Intraocular lens power calculation using a Placido disk-Scheimpflug tomographer in eyes that had previous myopic corneal excimer laser surgery.

Giacomo Savini1, Kenneth J Hoffer2, Domenico Schiano-Lomoriello2, Piero Barboni2.   

Abstract

PURPOSE: To assess the accuracy of different methods to calculate the intraocular lens (IOL) power from measurements obtained with an optical biometer combined with a Placido disk-Scheimpflug tomographer (Galilei G6) in eyes that had previous myopic excimer laser surgery.
SETTING: G.B. Bietti Foundation, Rome, Italy.
DESIGN: Prospective case series.
METHODS: When prerefractive surgery data were available, the following methods were evaluated: the Barrett True-K, total corneal power 1 (TCP 1) with double-K SRK/T, TCP with double-K Holladay 1, as well as the Masket and Savini methods. In all eyes, these "no-history" methods were studied: Barrett True-K No-History, TCP 1 using the American Society of Cataract and Refractive Surgery calculator, a ray-tracing program, and the Shammas No-History formula.
RESULTS: In eyes with historical data (n = 15), the Masket method, Savini method, and TCP 1 with double-K SRK/T produced the highest percentage of cases with an absolute prediction error in refraction of 0.50 diopters (D) or less and the lowest median absolute error (0.32 D, 0.28 D, and 0.25 D, respectively). In the whole cohort (N = 22), the most accurate no-history method was paraxial ray tracing, which had the highest percentage of eyes with an absolute prediction error of 0.50 D or less and the lowest median absolute error of 0.31 D.
CONCLUSION: Different solutions can provide accurate IOL power calculations when using the measurements obtained with the optical biometer combined with a Placido-dual Scheimpflug tomographer.
Copyright © 2018 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2018        PMID: 30049566     DOI: 10.1016/j.jcrs.2018.05.018

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  Accuracy of optimized Sirius ray-tracing method in intraocular lens power calculation.

Authors:  Li-Qing Wei; Ying-Hui Fu; Wei-Hua Pan; Li Nie; Yu Chen; Gui-Fang Liu; Zhen-Bin Qian
Journal:  Int J Ophthalmol       Date:  2022-02-18       Impact factor: 1.779

2.  Excimer Laser Corneal Refractive Surgery in the Clinic: A Systematic Review and Meta-analysis.

Authors:  Huang Zhang; Mingming Li; Zhimin Cen
Journal:  Comput Math Methods Med       Date:  2022-06-15       Impact factor: 2.809

3.  Intraocular Lens power calculation after laser refractive surgery: A Meta-Analysis.

Authors:  Hui Chen; Xinyi Chen; Hanle Wang; Zhi Fang; Ke Yao
Journal:  Sci Rep       Date:  2020-02-14       Impact factor: 4.379