Sunit Tolia1, Zubair Khan1, Gunjan Gholkar1, Marcel Zughaib1.
Abstract
BACKGROUND: Routine ambulatory echocardiographic estimates of left ventricular (LV) filling pressures are not cost-effective and are occasionally fraught with anatomic, physiologic as well as logistical limitations. The use of implantable hemodynamic devices such as CardioMEMS Heart Failure (HF) System has been shown to reduce HF-related readmission rates by remote monitoring of LV filling pressures. Little is known about the correlation between CardioMEMS and echocardiography-derived estimates of central hemodynamics.Entities:
Year: 2018 PMID: 30116638 PMCID: PMC6079522 DOI: 10.1155/2018/8568356
Source DB: PubMed Journal: Cardiol Res Pract ISSN: 2090-0597 Impact factor: 1.866
Baseline characteristics.
|
| |
|---|---|
| Age (mean ± SD) | 74 ± 9 |
| Gender ( | |
| Male | 10 (59%) |
| BMI (mean ± SD) | 31.7 ± 6.3 |
| Smoking ( | |
| Never smoker | 7 (41%) |
| Current smoker | 1 (6%) |
| Former smoker | 9 (53%) |
| CHF category | |
| HFrEF | 15 (75%) |
| HFpEF | 5 (25%) |
| Hypertension ( | 17 (100%) |
| Hyperlipidemia ( | 17 (100%) |
| Diabetes mellitus ( | 10 (59%) |
| Coronary artery disease ( | 14 (82%) |
| Peripheral arterial disease ( | 3 (18%) |
| Atrial fibrillation/other arrhythmias ( | 6 (35%) |
| Chronic lung disease ( | 9 (53%) |
| Chronic kidney disease ( | 4 (23%) |
|
| |
| Baseline medical therapy for congestive heart failure | |
| Beta-blockers ( | 17 (100%) |
| ACE-I or ARB ( | 7 (41%) |
| ARB/neprilysin inhibitor ( | 8 (47%) |
| Aldosterone receptor blocker ( | 12 (71%) |
| Diuretics ( | 17 (100%) |
| Other vasodilators ( | 2 (12%) |
| Digoxin ( | 3 (18%) |
Invasive and noninvasive hemodynamic data from CardioMEMS implant and simultaneous echocardiographic assessment.
|
| |
|---|---|
| Invasive hemodynamics at CardioMEMS implant | |
| PAsP, mmHg (mean ± SD) | 49 ± 16 |
| PAdP, mmHg (mean ± SD) | 21 ± 7 |
|
| |
| Hemodynamics by CardioMEMS | |
| PAsP, mmHg (mean ± SD) | 42 ± 8 |
| PAdP, mmHg (mean ± SD) | 18 ± 4 |
|
| |
| Estimated hemodynamics by echocardiogram | |
| PAsP, mmHg (mean ± SD) | 42 ± 8 |
| LAP, mmHg (mean ± SD) | 19 ± 5 |
LAP, left atrial pressure; mPAP, mean pulmonary artery pressure; PAdP, pulmonary artery diastolic pressure; PAsP, pulmonary artery systolic pressure.
Figure 1Bland–Altman analysis (a) and linear correlation (b) between CardioMEMS-derived pulmonary artery diastolic pressure (PAdP) and echocardiography-estimated left atrial pressure (LAP) by Nagueh formula (estimated LAP = 1.24 × (E/e') + 1.9).
Figure 2Bland–Altman analysis (a) and linear correlation (b) between CardioMEMS-derived pulmonary artery systolic pressure (PAsP) and echocardiography-estimated PAsP derived from tricuspid regurgitation jet velocity.