cataract surgery and in combined vitrectomy surgery.

Authors:  T Iwase; K Sugiyama
Journal:  Br J Ophthalmol       Date:  2006-08-16       Impact factor: 4.638

5.  [Refractive predictability and stability of three-piece versus single-piece intraocular lenses in patients with high axial myopia].

Authors:  A Frings; I Steinbrugger; E Wissiak; B Vidic; Y El-Shabrawi; N Ardjomand
Journal:  Ophthalmologe       Date:  2012-06       Impact factor: 1.059

6.  [Light adjustable lens. New options for customized correction of presbyopia].

Authors:  F H Hengerer; J Böcker; B H Dick; I Conrad-Hengerer
Journal:  Ophthalmologe       Date:  2012-07       Impact factor: 1.059

7.  Effect of a capsular tension ring on prevention of intraocular lens decentration and tilt and on anterior capsule contraction after cataract surgery.

Authors:  Minehiro Takimoto; Ken Hayashi; Hideyuki Hayashi
Journal:  Jpn J Ophthalmol       Date:  2008-11-11       Impact factor: 2.447

8.  Effect of haptic material and number of intraocular lens on anterior capsule contraction after cataract surgery.

Authors:  Sun Young Kim; Ji Wook Yang; Young Chun Lee; Su-Young Kim
Journal:  Korean J Ophthalmol       Date:  2013-01-15

9.  Influences of anterior capsule polishing on effective lens position after cataract surgery: a randomized controlled trial.

Authors:  Yang Gao; Guang-Fu Dang; Xu Wang; Lian Duan; Xin-Yi Wu
Journal:  Int J Clin Exp Med       Date:  2015-08-15

10.  Stability of Various Types of Aspheric Intraocular Lenses After Implantation: A One-Year Retrospective Study.

Authors:  Yuan Ning; Yushuang Shao; Jiangyue Zhao; Jinsong Zhang; Mingwu Wang; Yu Qin
Journal:  Int J Gen Med       Date:  2021-06-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.