Delia Waldenmaier1, Eva Zschornack1, Lucas Kalt2, Andreas Buhr2, Stefan Pleus1, Cornelia Haug1, Guido Freckmann1.
Abstract
INTRODUCTION: A new insulin pump system was tested under everyday conditions for user evaluation and safety assessment prior to its launch in Europe. This insulin pump is focused on easy handling and uses prefilled cartridges.Entities:
Keywords: CSII; insulin pump; user evaluation
Mesh:
Substances:
Year: 2018 PMID: 29998760 PMCID: PMC6313294 DOI: 10.1177/1932296818785386
Source DB: PubMed Journal: J Diabetes Sci Technol ISSN: 1932-2968
Subjects’ Currently Used Insulin Pump Systems.
| n (%) | |
|---|---|
| Insulin pump | |
| Medtronic Paradigm® Veo™ | 7 (20%) |
| Accu-Chek® Spirit Combo | 12 (34%) |
| Accu-Chek® Insight | 4 (11%) |
| Animas® Vibe® | 3 (8.6%) |
| mylife™ OmniPod® | 7 (20%) |
| MiniMed® 640G | 2 (5.7%) |
| Infusion set cannula | |
| Steel | 7 (20%) |
| Teflon | 21 (60%) |
| No infusion set | 7 (20%) |
| Insulin | |
| Humalog® | 20 (57%) |
| NovoRapid® | 12 (34%) |
| Other | 3 (8.6%) |
Figure 1.Subjects’ satisfaction with the new pump (after 30 days of use).
Figure 2.Subjects’ answers to questions about the insulin pump (A), functions (B), infusion set (C), and prefilled cartridge (D) (selected items). Due to rounding, minor differences between individual numbers and sums may occur.
Figure 3.Main reasons for the 77 unplanned infusion set replacements of 406 infusion set changes.
Occurrence of Alarms During Home Use Based on Data From Insulin Pump History.
| Alarm | Occurrence in all pumps | Per pump in 30 days |
|---|---|---|
| Total | 127 | 3.91 |
| Empty cartridge | 38 | 1.17 |
| Occlusion | 27 | 0.83 |
| Empty battery | 16 | 0.49 |
| Electronic alarm | 8 | 0.25 |
| Battery removed | 4 | 0.12 |
| Battery rejected | 3 | 0.09 |
| Auto stop | 2 | 0.06 |
Therapy Events During Home Use Based on Data from Insulin Pump History.
| Event | Occurrence in all subjects | Per subject per day |
|---|---|---|
| Total | 8997 | 9.24 |
| Immediate bolus | 5713 | 5.87 |
| Extended bolus | 63 | 0.06 |
| Combo bolus | 172 | 0.18 |
| Temporary basal rate | 310 | 0.32 |
| Switch of basal profile | 45 | 0.05 |
| Change of pump mode | 347 | 0.36 |
| Rewind | 376 | 0.39 |
| Priming | 621 | 0.64 |
| Others | 1350 | 1.39 |