Comparison of a radiofrequency-based wireless pressure sensor to swan-ganz catheter and echocardiography for ambulatory assessment of pulmonary artery pressure in heart failure.

Hugo E Verdejo1, Pablo F Castro, Roberto Concepción, Marcela A Ferrada, Mario A Alfaro, Milton E Alcaíno, Carlos C Deck, Robert C Bourge.   

Abstract

OBJECTIVES: The goal of this work was to evaluate the accuracy of a new heart failure (HF) sensor (HFS) (Heart Failure Sensor, CardioMEMS Inc., Atlanta, Georgia) pulmonary artery pressure (PAP) monitoring compared with Swan-Ganz (SG) (Hospira, Inc., Lake Forest, Illinois) catheterization and echocardiography (ECHO) in ambulatory HF patients.
BACKGROUND: There is an increasing interest in the development of ambulatory monitoring devices aiming to adjust therapy and prevent hospitalizations in HF patients.
METHODS: Twelve patients with HF and New York Heart Association functional class II to IV were included in this study. The HFS was deployed into the pulmonary artery under angiography, allowing wireless PAP measurement. Two independent blind operators performed 3 HFS measurements at each visit, with simultaneous ECHO at 2, 14, 30, 60, and 90 days. Swan-Ganz catheterization was performed at 0 and 60 days. Linear regression was used as a measure of agreement. Variability between methods and interobserver variability were evaluated by Bland-Altman analysis.
RESULTS: Mean age was 63 +/- 14.6 years. Systolic PAP was 64 +/- 22 mm Hg and 58 +/- 22 mm Hg for HFS and SG, respectively (p < 0.01). Both methods showed a significant correlation (r2 = 0.96 baseline, r2 = 0.90 follow-up, p < 0.01), with a mean difference of 6.2 +/- 4.5 mm Hg. Diastolic PAP was 23 +/- 14 mm Hg and 28 +/- 16 mm Hg for HFS and SG, respectively (r2 = 0.88 baseline, r2 = 0.48 follow-up, p < 0.01), with a mean difference of -1.6 +/- 6.8 mm Hg. Systolic PAP was 60 +/- 20 mm Hg and 62 +/- 12 mm Hg for HFS and ECHO, respectively (r2 = 0.75, p < 0.01), with a mean difference of -2.6 +/- 11 mm Hg. There was no significant interobserver difference.
CONCLUSIONS: The HFS provides an accurate method for PAP assessment in the intermediate follow-up of HF patients.

Entities:  

Mesh:

Year:  2007        PMID: 18154961     DOI: 10.1016/j.jacc.2007.06.061

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  31 in total

1.  [Diminishing borders between cardiology and cardiothoracic surgery: quo vadis?].

Authors:  T Schilling; R Bekeredjian; A Haverich; H A Katus
Journal:  Chirurg       Date:  2010-12       Impact factor: 0.955

Review 2.  Pathophysiology of the transition from chronic compensated and acute decompensated heart failure: new insights from continuous monitoring devices.

Authors:  Philip B Adamson
Journal:  Curr Heart Fail Rep       Date:  2009-12

3.  Hemodynamic-guided heart-failure management using a wireless implantable sensor: Infrastructure, methods, and results in a community heart failure disease-management program.

Authors:  Rita Jermyn; Amit Alam; Jessica Kvasic; Omar Saeed; Ulrich Jorde
Journal:  Clin Cardiol       Date:  2016-11-23       Impact factor: 2.882

4.  Device monitoring strategies in acute heart failure syndromes.

Authors:  Michael A Samara; W H Wilson Tang
Journal:  Heart Fail Rev       Date:  2011-09       Impact factor: 4.214

Review 5.  Home monitoring for heart failure management.

Authors:  Anh L Bui; Gregg C Fonarow
Journal:  J Am Coll Cardiol       Date:  2012-01-10       Impact factor: 24.094

6.  Therapeutic implications of implantable device-based monitoring of patients with heart failure.

Authors:  Eric Popjes; John P Boehmer
Journal:  Curr Treat Options Cardiovasc Med       Date:  2008-09

7.  Update on device technologies for monitoring heart failure.

Authors:  Bharat Singh; Stuart D Russell; Alan Cheng
Journal:  Curr Treat Options Cardiovasc Med       Date:  2012-10

8.  Identification of hemodynamically unstable arrhythmias using subcutaneous photoplethysmography.

Authors:  Robert G Turcott; Todd J Pavek
Journal:  J Cardiovasc Electrophysiol       Date:  2009-10-20

9.  Continuous in vivo blood pressure measurements using a fully implantable wireless SAW sensor.

Authors:  Olive H Murphy; Mohammad Reza Bahmanyar; Alessandro Borghi; Christopher N McLeod; Manoraj Navaratnarajah; Magdi H Yacoub; Christofer Toumazou
Journal:  Biomed Microdevices       Date:  2013-10       Impact factor: 2.838

10.  Noninvasive assessment of murine pulmonary arterial pressure: validation and application to models of pulmonary hypertension.

Authors:  Hélène B Thibault; Baptiste Kurtz; Michael J Raher; Rahamthulla S Shaik; Aaron Waxman; Geneviève Derumeaux; Elkan F Halpern; Kenneth D Bloch; Marielle Scherrer-Crosbie
Journal:  Circ Cardiovasc Imaging       Date:  2009-12-31       Impact factor: 7.792

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.