BACKGROUND: Complete blood count (CBC) results play an important role in peripheral blood smear (PBS) examinations. Many descriptions in PBS reports may simply be translated from CBC parameters. We developed a computer program that automatically generates a PBS draft report based on CBC parameters and age- and sex-matched reference ranges.Entities:
Keywords: Complete blood count; Draft report generator; Peripheral blood smear
Mesh:
Year: 2018 PMID: 30027693 PMCID: PMC6056392 DOI: 10.3343/alm.2018.38.6.512
Source DB: PubMed Journal: Ann Lab Med ISSN: 2234-3806 Impact factor: 3.464
Hematology analyzers tested
| Model | Manufacturer | Data file type | Data file generation | Location |
|---|---|---|---|---|
| XE-5000 | Sysmex, Kobe, Japan | xlsx | Instrument | KUMC Guro Hospital |
| XE-5000 | Sysmex, Kobe, Japan | csv | Instrument | KUMC Ansan Hospital |
| XN-9000 | Sysmex, Kobe, Japan | xlsx | Instrument | KUMC Ansan Hospital |
| DxH800 | Beckman Coulter, Brea, CA, USA | xls | Middleware (HemLink) | KUMC Guro Hospital |
| ADVIA 2120i | Siemens Healthcare Diagnostics, Eschborn, Germany | xlsx | LIS | Inha University Hospital |
Abbreviations: csv, comma separated values; KUMC, Korea University Medical Center; LIS, laboratory information system.
Fig. 1Draft report: one sample is selected from the Sample List (left, shaded), and instantly, the Instrument Data panel is auto populated with complete blood count parameters. A draft report is prepared in the Draft Report panel. The Microscopic Findings panel remains in its initial state at this point. Age- and sex-matched reference ranges can be applied, if age and sex are manually selected (at the top of the screen). In this example, the Excel data file was generated from Sysmex XE-5000 (Sysmex, Kobe, Japan).
Fig. 2Final report: A manual differential count was performed, and the results are typed into the Microscopic Findings panel (red box). All changes made in the Microscopic Findings (in this case, differential counts and poikilocytosis) are automatically reflected in the Draft Report. Work-up (w/u) for Epstein-Barr virus (EBV) and cytomegalovirus (CMV) markers is recommended, owing to the increased reactive lymphocyte count (arrow). The grading scale for poikilocytosis follows the International Council for Standardization in Haematology guidelines (red circle, −, ++, and +++ for elliptocytes) [3]. In this example, the Excel data file was generated from Sysmex XE-5000 (Sysmex, Kobe, Japan).
Recommendations automatically generated according to morphological findings
| Finding | Recommendation |
|---|---|
| Schistocytes, spherocytes | Work-up for hemolysis (LDH, DAT, haptoglobin, plasma Hb, etc.). |
| Rouleaux formation | Serum and urine PEP/IFE, if clinically indicated. |
| Blast | CBC & PBS follow up, BM examination if clinically indicated. |
| Abnormal lymphocytes | CBC & PBS follow up, BM examination if clinically indicated. |
| Reactive lymphocytes ( > 10%) | Work-up for viral infection such as EBV and CMV. |
| Platelet clumping | Rule out EDTA-induced pseudothrombocytopenia. |
| Resampling using Heparin or Citrate tube is recommended. |
Abbreviations: LDH, lactose dehydrogenase; DAT, direct antiglobulin test; PEP, protein electrophoresis; CBC, complete blood count; PBS, peripheral blood smear; IFE, immunofixation electrophoresis; BM, bone marrow; EBV, Epstein-Barr virus; CMV, cytomegalovirus; EDTA, ethylenediaminetetraacetic acid.
Fig. 3In this example, the input data file contains patient information, such as sex and age, enabling automatic adjustment of sex and age options of the program (red box). In this case, the patient is one year old. The hemoglobin, mean corpuscular volume (MCV), and white blood cell (WBC) differential counts would have indicated microcytic anemia and relative lymphocytosis had adult reference ranges been applied. However, they are interpreted as normal, following the age-specific reference ranges. In this example, the Excel data file was generated from HemLink, a middleware for Beckman Coulter hematology analyzers.