Effect of high-intensity curing lights on the polymerization of bulk-fill composites.

Mindy M Daugherty1, Wen Lien2, Michael R Mansell3, Douglas L Risk4, Daniel A Savett5, Kraig S Vandewalle6.   

Abstract

OBJECTIVES: The purpose of this study was to investigate the effect of high-irradiance light-curing-units (LCUs) on the depth-of-cure (DoC) and degree-of-polymerization (DoP) of bulk-fill composites (BFCs).
METHODS: The DoC of composites (Beautifil-Bulk, SHOFU; Filtek-Bulk-Fill, 3M ESPE; Tetric-EvoCeram-Bulk-Fill, Ivoclar; Sonic-Fill-2, Kerr; Venus-Bulk-Fill, Heraeus; Z250, 3M-ESPE) were measured according to ISO-4049 using high-irradiance LCUs (FlashMax-P3, CMS-Dental; SPEC3, Coltene) and conventional LCU (Paradigm, 3M-ESPE) for exposure times: 3/9-s, 3/20-s, and 10/20-s respectively. Using FTIR, the DoP per composite was measured at the bottom surface as a function of post-curing times for the LCUs at the same exposure times. Data was analyzed with nonlinear regression and ANOVA/Tukey.
RESULTS: Significant differences in DoC were found amongst the LCUs for the various exposure times. All BFCs failed to meet the DoC claimed by manufacturers and failed to satisfy ISO-4049 with the high-irradiance LCUs with 3-s exposures. Standard irradiance and 20-s exposures outperformed all other irradiance-exposure combinations for maximizing the DoC and DoP of BFCs. A minimum of 15.3J/cm2 radiant exposure was required to achieve an adequate maximum polymerization rate. Venus Bulk exhibited the highest DoC and DoP for any LCU-exposure-time combination. SIGNIFICANCE: Among the different combinations of BFCs and LCUs, DoC and DoP were always increased with longer exposure time, but there exists a theoretical radiant-exposure limit beyond which DoP or DoC remains unchanged. However, high DoC or DoP are not always associated with one another. Thus, the exposure-reciprocity law must be approached thoughtfully since irradiance and exposure can independently affect DoP and DoC.
Copyright © 2018 The Academy of Dental Materials. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Bulk-fill composites; Degree of polymerization; Depth of cure; High-intensity curing lights

Mesh:

Substances:

Year:  2018        PMID: 29960652     DOI: 10.1016/j.dental.2018.06.005

Source DB:  PubMed          Journal:  Dent Mater        ISSN: 0109-5641            Impact factor:   5.304


  10 in total

Review 1.  [Factors influencing clinical application of bulk-fill composite resin].

Authors:  Jing Xue
Journal:  Hua Xi Kou Qiang Yi Xue Za Zhi       Date:  2020-06-01

Review 2.  The influence of inorganic fillers on the light transmission through resin-matrix composites during the light-curing procedure: an integrative review.

Authors:  Rita Fidalgo-Pereira; Daniela Carpio; Orlanda Torres; Oscar Carvalho; Filipe Silva; Bruno Henriques; Mutlu Özcan; Júlio C M Souza
Journal:  Clin Oral Investig       Date:  2022-06-29       Impact factor: 3.606

3.  The Effect of Two Different Light-Curing Units and Curing Times on Bulk-Fill Restorative Materials.

Authors:  Gokcen Deniz Bayrak; Elif Yaman-Dosdogru; Senem Selvi-Kuvvetli
Journal:  Polymers (Basel)       Date:  2022-05-05       Impact factor: 4.967

4.  Effects of Different Light-curing Modes on the Compressive Strengths of Nanohybrid Resin-based Composites: A Comparative In Vitro Study.

Authors:  Franco Mauricio; Julia Medina; Luzmila Vilchez; Oscar Sotomayor; César Muricio-Vilchez; Frank Mayta-Tovalino
Journal:  J Int Soc Prev Community Dent       Date:  2021-04-15

5.  The effect of rapid high-intensity light-curing on micromechanical properties of bulk-fill and conventional resin composites.

Authors:  Matej Par; Danijela Marovic; Thomas Attin; Zrinka Tarle; Tobias T Tauböck
Journal:  Sci Rep       Date:  2020-06-29       Impact factor: 4.379

6.  Changes in color and contrast ratio of resin composites after curing and storage in water.

Authors:  Marlus da Silva Pedrosa; Fernando Neves Nogueira; Vitor de Oliveira Baldo; Igor Studart Medeiros
Journal:  Saudi Dent J       Date:  2021-02-20

7.  Effect of Light Curing Distance on Microhardness Profiles of Bulk-Fill Resin Composites.

Authors:  Fatin A Hasanain; Hani M Nassar; Reem A Ajaj
Journal:  Polymers (Basel)       Date:  2022-01-28       Impact factor: 4.329

8.  Evaluation of Microhardness of Two Bulk-fill Composite Resins Compared to a Conventional Composite Resin on surface and in Different Depths.

Authors:  Keyvan Saati; Sheida Khansari; Farnaz Mahdisiar; Sara Valizadeh
Journal:  J Dent (Shiraz)       Date:  2022-03

Review 9.  High-Power LED Units Currently Available for Dental Resin-Based Materials-A Review.

Authors:  Rita Almeida; Patricia Manarte-Monteiro; Joana Domingues; Carlos Falcão; Mariano Herrero-Climent; Blanca Ríos-Carrasco; Bernardo Ferreira Lemos
Journal:  Polymers (Basel)       Date:  2021-06-30       Impact factor: 4.329

10.  Infection Control Barrier and Curing Time as Factors Affecting the Irradiance of Light-Cure Units.

Authors:  Reem A Ajaj; Hani M Nassar; Fatin A Hasanain
Journal:  J Int Soc Prev Community Dent       Date:  2018-11-29
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.