Literature DB >> 30088532

NORMATIVE VALUES FOR A NEW WATER-PERFUSED HIGH RESOLUTION MANOMETRY SYSTEM.

Rogério Mariotto Bitetti da Silva1, Fernando A M Herbella1, Daniel Gualberto1.   

Abstract

BACKGROUND: Esophageal manometry is the most reliable method to evaluate esophageal motility. High resolution manometry (HRM) provides topographic contour colored plots (Clouse Plots) with simultaneous analysis from the pharynx to the stomach. Both solid state and water-perfused systems are available.
OBJECTIVE: This study aims to determinate the normative data for a new water-perfused HRM.
METHODS: HRM was made in 32 healthy volunteers after 8 hours fasting. HRM system used consisted of a 24-channel water-perfused catheter (Multiplex, Alacer Biomedica, São Paulo, Brazil). The reusable catheter is made of polyvinyl chloride (PVC) with 4.7 mm of diameter. Side holes connected to pressure transducers are spaced 2 cm for the analysis from the pharynx to the lower esophageal sphincter (LES). Holes are spaced 5 mm and 120° in a spiral disposition in the LES area. The sensors encompass 34 cm in total. Upper esophageal sphincter (UES) parameters studied were basal and relaxation pressures. Esophageal body parameters were distal contractile integral (DCI), distal latency (DL) and break. LES parameters studied were basal pressure, integrated residual pressure (IRP), total and abdominal length. Variables are expressed as mean ± standard deviation, median (interquartile range) and percentiles 5-95th.
RESULTS: All volunteers (17 males, aged 22-62 years) completed the study and tolerated the HRM procedure well. Percentiles 5-95th range were calculated: Upper Esophageal Sphincter (UES) basal pressure 16.7-184.37 (mmHg), DL: 6.2-9.1 (s), DCI: 82.72-3836.61 (mmHg.s.cm), break: <7.19 (cm), LES basal pressure: 4.89-37.16 (mmHg), IRP: 0.55-15.45 (mmHg).
CONCLUSION: The performance and normative values obtained for this low-cost water-perfused HRM seems to be adequate for clinical use.

Entities:  

Mesh:

Year:  2018        PMID: 30088532     DOI: 10.1590/S0004-2803.201800000-40

Source DB:  PubMed          Journal:  Arq Gastroenterol        ISSN: 0004-2803


  4 in total

1.  VALIDATION OF A NEW WATER-PERFUSED HIGH-RESOLUTION MANOMETRY SYSTEM.

Authors:  Rogério Mariotto; Fernando A M Herbella; Vera Lucia Ângelo Andrade; Francisco Schlottmann; Marco G Patti
Journal:  Arq Bras Cir Dig       Date:  2021-01-25

2.  The Upper Esophageal Sphincter Distensibility Index Measured Using Functional Lumen Imaging Probe Identifies Defective Barrier Function of the Upper Esophageal Sphincter.

Authors:  Lucie F Calderon; Meredith Kline; Marc Hersh; Kevin P Shah; Suprateek Kundu; Andrew Tkaczuk; Nancy McColloch; AnS Jain
Journal:  J Neurogastroenterol Motil       Date:  2022-07-30       Impact factor: 4.725

3.  Inclusion Body Myositis With Dysphagia Diagnosed Using Pharyngoesophageal High-Resolution Manometry.

Authors:  Toshimi Chiba
Journal:  ACG Case Rep J       Date:  2022-08-31

4.  A randomized clinical trial to assess the effectiveness of pre- and post-surgical pelvic floor physiotherapy for bowel symptoms, pelvic floor function, and quality of life of patients with rectal cancer: CARRET protocol.

Authors:  Cinara Sacomori; Luz Alejandra Lorca; Mónica Martinez-Mardones; Roberto Ignacio Salas-Ocaranza; Guillermo Patricio Reyes-Reyes; Marta Natalia Pizarro-Hinojosa; Jorge Plasser-Troncoso
Journal:  Trials       Date:  2021-07-13       Impact factor: 2.279

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.