Multiple Breath-hold Volumetric Modulated Arc Therapy Under Fluoroscopic Image Guidance with an Implanted Fiducial Marker: An Advanced Technique.

Tsuyoshi Takanaka1, Yoko Taima2, Yasushi Horichi3, Yasuhiro Kawamori3, Koji Nobata3, Kiyohide Kitagawa3, Azusa Norishima4, Kento Koshikawa4, Tatsuya Mito4, Satoshi Yoshida4, Tetsuya Kawanaka4, Katsuaki Matsutani4, Masahiro Kawahara4.   

Abstract

An advanced technique for multiple breath-hold volumetric modulated arc therapy (VMAT) has been proposed under fluoroscopic image guidance with a fiducial marker implanted close to a tumor. The marker coordinates on a digitally reconstructed radiography image at a gantry start angle, under a planned breath-hold condition, were transferred to the fluoroscopic image window. Then, a reference lateral line passing through the planned breath-hold marker position was drawn on the fluoroscopic image. Additional lateral lines were further added on both sides of the reference line with a distance of 3 mm as a tolerance limit for the breath-hold beam delivery. Subsequently, the patient was asked to breathe in slowly under fluoroscopy. Immediately after the marker position on the fluoroscopic image moved inside the tolerance range, the patient was asked to hold the breath and the VMAT beam was delivered. During the beam delivery, the breath-hold status was continuously monitored by checking if the deviation of the marker position exceeded the tolerance limit. As long as the marker stayed within the tolerance range, a segmented VMAT delivery continued for a preset period of 15 to 30 seconds depending on the breath-hold capability of each patient. As soon as each segmented delivery was completed, the beam interrupt button was pushed; subsequently, the patient was asked for free breathing. This procedure was repeated until all the segmented VMAT beams were delivered. A lung tumor case is reported here as an initial study. The proposed technique may be clinically advantageous for treating respiratory moving tumors including lung tumor, liver cancer, and other abdominal cancers.

Entities:  

Keywords:  breath-hold; fiducial markers; fluoroscopy; respiratory motion management; vmat

Year:  2018        PMID: 29951345      PMCID: PMC6019328          DOI: 10.7759/cureus.2499

Source DB:  PubMed          Journal:  Cureus        ISSN: 2168-8184


  3 in total

1.  The use of active breathing control (ABC) to reduce margin for breathing motion.

Authors:  J W Wong; M B Sharpe; D A Jaffray; V R Kini; J M Robertson; J S Stromberg; A A Martinez
Journal:  Int J Radiat Oncol Biol Phys       Date:  1999-07-01       Impact factor: 7.038

2.  Reproducibility of liver position using active breathing coordinator for liver cancer radiotherapy.

Authors:  Cynthia Eccles; Kristy K Brock; Jean-Pierre Bissonnette; Maria Hawkins; Laura A Dawson
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-03-01       Impact factor: 7.038

3.  Reproducibility of diaphragm position assessed with a voluntary breath-holding device.

Authors:  Shigeyuki Takamatsu; Tsuyoshi Takanaka; Tomoyasu Kumano; Eiichi Mizuno; Satoshi Shibata; Shizuko Ohashi; Yuichi Kurata; Shinichi Ueda; Naoki Hori; Saori Shouji; Kimiya Noto; Hironori Kojima; Osamu Matsui
Journal:  Jpn J Radiol       Date:  2013-03-23       Impact factor: 2.374

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.