| Literature DB >> 29951141 |
Shizuka Tada1, Masayoshi Shibata1, Seiko Ohno1, Yasunobu Haruki1, Hironori Murakami2, Daisuke Hotta2, Masanori Nojima3, Gregory W Ruhnke4.
Abstract
BACKGROUND: The optimal implantation site of a new implantable cardiac monitor (ICM) named Reveal LINQ® may be limited based on a sufficient amplitude of R wave potential (AEP) acquisition because it is the same anatomic area used for transthoracic echocardiography (TTE).Entities:
Keywords: devices; female; human; implantable loop recorder; transthoracic echocardiography
Year: 2018 PMID: 29951141 PMCID: PMC6009771 DOI: 10.1002/joa3.12037
Source DB: PubMed Journal: J Arrhythm ISSN: 1880-4276
Figure 1Measurements of the amplitude of electrical potential as measured by the R wave. Two small electrodes were attached parasternally in the 4th intercostal space (ICS) in positions similar to those of ECG precordial leads V2 and V3 2 cm lateral of the left sternal border (A). They were then moved to the 5th ICS, again placed in a parasternal position (B). Finally, they were attached superiorly in a sagittal plane relative to the left sternal border at the 4th ICS in the same parasternal position (C)
Figure 2The specific measure indicating the actual electrode sites of the Reveal LINQ®
Figure 3Measurements of the amplitude of electrical potential based on the R wave ascertained after connection of the electrodes with a pacemaker programmer 2090W (Medtronic)
Figure 4The figure shows that an assessment method of grading interference with transthoracic echocardiography performance upon placement of an imitation Reveal LINQ® in the 4th intercostal space
Age, gender, height, weight, and BMI of all enrolled subjects
| No. | Age (years) | Gender | Height (cm) | Weight (kg) | BMI |
|---|---|---|---|---|---|
| 1. | 23 | M | 168.0 | 54.9 | 19.5 |
| 2. | 27 | M | 179.6 | 75.0 | 23.3 |
| 3. | 29 | F | 175.0 | 64.0 | 20.9 |
| 4. | 31 | M | 169.9 | 51.0 | 17.7 |
| 5. | 35 | M | 174.9 | 78.0 | 25.5 |
| 6. | 37 | F | 161.7 | 60.0 | 23.0 |
| 7. | 41 | M | 170.0 | 81.0 | 28.0 |
| 8. | 43 | F | 169.1 | 83.6 | 29.2 |
| 9. | 48 | F | 173.0 | 65.0 | 21.7 |
| 10. | 51 | M | 156.5 | 58.7 | 24.0 |
| 11. | 51 | M | 155.6 | 60.7 | 25.1 |
| 12. | 52 | F | 163.0 | 55.0 | 20.7 |
| 13. | 52 | F | 160.0 | 56.6 | 22.1 |
| 14. | 58 | F | 158.8 | 57.3 | 22.7 |
| 15. | 59 | M | 157.5 | 49.0 | 19.8 |
| 16. | 62 | M | 151.0 | 44.2 | 19.4 |
| 17. | 67 | F | 155.0 | 54.5 | 22.7 |
| 18. | 71 | M | 156.2 | 53.0 | 21.7 |
| Average | 45.3 ± 13.6 | 164.2 ± 8.4 | 61.2 ± 11.3 | 22.6 ± 3.0 |
The amplitude of electrical potential of the R wave with the 3 site‐setting combinations in each position among all enrolled subjects
| Position | A/a | B/a | A/b |
|---|---|---|---|
| Supine | 0.84 ± 0.28 | 0.73 ± 0.36 | 0.33 ± 0.19 |
| Right lateral decubitus | 0.63 ± 0.32 | 0.70 ± 0.39 | 0.23 ± 0.18 |
| Left lateral decubitus | 0.64 ± 0.27 | 0.64 ± 0.37 | 0.22 ± 0.21 |
| Sitting | 0.55 ± 0.31 | 0.52 ± 0.36 | 0.19 ± 0.17 |
| Standing | 0.57 ± 0.31 | 0.56 ± 0.40 | 0.18 ± 0.20 |
P < .001; this indicates a significant difference between A/a and A/b in each position.
P < .001; this indicates a significant difference between B/a and A/b in each position.
P < .001; this indicates a significant difference in the supine position relative to all other positions. Values are expressed as mean ± standard deviation.
Grade of interference with transthoracic echocardiography (TTE) performance after placement of the imitation ICM of the Reveal LINQ® in setting a either at sites A or B
| No. | Site A | Site B |
|---|---|---|
| 1. | 1 | 0 |
| 2. | 1 | 0 |
| 3. | 1 | 0 |
| 4. | 1 | 0 |
| 5. | 2 | 0 |
| 6. | 1 | 0 |
| 7. | 0 | 0 |
| 8. | 2 | 0 |
| 9. | 1 | 0 |
| 10. | 0 | 0 |
| 11. | 1 | 0 |
| 12. | 2 | 2 |
| 13. | 1 | 0 |
| 14. | 0 | 0 |
| 15. | 0 | 0 |
| 16. | 2 | 2 |
| 17. | 2 | 2 |
| 18. | 2 | 0 |
| Average | 1.1 | 0.3 |
Grade 2, a TTE transducer overlays ≥1/2 of the imitation ICM; grade 1, a TTE transducer overlays <1/2 of the imitation ICM; grade 0, a TTE transducer does not overlay the imitation ICM at all.
P = .0006 vs. site A.
End‐diastolic dimension, end‐systolic dimension and spherical index of the left ventricle, and left atrial dimension obtained by transthoracic echocardiography in setting A either at sites A or B before and after implantation of the imitation Reveal LINQ®
| Pre | Site A | Site B | |
|---|---|---|---|
| Dd (mm) | 45.9 ± 3.2 | 45.4 ± 3.2 | 45.5 ± 3.0 |
| Ds (mm) | 27.7 ± 3.6 | 27.9 ± 3.7 | 28.0 ± 3.8 |
| SI (%) | 105.8 ± 9.7 | 106.6 ± 10.8 | 105.1 ± 9.8 |
| LAD (mm) | 31.6 ± 4.3 | 31.7 ± 3.8 | 31.0 ± 3.3 |
Dd, end‐diastolic dimension of the left ventricle; Ds, end‐systolic dimension of the left ventricle; SI, spherical index of the left ventricle; LAD, left atrial dimension.
Values are expressed as mean ± standard deviation.
P = .028 vs. Dd prior to placement of imitation ICM.