Feasibility and challenges of using multiple breath washout in COPD.

Alan S Bell1,2, Philip J Lawrence1, Dave Singh1,2,3, Alexander Horsley2,3,4.   

Abstract

Background: Lung clearance index (LCI), derived from multiple-breath washout (MBW), is a well-established assessment of ventilation inhomogeneity in cystic fibrosis but has not been widely applied in other conditions characterized by heterogeneous airways disease, such as COPD. The aim of this study was to evaluate the sensitivity, repeatability, and practicality of LCI in patients with COPD.
Methods: Fifty-four COPD patients completed MBW using nitrogen as the washout tracer gas (MBWN2, measured using an Exhalyzer™ device), spirometry, and plethysmography. Twenty patients repeated MBWN2, MBWSF6 (using a separate Innocor™ gas analyzer to measure washout of the exogenous trace sulphur hexafluoride), and spirometry at a second visit ≥24 hours later.
Results: Mean (SD) COPD LCI measured by nitrogen washout (LCIN2) was 12.1 (2.2); mean (SD) LCI Z-score 5.8 (2.0). LCIN2 increased across Global Initiative for Obstructive Lung Disease stages 1 to 3 and was abnormal (Z-score >1.65) in all COPD patients, including those with forced expiratory volume in 1 second (FEV1) ≥80% predicted. LCI was repeatable (median intra-test coefficient of variation 4.1%) and reproducible (limits of agreement -1.8 to 1.6) after mean of 16 days. Functional residual capacity (FRC) measurements were significantly greater using nitrogen than SF6 or plethysmography: mean FRC measured by nitrogen washout (FRCN2) 139% predicted versus FRC measured by plethysmography 125% predicted, p<0.0001.
Conclusion: LCI is most suitable as a measure of early airways disease in COPD in those with well-preserved FEV1, with similar repeatability and limitations to that observed in cystic fibrosis. Using the Exhalyzer system to perform MBWN2, however, appeared to substantially over-read FRC. This discrepancy needs addressing before FRCN2 measurements made using this device can be reliably deployed.

Entities:  

Keywords:  COPD; functional residual capacity; lung clearance index; lung physiology; lung volumes; multiple breath washout

Mesh:

Substances:

Year:  2018        PMID: 30022817      PMCID: PMC6044337          DOI: 10.2147/COPD.S164285

Source DB:  PubMed          Journal:  Int J Chron Obstruct Pulmon Dis        ISSN: 1176-9106


Introduction

Multiple-breath washout (MBW) testing is now well established in cystic fibrosis (CF) research as a practical and reproducible measure of airways physiology that is both more sensitive at detecting the presence of disease and at monitoring treatment responses compared with conventional spirometry.1 MBW involves tidal breathing to wash an inert tracer gas from the lungs and generates measures of lung volume (functional residual capacity; FRC) as well as of gas mixing efficiency.2 The most important and commonly reported outcome is the lung clearance index (LCI), a summary measure of ventilation heterogeneity. It has been known for some time that COPD results in abnormal gas mixing,3 and also that ventilation heterogeneity is an early feature of smoking-related lung disease.4 Ventilation heterogeneity has been demonstrated in COPD using ventilation-MRI, and seems to relate to other measures of disease severity.5 However, the applicability of current MBW technologies for measuring ventilation heterogeneity in COPD has not been studied. The European CF Society have recently published recommendations for use of MBW in clinical trials in CF6 and there is a drive for clinical application of these measurements.7 MBW is now available in several CF centers in the US and Europe, the majority using nitrogen washout systems, which derive nitrogen concentration from measurements of expired O2 and CO2. Nitrogen washout is performed by breathing 100% oxygen. This differs from the majority of recent formative MBW studies where SF6 was used as the tracer gas, with washout by breathing room air. Different technologies seem to deliver different measures of LCI and FRC,8 and these are not interchangeable even between nitrogen washout systems.9 The role of MBW in COPD may differ from that in CF, being potentially more relevant in disease phenotyping and assessing treatment response than in detecting and preventing early disease progression. For MBW to be adopted in this way, it must be sensitive to disease identification, practical to perform,10 and reproducible.11 The applicability of MBW in adults with COPD has not previously been assessed. We hypothesized that MBW would be a practical, repeatable, and potentially informative test in adults with COPD. The aims of this study were, therefore, as follows: To assess practical limitations of MBW in patients diagnosed with COPD, including test time and success rates. To assess intra-visit repeatability and inter-visit reproducibility of LCI and FRC. To compare MBW outcomes with other lung function measures in patients diagnosed with COPD. To compare indices from nitrogen washout with those from SF6 washout.

Materials and methods

This study was performed at the Medicines Evaluation Unit, University Hospital of South Manchester, UK. COPD patients were aged >40 years and had a diagnosis of COPD in accordance with the 2013 Global Initiative for Obstructive Lung Disease (GOLD) recommendations.12 This included a smoking history ≥10 pack-years with typical symptoms (1 or more of productive cough, breathlessness, and wheeze) and evidence of airflow obstruction (forced expiratory volume in 1 second [FEV1]/forced vital capacity <0.7). Additional exclusion criteria included any respiratory tract infection or COPD exacerbations within 6 weeks of assessment. Patients diagnosed with COPD continued all regular pharmacological therapies but restricted short-acting inhaled bronchodilators for a minimum of 8 hours, and long-acting inhaled bronchodilators for a minimum of 24 hours prior to assessments. The study was approved by the Greater Manchester South research ethics committee (reference: 10/H1003/108) and written informed consent was obtained from each subject.

Study design

Fifty-four patients diagnosed with COPD performed impulse oscillometry, MBW, gas transfer, whole-body plethysmography, spirometry, and a 6-minute walk test (6MWT) of exercise capacity (in that order). Subjects were invited to return for repeat MBW testing >24 hours later.

Assessments

Detailed descriptions of each of the measurements can be found in the online supplement. Multiple-breath nitrogen washout (MBWN2) was performed as previously described8 using an open circuit, bias flow system (Exhalyzer D®, Eco-Medics AG, Duernten, Switzerland). Analysis was performed using the associated software (Spiroware 3.1 EcoMedics AG). MBWN2 was performed 3 times and reported values for FRC measured by nitrogen washout (FRCN2) and LCI measured by nitrogen washout (LCIN2) are the average of at least 2 reproducible repeats. Age-matched Z-scores for LCIN2 were taken from those provided by the manufacturers. A subset of patients performed additional MBW of SF6 (MBWSF6). This was carried out using an open circuit washin of 0.2% SF6 followed by washout breathing room air, using a modified Innocor analyzer as previously described.13 Both MBWN2 and MBWSF6 were performed according to consensus recommendations.14 Spirometry was performed according to the American Thoracic Society/European Respiratory Society recommendations15 using a Sensormedics Vmax spirometer (Sensormedics Corporation, Yorba Linda, CA, USA). Reference values used were those of the 2012 Global Lung Initiative.16 Impulse oscillometry was performed using a Masterscreen impulse oscillometer (IOS; Erich Jaeger, Hoechenberg, Germany) in accordance with consensus recommendations.17 Plethysmographic lung volumes were measured in accordance with the American Thoracic Society/European Respiratory Society recommendations16 using a constant volume whole-body plethysmograph (Autobox 6200 DL, Sensor-medics Corporation). Carbon monoxide diffusing capacity (DLCO) and carbon monoxide transfer coefficient (KCO) were measured using the Vmax 22 instrument (Sensormedics Corporation), according to ATS/ERS recommendations.18 The reported value is the mean of at least 2 successful readings within 10% of the highest value. The 6MWT test was conducted to a standardized protocol in accordance with current guidelines using a 20-meter course.19

Statistical analysis

A sample size of 10 patients would give 99% power to detect a 20% difference in LCI over historical control data and a sample size of 22 would be required to detect a 10% difference in LCI. We were interested, however, in how LCI compared with lung physiology across a range of disease severity and estimated that 50 patients would be sufficient to make reasonable inferences about how airway markers compared, and would give us enough subjects to get repeat LCI measurements on at least 20. Data were analyzed using Prism (GraphPad Software Inc., San Diego, CA, USA). Parametric data were expressed as mean (SD) and non-parametric data expressed as median with interquartile range (IQR). For comparison of physiology parameters with GOLD stage, a 1-way analysis of variance was conducted, followed by Tukey’s multiple comparison test to compare pairs of data. Correlations were assessed using a Pearson’s correlation coefficient or a Spearman’s rank correlation coefficient. Intra-visit repeatability was determined by calculating the coefficient of variation (CV) of the 3 recordings for each subject. Inter-visit reproducibility was assessed using Bland–Altman analysis20 and intra-class correlation coefficient (ICC). ICC was calculated using SPSS (version 22, IBM SPSS inc, Armonk, NY, USA) using a 2-way mixed analysis of absolute agreement. For the comparison of MBW methods, outcomes were compared using a paired t-test and the agreement between the 2 systems assessed using Bland–Altman analysis.20 Significance level was set at p=0.05.

Results

Fifty-four patients diagnosed with COPD (37 male) performed assessment of lung function and MBW. Mean (SD) age of subjects was 66 (7) years, with a median smoking history of 42 pack-years (full demographics presented in Table 1). Seventeen (31%) of the COPD patients were frequent exacerbators, defined as having ≥2 exacerbations in the last 12 months.
Table 1

Clinical characteristics and lung function measurements

Clinical characteristicsCOPD (n=54)
Absolute values% predicted
Sex (male/female)37/17
Age (years)66 (7)
Height (m)1.67 (0.10)
Weight (kg)78.8 (18.6)
Body mass index (kg/m2)27.8 (4.9)
Smoking history (pack years)a42.1 (31.7–55.5)
Spirometry
 FEV1 (L)1.68 (0.6)63.1 (18.4)
 FEF25%–75% (L/s)0.66 (0.35)22.0 (11.2)
 FVC (L)3.62 (1.0)106.6 (17.8)
 FEV1/FVC0.48 (0.13)
Multiple breath nitrogen washout
 MBWN2 test time (mins)38.4 (19.9)
 FRCN2 (L)4.35 (1.27)135.7 (36.8)
 LCIN212.1 (2.2)158.9 (29.3)
Plethysmography
 TLC (L)6.30 (1.27)105.5 (15.4)
 Inspiratory capacity (L)2.15 (0.59)78.4 (15.2)
 FRCpleth (L)3.99 (1.06)123.0 (29.8)
 RV (L)2.95 (0.97)127.9 (40.7)
 RV/TLC %46.6 (10.5)
Gas transfer
 DLCO (mmol⋅min−1⋅kPa−1)5.23 (1.78)63.1 (18.4)
 KCO (mmol⋅min−1⋅kPa−1)1.07 (0.28)79.6 (23.2)
Forced oscillation
 R5–R20 a (kPa/L/s)0.17 (0.11–0.25)
 X5 (kPa/L/s)−0.30 (0.16)
 CAT scorea18.0 (9.0–25.3)

Note: Parametric data are expressed as mean and SD except

median and inter-quartile range.

Abbreviations: CAT, COPD Assessment Test; DLco, carbon monoxide diffusing capacity; FEF, forced expiratory flow; FEV1, forced expiratory volume in 1 second; FRC, functional residual capacity; FRCN2, FRC measured by nitrogen washout; FRCpleth, FRC measured by plethysmography; FVC, forced vital capacity; KCO, carbon monoxide transfer coefficient; LCIN2, lung clearance index measured by nitrogen washout; MBWN2, multiple-breath washout using nitrogen; RV, residual volume; TLC, total lung capacity.

Practicality of MBWN2

Four out of the 54 patients diagnosed with COPD (7%) were unable to complete MBWN2 due to leak at the mouthpiece, dry throat, or a sensation of dyspnea when breathing high-flow O2. Nine repeat measurements were discarded for failure to meet quality control criteria. Patients with high minute ventilation would sometimes re-inspire expired air or room air from the end of the flow-past circuit. Overall, success rate was similar to that achieved for the other physiological assessments (Table S1). Median (IQR) total test time in COPD patients was 36.3 (23.8–44.9) minutes. Test time increased in those with higher LCI, r2=0.24, p=0.0006 (Figure S1).

Repeatability and reproducibility of MBW in COPD

The median (IQR) intra-test CV of triplicate repeat FRCN2 measurements on the same visit was 4.3 (1.7–6.3) for FRC and 4.1 (2.2–7.1) for LCI. Twenty clinically stable patients diagnosed with COPD performed repeat measurements of MBWN2 in triplicate after a mean (SD) interval of 16 (13) days. There was a mean difference in FRCN2 between visits of −0.4% predicted (absolute), 95% limits of agreement −28.9% to 28.1% (Figure S2). ICC for repeat FRC was 0.90 (95% CI: 0.76–0.96). For LCIN2, the mean difference between visits was −0.11 with limits of agreement −1.79 to 1.59, approximately equivalent to ±15% of initial LCI (Figure S3). ICC for LCI was 0.94 (95% CI: 0.86–0.98).

Comparison of LCI with other lung physiology assessments

LCIN2 was raised in all patients with COPD (Z-score >1.65), including 10 (20% of the COPD patients) with GOLD stage 1 (defined as FEV1 >80% predicted).21 LCIN2 showed significant inverse correlation with FEV1 Z-score (r2=0.36, p<0.0001; Figure 1) and forced expiratory flow (FEF)25%–75% (r2=0.36, p<0.0001). LCIN2 also showed significant positive correlations with R5–R20 (r2 =0.17, p=0.009), residual volume/total lung capacity (RV/TLC [%]) (r2=0.16, p=0.02), Raw (r2=0.30, p=0.0006), and significant negative correlations with X5 (r2=0.35, p=0.0006), sGaw (r2 =0.18, p=0.012) and DLCO (r2=0.34, p=0.0001). There was no correlation between LCIN2 and 6MWT distance. Full details of these correlations are presented in Table S2 and Figures S4–S11.
Figure 1

COPD LCIN2 Z-score versus FEV1 Z-score.

Notes: Z-scores were calculated using the GLI 2012 reference equations for FEV117 and those provided by Ecomedics for LCI. The horizontal dotted line represents the upper limit of normal for LCIN2. The vertical dotted line represents the FEV1 Z-score of −1.65 (lower limit of normal range).

Abbreviations: FEV1, forced expiratory volume in 1 second; GLI, global lung initiative; LCIN2, lung clearance index measured by nitrogen washout.

LCIN2 also increased significantly with greater GOLD stage. Mean (SD) LCIN2 in GOLD stage 1 was 10.2 (1.2), versus GOLD stage 2 (FEV1 50%–80%), 11.9 (1.8) p=0.048, or GOLD stage 3 (FEV1 30%–50%), 14.19 (2.3) p<0.0001 (Figure 2). There was only a single subject with GOLD stage 4 COPD, omitted from this analysis. Difference between lung physiology outcomes for the different GOLD stages (1–3) is presented in the OLS (Table S3). In addition to LCI, there was statistically significant separation between all 3 GOLD stages for sGaw and FEF25%–75%.
Figure 2

LCIN2 versus COPD GOLD stage.

Abbreviations: GOLD, Global Initiative for Obstructive Lung Disease; LCIN2, lung clearance index measured by nitrogen washout

Comparison of FRC assessments

Seventeen patients diagnosed with COPD completed separate sequential assessments of FRC and LCI using both N2 and SF6 as the washout gases (Table 2). There was no significant difference in LCI measured by sulphur hexafluoride washout (LCISF6) and LCIN2 (11.9 [3.1] versus 12.3 [2.2], p=0.4). FRCN2, however, was significantly elevated compared with FRC measured by sulphur hexafluoride washout (FRCSF6). Mean (SD) of FRCN2 was 136.8 (42.8) % predicted compared with 112.4 (28.0) % for FRCSF6, p=0.002. The difference between FRCN2% and FRCSF6% increased disproportionally as FRC% increased. The mean difference was 24.4 (26.4) % predicted with limits of agreement −27.3 to 76.1 % predicted (Figure S12).
Table 2

Clinical characteristics and lung function measurements of COPD patients with paired MBW assessments

Clinical parameterCOPD (n=17)p-value
FRC (L)
 FRCN24.24 (1.47)
 FRCSF63.49 (1.00)0.002
FRC (%) predicted
 FRCN2136.8 (42.8)0.002
 FRCSF6112.4 (28.0)
Lung clearance index
 LCIN212.3 (2.2)
 LCISF611.9 (3.1)0.4
FEV1(%)64.2 (21.1)
FVC (%)101.4 (13.6)
FEV1/FVC0.5 (0.1)

Note: Parametric data are expressed as mean and SD, and comparisons were performed using Student’s t-test.

Abbreviations: FEV1, forced expiratory volume in 1 second; FVC, forced vital capacity; FRC, functional residual capacity; FRCN2, FRC measured by nitrogen washout; FRCSF6, FRC measured by sulphur hexafluoride washout; LCIN2, lung clearance index measured by nitrogen washout; LCISF6, lung clearance index measured by sulphur hexafluoride washout; MBW, multiple-breath washout.

FRC values measured by plethysmography and nitrogen washout were also compared for the whole cohort at visit 1. Overall, FRCN2 was significantly greater than FRC measured by plethysmography (FRCpleth) (139.3% versus 124.5%, p<0.0001). Mean difference between FRCN2 and FRCpleth was 14.8% predicted (limits of agreement −17.3 to 46.8; Figure 3), with a disproportionate increase in FRCN2 in those with higher FRC % predicted.
Figure 3

Bland–Altman comparison between FRCN2 and FRCpleth in subjects with COPD.

Notes: The central dotted line represents the mean difference, and the upper and lower dotted lines represent the limits of agreement. The vertical line represents hyper-inflation, defined as FRCN2 >120% predicted. Data points plotted above the x-axis represent subjects in which FRCN2 was greater than FRCpleth.

Abbreviations: FRC, functional residual capacity; FRCN2, FRC measured by nitrogen washout; FRCpleth, FRC measured by plethysmography.

Discussion

This is the first study to compare different MBW techniques, alongside detailed physiological phenotyping, in patients diagnosed with COPD. The specific objectives were to understand more about the repeatability and practical limitations of MBW in COPD. This study found LCI to be sensitive to detect mild COPD, being elevated even in those in GOLD stage 1. Intra-visit repeatability was similar to that reported in CF studies8,13 and inter-visit reproducibility was similar to that of CF adults.22 Finally, LCI has physiologically valid correlations with other small airways measurements (including X5, R5–R20, RV/TLC, Raw) as well as other markers of COPD severity, such as DLCO. This accords with recent work showing that smokers with preserved FEV1 had evidence of small airways damage, gas trapping, and impaired gas transfer.23,24 Abnormality in single-breath nitrogen washout in symptomatic smokers and in MBW parameters in asymptomatic smokers (>10 pack years) has also been previously demonstrated.24,25 The particular advantages of LCI as an outcome for assessing these early changes is that the methodology is standardized,14 with good reproducibility, and with apparently good sensitivity. This last feature is harder to quantify, however, since there is no gold-standard definition of COPD that does not involve spirometry. It is nevertheless reassuring that LCI was abnormal (Z-score >1.65) in all the subjects with previously defined COPD and that it appeared to be more sensitive in this regard at detecting early airway change than the other lung physiology assessments. As with severe CF, however, the measurement appears less useful in those with more advanced COPD. In these patients, the test takes much longer to complete and the overall ventilation heterogeneity signal likely reflects a mixed pattern of disease. Thus, any signal from potentially reversible small airways disease will be overwhelmed by the signal from other lung regions where damage is more advanced and irreversible. These data and observations, therefore, support a role for MBW in measuring disease severity in patients with mild-to-moderate (GOLD stage 1–2) COPD, and possibly in detecting disease before it fulfills current spirometry-driven criteria. An important caveat, however, exists for the specific nitrogen washout system used in this study. An unanticipated finding was that the measures of FRCN2 were far greater not only than those of FRCSF6 but also FRCpleth. Plethysmography measures all gas within the thoracic cavity and should, therefore, always measure a greater value for “FRC” than that which is ventilated during the quiet tidal breathing of a MBW test. FRCpleth and FRCN2 are similar in healthy subjects,26 but increasing FRCpleth should be apparent in disease as gas trapping worsens, leading to an even greater discrepancy between tidally ventilated and total end-expiratory lung volumes.27 In this study, however, we have observed the opposite. Although both FRCpleth and FRCN2 were elevated in COPD, FRCN2 was higher than FRCpleth with an offset of almost 20%. This difference also worsened in those with more severe disease, implying that tidally ventilated lung volume becomes increasingly greater than total lung volume as disease progresses. Clearly, this conclusion cannot be correct. A recent study compared 2 similar indirect N2 devices (Exhalyzer D versus EasyOne Pro, NDD Medical Technologies, Zurich, Switzerland). A difference of 30% in FRC values was found between these systems, with the Exhalyzer D generating higher mean values,9 consistent with the observations in this study. This suggests an over-reading of FRC by the Exhalyzer system, which may be device-specific. Recent attempts to correct for excretion of body nitrogen did not fully account for this error,28,29 and are not included in this analysis. The limitations of this same MBWN2 system for measuring lung volumes have also recently been recognized in patients with CF,30 and seem to have occurred despite promising in vitro data for this method.7 Importantly, accurate FRC measurement is essential for all MBW outcome measurements, including LCI. Reassuringly, however, the SF6 washout gave similar values for LCI despite much lower values for FRC. This may be because both FRC and cumulative expired volume are overestimated to the same extent with the nitrogen washout, but importantly suggests no loss of sensitivity by using SF6. In this study, we have assessed patients diagnosed with COPD over a wide range of severity, including at repeat time points. The lack of a simple positive control for COPD, however, means that we cannot easily assess what the response to therapy should be. We also did not include a healthy control population, but have instead used the reference data from previous age-matched cohorts provided by the manufacturer to calculate Z-scores. LCI was assessed on site, using the on-board software of the Exhalyzer-D device. Measurements were not subject to external over-read as recently recommended in clinical trials,6 although the research team has considerable experience with MBW as a technique.

Conclusion

We have shown that LCI offers substantial promise as a measurement in COPD. The key practical features of MBW assessment in COPD are similar to those in CF, and LCI appears to work best as a measure of early airways disease in those with well-preserved FEV1. However, the specific nitrogen washout system used here exhibited major limitations in terms of the accuracy of the FRC measurements, something that has not been adequately accounted for and which has significant implications for clinical trials and trial networks already in progress. Considerable caution should, therefore, be employed in interpreting nitrogen washout outcomes until the reasons for this error are better understood.
  27 in total

1.  ATS statement: guidelines for the six-minute walk test.

Authors: 
Journal:  Am J Respir Crit Care Med       Date:  2002-07-01       Impact factor: 21.405

2.  Standardisation of the single-breath determination of carbon monoxide uptake in the lung.

Authors:  N Macintyre; R O Crapo; G Viegi; D C Johnson; C P M van der Grinten; V Brusasco; F Burgos; R Casaburi; A Coates; P Enright; P Gustafsson; J Hankinson; R Jensen; R McKay; M R Miller; D Navajas; O F Pedersen; R Pellegrino; J Wanger
Journal:  Eur Respir J       Date:  2005-10       Impact factor: 16.671

3.  Consensus statement for inert gas washout measurement using multiple- and single- breath tests.

Authors:  Paul D Robinson; Philipp Latzin; Sylvia Verbanck; Graham L Hall; Alexander Horsley; Monika Gappa; Cindy Thamrin; Hubertus G M Arets; Paul Aurora; Susanne I Fuchs; Gregory G King; Sooky Lum; Kenneth Macleod; Manuel Paiva; Jane J Pillow; Sarath Ranganathan; Sarah Ranganathan; Felix Ratjen; Florian Singer; Samatha Sonnappa; Janet Stocks; Padmaja Subbarao; Bruce R Thompson; Per M Gustafsson
Journal:  Eur Respir J       Date:  2013-02-08       Impact factor: 16.671

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

Review 5.  Lung clearance index: evidence for use in clinical trials in cystic fibrosis.

Authors:  L Kent; P Reix; J A Innes; S Zielen; M Le Bourgeois; C Braggion; S Lever; H G M Arets; K Brownlee; J M Bradley; K Bayfield; K O'Neill; D Savi; D Bilton; A Lindblad; J C Davies; I Sermet; K De Boeck
Journal:  J Cyst Fibros       Date:  2013-12-05       Impact factor: 5.482

6.  Emphysematous changes and normal variation in smokers and COPD patients using diffusion 3He MRI.

Authors:  Andrew J Swift; Jim M Wild; Stan Fichele; Neil Woodhouse; Sally Fleming; Judith Waterhouse; Rod A Lawson; Martyn N J Paley; Edwin J R Van Beek
Journal:  Eur J Radiol       Date:  2005-06       Impact factor: 3.528

7.  Feasibility and variability of measuring the Lung Clearance Index in a multi-center setting.

Authors:  Susanne I Fuchs; Helmut Ellemunter; Johannes Eder; Uwe Mellies; Jörg Grosse-Onnebrink; Burkhard Tümmler; Doris Staab; Andrea Jobst; Matthias Griese; Jan Ripper; Ernst Rietschel; Susanne Zeidler; Frank Ahrens; Monika Gappa
Journal:  Pediatr Pulmonol       Date:  2011-12-13

Review 8.  Putting lung function and physiology into perspective: cystic fibrosis in adults.

Authors:  Alex Horsley; Salman Siddiqui
Journal:  Respirology       Date:  2014-09-15       Impact factor: 6.424

9.  Evaluation of the impact of alveolar nitrogen excretion on indices derived from multiple breath nitrogen washout.

Authors:  Niklas Nielsen; Jorgen G Nielsen; Alex R Horsley
Journal:  PLoS One       Date:  2013-09-09       Impact factor: 3.240

10.  Correcting for tissue nitrogen excretion in multiple breath washout measurements.

Authors:  Mica Kane; Jonathan H Rayment; Renee Jensen; Reginald McDonald; Sanja Stanojevic; Felix Ratjen
Journal:  PLoS One       Date:  2017-10-11       Impact factor: 3.240

View more
  9 in total

1.  Effect of hypertonic saline on mucociliary clearance and clinical outcomes in chronic bronchitis.

Authors:  William D Bennett; Ashley G Henderson; Agathe Ceppe; Kirby L Zeman; Jihong Wu; Christine Gladman; Fred Fuller; Stephen Gazda; Brian Button; Richard C Boucher; Scott H Donaldson
Journal:  ERJ Open Res       Date:  2020-08-11

2.  Assessment of Ventilatory Heterogeneity in Chronic Obstructive Pulmonary Disease Using the Inspired Sinewave Test.

Authors:  Richard M Bruce; Phi Anh Phan; Marzia Rigolli; Minh C Tran; Edmund Pacpaco; Najib M Rahman; Andrew D Farmery
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2021-02-23

3.  A randomised controlled trial to investigate the use of high-frequency airway oscillations as training to improve dyspnoea in COPD.

Authors:  Enya Daynes; Neil Greening; Salman Sidiqqui; Sally Singh
Journal:  ERJ Open Res       Date:  2019-07-29

4.  Lung clearance index in detection of post-transplant bronchiolitis obliterans syndrome.

Authors:  Madeleine Driskel; Alex Horsley; Laurice Fretwell; Nigel Clayton; Mohamed Al-Aloul
Journal:  ERJ Open Res       Date:  2019-10-15

5.  Simultaneous sulfur hexafluoride and nitrogen multiple-breath washout (MBW) to examine inherent differences in MBW outcomes.

Authors:  Katie J Bayfield; Alex Horsley; Eric Alton; Samantha Irving; Andrew Bush; Jane C Davies
Journal:  ERJ Open Res       Date:  2019-11-04

6.  Effect of intermittent inspiratory leaks on measurement of lung clearance index using nitrogen and sulfur hexafluoride.

Authors:  Carl A Whitfield; Oliver E Jensen; Alex Horsley
Journal:  ERJ Open Res       Date:  2018-10-19

7.  The validity of shortened multiple-breath washout testing using sulfur hexafluoride in the assessment of patients with COPD.

Authors:  Enya Daynes; Neil Greening; John Owers-Bradley; Sally J Singh; Salman Siddiqui
Journal:  ERJ Open Res       Date:  2021-08-02

8.  Lung clearance index in healthy volunteers, measured using a novel portable system with a closed circuit wash-in.

Authors:  Alex R Horsley; Amnah Alrumuh; Brooke Bianco; Katie Bayfield; Joanne Tomlinson; Andrew Jones; Anirban Maitra; Steve Cunningham; Jaclyn Smith; Catherine Fullwood; Anand Pandyan; Francis J Gilchrist
Journal:  PLoS One       Date:  2020-02-25       Impact factor: 3.240

9.  Longitudinal assessment of lung clearance index to monitor disease progression in children and adults with cystic fibrosis.

Authors:  Alex R Horsley; John Belcher; Katie Bayfield; Brooke Bianco; Steve Cunningham; Catherine Fullwood; Andrew Jones; Anna Shawcross; Jaclyn A Smith; Anirban Maitra; Francis J Gilchrist
Journal:  Thorax       Date:  2021-07-22       Impact factor: 9.139

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.