Martin N Stienen1,2, Jorn Fierstra1,2, Athina Pangalu2,3, Luca Regli1,3, Oliver Bozinov1,3.
Abstract
BACKGROUND: Recently, the use of intraoperative magnetic resonance imaging (ioMRI) has evolved in neurosurgery. Challenges related to ioMRI-augmented procedures are significant, since the magnetic field creates a potentially hazardous environment. Strict safety guidelines in the operating room (OR) are necessary. Checklists can minimize errors while increasing efficiency and improving workflow.Entities:
Keywords: 2-room concept; Complications; Infection; Safety; Surgical checklist; Transsphenoidal surgery; intraoperative magnetic resonance imaging
Year: 2019 PMID: 30099512 PMCID: PMC6531895 DOI: 10.1093/ons/opy205
Source DB: PubMed Journal: Oper Neurosurg (Hagerstown) ISSN: 2332-4252 Impact factor: 2.703
FIGURE 1.The setting of the ioMRI-OR is outlined A, where the OR is separated from the ioMRI suite by a sliding double-doorB.
Checklist for Open Microsurgical Tumor Resections Using the ioMRI
| Before sterile draping | |
|---|---|
| Questions to OR/circulation nurse | Response |
| Lower head coil installed?* | Water-tightly packed (cover: 80 × 80 cm). |
| Sticky cloth (75 × 90 cm)a?* | Attached to the shoulder. |
| 2 × Moltexb?* | Placed before and behind pins. |
| PEEK pins installedc?** | Quantity: |
| Earplugs?** | Applied bilaterally. |
| Electrocautery plate installed?* | Location: |
| Hands, Feet?* | Positioned straight. |
| Phantom?* | Fits well. |
| Questions to anesthesiology team | Response |
| Electrical cables?** | Not in contact with the patient. |
| Tracheal tube?* | Positioned flat. |
| Hose?* | Not fixed with towel clamps. |
| Patient positioned on a dry cover (after insertion of the urinary catheter)?* | Double-checked. |
| Questions to neuromonitoring team | Response |
| Skin-skin contact?** | Isolation material installed. |
| Electrical cables?** | Fixed over transfer table. |
|
| |
|
|
|
| Sterile towel clamps?** | Quantity: |
|
| |
|
|
|
| Scar?*** | Covered hermetic with sterile film. |
| Collection bag fluid below head?* | Removed. |
| Sterile towel clamps?** | Removed. Quantity: |
| Navigation star?** | Removed. |
| Leyla adapter?** | Removed. |
| Electrocautery plate and cable?** | Removed. |
| Neuromonitoring device?** | Removed from the table. |
|
| |
|
|
|
| Upper head coil installed?* | Installed. Mind the “little man”! (°) |
| Cover 125cmd?* | Overlaid. |
| Covering sheets?* | Shortened. |
| White transfer plate? (°)* | Visible. |
| OR table, docking?* | Brought into transfer position. |
| Sliding door MRI?* | Free of obstacles. |
| Coils? Connectors?* | Exposed. |
| Date: __________________ | |
| Signature: ___________________ | Patient etiquette: |
The * indicate the main rationale: * = organizational; ** = magnetic field strength-related; *** = sterility. Most steps are relevant to any 2-room ioMRI-OR suite, whereas some are specifically relevant for the Siemens magnet (Siemens Healthineers, Erlangen, Germany) with NORAS (Noras MRI products, Höchberg, Germany) head coil (marked with °).
aProduct ID 2626, Mundus Medici GmbH, Düsseldorf, Germany; bProduct ID 161008, MoliNea Plus, Hartmann, Neuhausen, Switzerland; cPEEK pins reusable, NORAS MRI products, Höchberg, Germany; dProduct ID 401008125, Schlieren, Switzerland.
Checklist for Transsphenoidal Pituitary Tumor Resections Using the ioMRI Suite
| Before sterile draping | |
|---|---|
| Questions to OR/circulation nurse | Response |
| Lower head coil installed?* | Water-tightly packed (cover: 80 × 80 cm). |
| Sticky cloth (75 × 90 cm)a?* | Attached to the shoulder. |
| 2 ×Moltexb?* | Placed before and behind pins. |
| PEEK pins installedc?** | Quantity: |
| Earplugs?** | Applied bilaterally. |
| Electrocautery plate installed?* | Location: |
| Hands, Feet?* | Positioned straight. |
| Phantom?* | Fits well. |
| Questions to anesthesiology team | Response |
| Electrical cables?** | Not in contact with the patient. |
| Neck?* | Tamponated with fabric. |
| Tracheal tube?* | Positioned flat. |
| Hose?* | Not fixed with towel clamps. |
| Patient positioned on a dry cover (after insertion of the urinary catheter)?* | Double-checked. |
|
| |
|
|
|
| Sterile towel clamps?** | Quantity: |
| “Spitztupfer” installedd?* | Quantity: |
|
| |
|
|
|
| Scar?*** | “Netzel” implantede |
| Collection bag fluid below head?* | Removed. |
| Sterile towel clamps?** | Removed. Quantity: |
| Navigation star?** | Removed. |
| Electrocautery plate and cable?** | Removed. |
|
| |
|
|
|
| Upper head coil installed?* | Installed. Mind the “little man”! (°) |
| Cover 125cmf?* | Overlaid. |
| Covering sheets?* | Shortened. |
| White transfer plate? (°)* | Visible. |
| OR table, docking?* | Brought into transfer position. |
| Sliding door MRI?* | Free of obstacles. |
| Coils? Connectors?* | Exposed. |
| Date: __________________ | |
| Signature: ___________________ | Patient etiquette: |
The ** indicate the main rationale: * = organizational; ** = magnetic field strength-related; *** = sterility. Most steps are relevant to any 2-room ioMRI-OR suite, whereas some are specifically relevant for the Siemens magnet (Siemens Healthineers, Erlangen, Germany) with NORAS (Noras MRI products, Höchberg, Germany) head coil (marked with °)
aProduct ID 2626, Mundus Medici GmbH, Düsseldorf, Germany; bProduct ID 161008, MoliNea Plus, Hartmann, Neuhausen, Switzerland; cPEEK pins reusable, NORAS MRI products, Höchberg, Germany; dProduct ID 35003, Novimed Medizintechnik, Dietikon, Switzerland; e Product ID 35004, Novimed Medizintechnik, Dietikon, Switzerland; f Product ID 401008125, Schlieren, Switzerland.
FIGURE 2.Critical steps are taken before draping. A, It needs to be ensured that the setup consisting of the lower and upper head coil is complete.B, MRI-compatible PEEK pins for the head clamp are used.C, We make sure the phantom for the MRI channel fits well over the upper head coil.
FIGURE 3.Critical steps are taken before transfer. A, The skin is only provisionally readapted by the surgeon after exact hemostasis.B, The skin is then covered hermetically with a sterile film and the bag collecting fluid below the head is removed, along with all sterile towel clamps, navigation star, Leyla adapter, electrocautery plate, and neuromonitoring devices. The cover for the transfer is then placed(C, for transcranial;D, for transsphenoidal surgery). E, Prior to moving the patient, the upper head coil is installed.F, Covers are placed and shortened to avoid getting stuck during transfer.