| Literature DB >> 30065507 |
Ammar A Oglat1, M Z Matjafri1, Nursakinah Suardi1, Mohammad A Oqlat2, Mostafa A Abdelrahman3, Ahmad A Oqlat3.
Abstract
Medical Doppler ultrasound is usually utilized in the clinical adjusting to evaluate and estimate blood flow in both the major (large) and the minor (tiny) vessels of the body. The normal and abnormal sign waveforms can be shown by spectral Doppler technique. The sign waveform is individual to each vessel. Thus, it is significant for the operator and the clinicians to understand the normal and abnormal diagnostic in a spectral Doppler show. The aim of this review is to explain the physical principles behind the medical Doppler ultrasound, also, to use some of the mathematical formulas utilized in the medical Doppler ultrasound examination. Furthermore, we discussed the color and spectral flow model of Doppler ultrasound. Finally, we explained spectral Doppler sign waveforms to show both the normal and abnormal signs waveforms that are individual to the common carotid artery, because these signs are important for both the radiologist and sonographer to perceive both the normal and abnormal in a spectral Doppler show.Entities:
Keywords: Doppler ultrasound terminology; physics of ultrasound; sonography; spectral Doppler of common carotid artery
Year: 2018 PMID: 30065507 PMCID: PMC6029191 DOI: 10.4103/JMU.JMU_11_17
Source DB: PubMed Journal: J Med Ultrasound ISSN: 0929-6441
Figure 1Basic components of ultrasound probe (transducer)
Figure 2Group of images explaining the Doppler Effect
Figure 3Spectral Doppler evaluation of uterine artery
Main functions of Doppler device parameter
| Name of Doppler device parameter | Main functions |
|---|---|
| Mean velocity evaluation | This exam supply evaluation of the precision of the Doppler system’s evaluate of the mean Doppler scattered speed. In addition to the precision of the colour Doppler evaluate of the mean scattered speed |
| Range gate | Helps reveal blood flow signal wave |
| Sample volume or sample length | This exam evaluate the sensitivity of range gate to make sure if it is extreme sensitive at the centre position of the gate |
| Maximum velocity precision | This exam supply evaluation of the precision of the Doppler system’s evaluate of the maximum Doppler scattered speed. In addition to the precision of the reveal the degree of arterial narrowing or stenosis |
| Lowest detectable speed | This exam supply evaluation of the lowest speed that it is likely to show unambiguously |
| Highest detectable speed | This exam supply evaluation of the highest speed that it is likely to show unambiguously on both the colour Doppler image or on the PW Doppler spectrum. The highest speed with some diseases or stenosis may be reach up to 500-600 cm/s and can show this speed on the spectrum without aliasing |
| Spectral broadening | This exsam supply evaluation of the spectral Doppler broadening which cause by range of angles |
| Flow direction | This exam supply ability of the differentiate between flow towards and away from the probe |
| Angle correction | This exam supply ability to measures the accuracy of the angle correction the device |
| Wall filter | This exam remove intense signals from the vessel wall motion |
Source from: Browne, J.E., A review of Doppler ultrasound quality assurance protocols and test devices. Physica Medica, 2014. 30(7): p. 742-751. PW: Pulsed wave
Figure 4Phasicity, graphs clarify the different waveforms. The expressions terms utilized to describe the grade of waveform ripple empirically describe the speed and acceleration features of the waveform. Keep in mind that all of pulsatile, phasic and nonphasic flow waveforms all of them have Phasicity [19]