Abstract
Sepsis and septic shock are major healthcare problems, resulting in high morbidity and mortality. The Surviving Sepsis Campaign (SSC), which standardised the approach to sepsis, was recently updated. Strategies to decrease the systemic inflammatory response have been proposed to modulate organ dysfunctions. Endotoxin, derived from the membrane of Gram-negative bacteria, is considered a major factor in the pathogenesis of sepsis. Endotoxin adsorption, if effective, has the potential to reduce the biological cascade of Gram-negative sepsis. We present a case of a 64-year-old man with severe Gram-negative sepsis, following purulent peritonitis secondary to rectosigmoid adenocarcinoma. To reduce the amplitude of the general effects of endotoxins we used a novel device, the Alteco® LPS Adsorber (Alteco Medical AB, Lund, Sweden), for lipopolysaccharide (LPS) adsorption. The efficacy markers were: the overall haemodynamic profile, translated into decreased vasopressor requirements, the normalisation of the cardiac index, the systemic vascular resistance index combined with the lactate level and the reduction in procalcitonin (PCT) levels. A decrease in the sequential organ failure assessment (SOFA) score at twenty-four hours was demonstrated. The clinical course following treatment was favourable for the days immediately following the treatment.This was attributed to the removal of endotoxin from the systemic circulation. The patient died one week after the endotoxin removal session, developing an ischemic bowel perforation with subsequent multiple organ failures.Entities:
Keywords: Alteco® LPS Adsorber; Gram-negative abdominal sepsis; Haemadsorption; endotoxin; haemodynamics
Year: 2017 PMID: 29967875 PMCID: PMC5769914 DOI: 10.1515/jccm-2017-0010
Source DB: PubMed Journal: J Crit Care Med (Targu Mures) ISSN: 2393-1817
Fig. 4LPS Adsorber was connected to an extracorporeal circuit including an HF 440 blood pump. (Infomed, Geneva, Switzerland)
Fig. 1Changes in the measured haemodynamic and biochemical data before and after treatment with the LPS Adsorber
Fig. 2Changes in the PCT values before and after treatment with the LPS Adsorber
Fig. 3Organ dysfunction indicated by the SOFA score before and after endotoxin elimination therapy