Literature DB >> 29958755

Risk Factors Associated With Early Complications of Revision Surgery for Head-Neck Taper Corrosion in Metal-on-Polyethylene Total Hip Arthroplasty.

Young-Min Kwon1, Daniel Rossi1, John MacAuliffe1, Yun Peng1, Paul Arauz1.   

Abstract

BACKGROUND: Adverse local tissue reactions (ALTRs) due to tribocorrosion of head-neck taper junctions in contemporary metal-on-polyethylene (MoP) total hip arthroplasty (THA) are emerging as an important reason for failure requiring revision surgery. This study aimed at: (1) reporting early complication rates and outcome, and (2) identifying risk factors associated with complications of revision surgery for head-neck taper corrosion in patients with MoP THA.
METHODS: Forty consecutive revision surgeries in 39 patients (male, 16; female, 23) with MoP THA were evaluated. The follow-up period after revision was a minimum of 14 months (range, 14-45). The indication for revision surgery was the presence of symptomatic ALTR on magnetic resonance imaging (MRI) with elevated metal ion levels.
RESULTS: The overall complication rate was 25% and the reoperation rate was 10%. The median serum level of cobalt ions decreased significantly 8.2 μg/L (1.2-56.1 μg/L) pre-revision to 3.1 μg/L (0.2-14.0 μg/L) post-revision (P < .01). High fretting and corrosion grades (Goldberg score ≥ 3) were observed in 82% of retrieved implants. MRI findings of solid lesion(s) with abductor deficiency (P < .01) and intraoperative tissue damage at revision (P = .02) were significantly associated with post-revision complications.
CONCLUSION: A high rate of early complications (25%) and re-revisions (10%) was observed after revision of ALTR associated with head-neck taper corrosion. Pre-revision MRI findings of solid lesion(s) with abductor deficiency and intraoperative tissue damage were risk factors associated with the occurrence of a complication after revision surgery. This information provides clinically useful information for clinical decision-making and preoperative counseling of MoP THA patients undergoing revision surgery for head-neck taper corrosion.
Copyright © 2018 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  adverse local tissue reaction; complications; head-neck taper corrosion; metal-on-polyethylene total hip arthroplasty; revision surgery; risk factors

Mesh:

Substances:

Year:  2018        PMID: 29958755     DOI: 10.1016/j.arth.2018.05.046

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  7 in total

Review 1.  What Surgeons Need to Know About Adverse Local Tissue Reaction in Total Hip Arthroplasty.

Authors:  Deborah J Hall; Robin Pourzal; Joshua J Jacobs
Journal:  J Arthroplasty       Date:  2020-01-15       Impact factor: 4.757

2.  Pseudotumor recurrence in a post-revision total hip arthroplasty with stem neck modularity: A case report.

Authors:  Bhumit R Desai; Gonzalo E Sumarriva; George F Chimento
Journal:  World J Orthop       Date:  2020-02-18

3.  Outcome of revision surgery for adverse local tissue reactions in patients with recalled total hip arthroplasty.

Authors:  Christian Klemt; Sakkadech Limmahakhun; Georges Bounajem; Christopher M Melnic; Michael J Harvey; Young-Min Kwon
Journal:  Arch Orthop Trauma Surg       Date:  2021-04-09       Impact factor: 2.928

4.  Standardizing terms for tribocorrosion-associated adverse local tissue reaction in total hip arthroplasty.

Authors:  Brian J McGrory; Joshua J Jacobs; Young-Min Kwon; Yale Fillingham
Journal:  Arthroplast Today       Date:  2020-03-03

Review 5.  The Diagnosis and Treatment of Acetabular Bone Loss in Revision Hip Arthroplasty: An International Consensus Symposium.

Authors:  Peter K Sculco; Timothy Wright; Michael-Alexander Malahias; Alexander Gu; Mathias Bostrom; Fares Haddad; Seth Jerabek; Michael Bolognesi; Thomas Fehring; Alejandro Gonzalez DellaValle; William Jiranek; William Walter; Wayne Paprosky; Donald Garbuz; Thomas Sculco
Journal:  HSS J       Date:  2021-09-28

6.  Effect of head size and rotation on taper corrosion in a hip simulator.

Authors:  Christian M Wight; Cari M Whyne; Earl R Bogoch; Radovan Zdero; Ryan M Chapman; Douglas W van Citters; William R Walsh; Emil Schemitsch
Journal:  Bone Jt Open       Date:  2021-11

7.  Synovial Fluid Metal Ion Levels are Superior to Blood Metal Ion Levels in Predicting an Adverse Local Tissue Reaction in Failed Total Hip Arthroplasty.

Authors:  Matthew T Houdek; Michael J Taunton; Cody C Wyles; Paul J Jannetto; David G Lewallen; Daniel J Berry
Journal:  J Arthroplasty       Date:  2021-05-01       Impact factor: 4.435

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.