Szymon Bialka1, Maja Copik1, Katarzyna Rybczyk1, Aleksander Owczarek2, Ewa Jedrusik2, Damian Czyzewski3, Marek Filipowski3, Eva Rivas4,5, Kurt Ruetzler4, Lukasz Szarpak6, Hanna Misiolek1.
Abstract
BACKGROUND: Electrical impedance tomography (EIT) is a tool to monitor regional ventilation distribution in patient's lungs under general anesthesia. The objective of this study was to assess the regional ventilation distribution using different driving pressures (DP) during high frequency jet ventilation (HFJV).Entities:
Keywords: Driving pressure; High frequency; Inspiratory pressure; Jet ventilation
Mesh:
Year: 2018 PMID: 30064377 PMCID: PMC6069840 DOI: 10.1186/s12871-018-0552-2
Source DB: PubMed Journal: BMC Anesthesiol ISSN: 1471-2253 Impact factor: 2.217
Fig. 1Flow diagram of the study
Results of parametric analysis of variance with repeated measurements and descriptive statistics
| DP 1.5 | DP 2.5 | PGROUP | PTIME | PINTERACTION | |||||
|---|---|---|---|---|---|---|---|---|---|
| 0 min | 3 min | 5 min | 0 min | 3 min | 5 min | ||||
| SBP [mmHg] | 114.5 ± 22.8 | 113.2 ± 18.9 | 110.9 ± 16.7 | 116.1 ± 24.8 | 113.0 ± 23.4 | 108.6 ± 22.2 | 0.95 |
| 0.55 |
| DBP [mmHg] | 76.3 ± 14.6 | 73.3 ± 10.8 | 72.5 ± 10.1 | 76.2 ± 14.0 | 74.5 ± 13.6 | 71.1 ± 13.8 | 0.97 |
| 0.60 |
| MBP [mmHg] | 92.0 ± 16.7 | 89.6 ± 14.0 | 87.8 ± 12.3 | 92.7 ± 19.1 | 90.5 ± 15.1 | 87.6 ± 17.2 | 0.89 |
| 0.92 |
| HR [beats/min] | 80.7 ± 13.0 | 79.3 ± 13.3 | 78.9 ± 12.9 | 80.9 ± 12.9 | 80.7 ± 13.6 | 79.5 ± 14.4 | 0.83 | 0.06 | 0.56 |
| SpO2 [%] | 97.5 ± 2.9 | 97.5 ± 3.2 | 96.9 ± 3.8 | 98.3 ± 1.7 | 98.0 ± 2.5 | 98.1 ± 2.1 | 0.23 | 0.20 | 0.18 |
| PIP [cm H2O] | 6.2 ± 2.8 | 6.3 ± 2.7 | 6.3 ± 2.7 | 9.4 ± 4.3 | 9.3 ± 4.2 | 9.5 ± 4.2 |
| 0.47 | 0.25 |
| MAP [cm H2O] | 2.5 ± 1.3 | 2.5 ± 1.3 | 2.4 ± 1.2 | 4.5 ± 2.2 | 4.4 ± 2.1 | 4.4 ± 2.1 |
| 0.28 | 0.75 |
| TV [mL] | 121.6 ± 6.8 | 121.5 ± 6.7 | 121.5 ± 6.8 | 149.2 ± 6.5 | 150.0 ± 5.3 | 149.8 ± 5.2 |
| 0.42 | 0.44 |
| MV [L] | 21.8 ± 1.2 | 21.7 ± 1.1 | 22.0 ± 1.5 | 26.5 ± 0.9 | 26.6 ± 0.6 | 26.6 ± 0.6 |
| 0.28 | 0.43 |
Interaction – between group and times
Fig. 2Comparison Systolic Arterial Pressure (SBP, mmHg), Diastolic Arterial Pressure (DAP, mmHg) and Mean Arterial Pressure (MAP, mmHg) in 5th minute observed time in DP 1,5 and DP 2, 5 group
Fig. 3Comparison Peak Inspiratory Pressure (PIP, cm H2O) and Mean Inspiratory Pressure (MIP, cm H2O) in 5th minute observed time in DP 1, 5 and DP 2,5 group
Fig. 4Comparison of Electrical Impedance Tomography (EIT) images in ROI 1 to ROI 4 in DP 1, 5 and DP 2,5 group