Literature DB >> 30100943

Capabilities of the Monte Carlo Simulation Codes for Modeling of a Small Animal SPECT Camera.

Alireza Sadremomtaz1, Zeinab Telikani1.   

Abstract

PURPOSE: This study aims to compare Monte Carlo-based codes' characteristics in the determination of the basic parameters of a high-resolution single photon emission computed tomography (HiReSPECT) scanner.
METHODS: The geometry of this dual-head gamma camera equipped with a pixelated CsI(Na) scintillator and lead hexagonal hole collimator were accurately described in the GEANT4 Application for the Tomographic Emission (GATE), Monte Carlo N-particle extended (MCNP-X), and simulation of imaging nuclear detectors (SIMIND) codes. We implemented simulation procedures similar to the experimental test for calculation of the energy spectra, spatial resolution, and sensitivity of HiReSPECT by using 99mTc sources.
RESULTS: The energy resolutions simulated by SIMIND, MCNP-X, and GATE were 17.53, 19.24, and 18.26%, respectively, while it was calculated at 19.15% in experimental test. The average spatial resolutions of the HiReSPECT camera at 2.5 cm from the collimator surface simulated by SIMIND, MCNP-X, and GATE were 3.18, 2.9, and 2.62 mm, respectively, while this parameter was reported at 2.82 mm in the experiment test. The sensitivities simulated by SIMIND, MCNP-X, and GATE were 1.44, 1.27, and 1.38 cps/μCi, respectively, on the collimator surface.
CONCLUSIONS: Comparison between simulation and experimental results showed that among these MC codes, GATE enabled to accurately model realistic SPECT system and electromagnetic physical processes, but it required more time and hardware facilities to run simulations. SIMIND was the most flexible and user-friendly code to simulate a SPECT camera, but it had limitations in defining the non-conventional imaging device. The most important characteristics like time and speed of simulation, preciseness of results, and user-friendliness should be considered during simulations.

Entities:  

Keywords:  GATE; Gama camera; HiReSPECT; MCNP; SIMIND

Year:  2018        PMID: 30100943      PMCID: PMC6066489          DOI: 10.1007/s13139-018-0530-0

Source DB:  PubMed          Journal:  Nucl Med Mol Imaging        ISSN: 1869-3474


  13 in total

1.  Estimation of fan beam and parallel beam parameters in a wire mesh design.

Authors:  Abdollah Khorshidi; Mansour Ashoor; S Hamed Hosseini; Azimeh Rajaee
Journal:  J Nucl Med Technol       Date:  2012-01-11

Review 2.  Evaluation of collimators' response: round and hexagonal holes in parallel and fan beam.

Authors:  Abdollah Khorshidi; Mansour Ashoor; S Hamed Hosseini; Azimeh Rajaee
Journal:  Prog Biophys Mol Biol       Date:  2012-05-16       Impact factor: 3.667

3.  GATE V6: a major enhancement of the GATE simulation platform enabling modelling of CT and radiotherapy.

Authors:  S Jan; D Benoit; E Becheva; T Carlier; F Cassol; P Descourt; T Frisson; L Grevillot; L Guigues; L Maigne; C Morel; Y Perrot; N Rehfeld; D Sarrut; D R Schaart; S Stute; U Pietrzyk; D Visvikis; N Zahra; I Buvat
Journal:  Phys Med Biol       Date:  2011-01-20       Impact factor: 3.609

4.  The effect of energy and source location on gamma camera intrinsic and extrinsic spatial resolution: an experimental and Monte Carlo study.

Authors:  Maria Holstensson; Mike Partridge; Susan E Buckley; Glenn D Flux
Journal:  Phys Med Biol       Date:  2010-03-02       Impact factor: 3.609

5.  Resolution-recovery-embedded image reconstruction for a high-resolution animal SPECT system.

Authors:  Navid Zeraatkar; Salar Sajedi; Mohammad Hossein Farahani; Hossein Arabi; Saeed Sarkar; Pardis Ghafarian; Arman Rahmim; Mohammad Reza Ay
Journal:  Phys Med       Date:  2014-06-27       Impact factor: 2.685

6.  SIMIND Monte Carlo simulation of a single photon emission CT.

Authors:  M T Bahreyni Toossi; J Pirayesh Islamian; M Momennezhad; M Ljungberg; S H Naseri
Journal:  J Med Phys       Date:  2010-01

7.  EduGATE - basic examples for educative purpose using the GATE simulation platform.

Authors:  Uwe Pietrzyk; Abdelhamid Zakhnini; Markus Axer; Sophie Sauerzapf; Didier Benoit; Michaela Gaens
Journal:  Z Med Phys       Date:  2012-08-19       Impact factor: 4.820

8.  Modulation transfer function assessment in parallel beam and fan beam collimators with square and cylindrical holes.

Authors:  Abdollah Khorshidi; Mansour Ashoor
Journal:  Ann Nucl Med       Date:  2014-02-26       Impact factor: 2.668

9.  Implementation of angular response function modeling in SPECT simulations with GATE.

Authors:  P Descourt; T Carlier; Y Du; X Song; I Buvat; E C Frey; M Bardies; B M W Tsui; D Visvikis
Journal:  Phys Med Biol       Date:  2010-04-14       Impact factor: 3.609

10.  Performance evaluation of a newly developed high-resolution, dual-head animal SPECT system based on the NEMA NU1-2007 standard.

Authors:  Vahideh Moji; Navid Zeratkar; Mohammad Hossein Farahani; Mahmoud Reza Aghamiri; Salar Sajedi; Behnoosh Teimourian; Pardis Ghafarian; Saeed Sarkar; Mohammad Reza Ay
Journal:  J Appl Clin Med Phys       Date:  2014-11-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.