Stent and leaflet stresses in 26-mm, third-generation, balloon-expandable transcatheter aortic valve.

Yue Xuan1, Danny Dvir2, Zhongjie Wang1, Trek Mizoguchi1, Jian Ye3, Julius M Guccione1, Liang Ge1, Elaine E Tseng4.   

Abstract

OBJECTIVES: Transcatheter aortic valve replacement has proven successful in treating intermediate-risk, high-risk, and inoperable patients with severe aortic stenosis. Third-generation, balloon-expandable transcatheter aortic valves were developed with an outer sealing skirt to reduce paravalvular leakage. As transcatheter aortic valve replacement use expands, long-term durability questions remain. Valve design influences durability, where regions of increased leaflet stress are vulnerable to early degeneration. However, third-generation transcatheter aortic valve stresses are unknown. Our goals were to determine the stent and leaflet stresses of third-generation, balloon-expandable transcatheter aortic valves.
METHODS: The commercial 26-mm Edwards SAPIEN 3 valve (Edwards Lifesciences, Inc, Irvine, Calif) underwent high-resolution micro-computed tomography scanning to develop a precise 3-dimensional geometric mesh of the stent and valve. Leaflet material properties were obtained from surgical bioprostheses, and stent material properties were based on cobalt-chromium. Simulations of systemic pressure loading were performed, and stress was calculated using finite element analyses.
RESULTS: At diastole, maximum and minimum principal stresses on transcatheter aortic valve leaflets were 2.7 MPa and -0.47 MPa, respectively. Peak leaflet stresses were observed at upper leaflet commissures, at their connection to the stent. Maximum and minimum principal stresses for the stent were 38.2 MPa and -44.4 MPa, respectively, at 80 mm Hg and were located just below the commissural stent.
CONCLUSIONS: Stress analysis of the 26-mm SAPIEN 3 valve using exact geometry from high-resolution scans demonstrated that peak stresses for both transcatheter aortic valve stent and leaflets were present at commissural tips where leaflets were attached. These regions would be most likely to initiate degeneration. The Dacron skirt had minimal effect on stresses on leaflets and stent. Published by Elsevier Inc.

Entities:  

Keywords:  durability; leaflet; stress; transcatheter aortic valve

Mesh:

Year:  2018        PMID: 30041923     DOI: 10.1016/j.jtcvs.2018.04.115

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  5 in total

1.  Stress in the time of transcatheter aortic valve replacement.

Authors:  Muralidhar Padala
Journal:  J Thorac Cardiovasc Surg       Date:  2018-11-08       Impact factor: 5.209

2.  Percutaneous aortic and mitral valve repair - from bench testing to simulators and clinical data: Percutaneous aortic and mitral valve repair.

Authors:  Paul T L Chiam; Huay Cheem Tan
Journal:  AsiaIntervention       Date:  2020-12-02

3.  Impact of Transcatheter Aortic Valve Size on Leaflet Stresses: Implications for Durability and Optimal Grey Zone Sizing.

Authors:  Yue Xuan; Danny Dvir; Andrew D Wisneski; Zhongjie Wang; Jian Ye; Julius M Guccione; Liang Ge; Elaine E Tseng
Journal:  AsiaIntervention       Date:  2020-12

4.  Stent and leaflet stresses across generations of balloon-expandable transcatheter aortic valves.

Authors:  Yue Xuan; Danny Dvir; Zhongjie Wang; Jian Ye; Julius M Guccione; Liang Ge; Elaine E Tseng
Journal:  Interact Cardiovasc Thorac Surg       Date:  2020-06-01

5.  Aortic valve neocuspidization with in-body tissue-engineered autologous membranes: preliminary results in a long-term goat model.

Authors:  Takayuki Kawashima; Tadashi Umeno; Takeshi Terazawa; Tomoyuki Wada; Takashi Shuto; Haruto Nishida; Hirofumi Anai; Yasuhide Nakayama; Shinji Miyamoto
Journal:  Interact Cardiovasc Thorac Surg       Date:  2021-05-27
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.