Seth Kenkel1,2, Anirudh Mittal1,3, Shachi Mittal1,3, Rohit Bhargava1,2,3,4. 1. Beckman Institute for Advanced Science and Technology , University of Illinois at Urbana-Champaign , Urbana , Illinois 61801 , United States. 2. Department of Mechanical Engineering , University of Illinois at Urbana-Champaign , Urbana , Illinois 61801 , United States. 3. Department of Bioengineering , University of Illinois at Urbana-Champaign , Urbana , Illinois 61801 , United States. 4. Department of Chemical and Biomolecular Engineering, Department of Electrical and Computer Engineering, and Department of Chemistry , University of Illinois at Urbana-Champaign , Urbana , Illinois 61801 , United States.
Abstract
Nanoscale topological imaging using atomic force microscopy (AFM) combined with infrared (IR) spectroscopy (AFM-IR) is a rapidly emerging modality to record correlated structural and chemical images. Although the expectation is that the spectral data faithfully represents the underlying chemical composition, the sample mechanical properties affect the recorded data (known as the probe-sample-interaction effect). Although experts in the field are aware of this effect, the contribution is not fully understood. Further, when the sample properties are not well-known or when AFM-IR experiments are conducted by nonexperts, there is a chance that these nonmolecular properties may affect analytical measurements in an uncertain manner. Techniques such as resonance-enhanced imaging and normalization of the IR signal using ratios might improve fidelity of recorded data, but they are not universally effective. Here, we provide a fully analytical model that relates cantilever response to the local sample expansion which opens several avenues. We demonstrate a new method for removing probe-sample-interaction effects in AFM-IR images by measuring the cantilever responsivity using a mechanically induced, out-of-plane sample vibration. This method is then applied to model polymers and mammary epithelial cells to show improvements in sensitivity, accuracy, and repeatability for measuring soft matter when compared to the current state of the art (resonance-enhanced operation). Understanding of the sample-dependent cantilever responsivity is an essential addition to AFM-IR imaging if the identification of chemical features at nanoscale resolutions is to be realized for arbitrary samples.
Nanoscale topological imaging using atomic force microscopy (AFM) combined with infrared (IR) spectroscopy (AFM-IR) is a rapidly emerging modality to record correlated structural and chemical images. Although the expectation is that the spectral data faithfully represents the underlying chemical composition, the sample mechanical properties affect the recorded data (known as the probe-sample-interaction effect). Although experts in the field are aware of this effect, the contribution is not fully understood. Further, when the sample properties are not well-known or when AFM-IR experiments are conducted by nonexperts, there is a chance that these nonmolecular properties may affect analytical measurements in an uncertain manner. Techniques such as resonance-enhanced imaging and normalization of the IR signal using ratios might improve fidelity of recorded data, but they are not universally effective. Here, we provide a fully analytical model that relates cantilever response to the local sample expansion which opens several avenues. We demonstrate a new method for removing probe-sample-interaction effects in AFM-IR images by measuring the cantilever responsivity using a mechanically induced, out-of-plane sample vibration. This method is then applied to model polymers and mammary epithelial cells to show improvements in sensitivity, accuracy, and repeatability for measuring soft matter when compared to the current state of the art (resonance-enhanced operation). Understanding of the sample-dependent cantilever responsivity is an essential addition to AFM-IR imaging if the identification of chemical features at nanoscale resolutions is to be realized for arbitrary samples.
Atomic force microscopy (AFM)
techniques, including photoinduced-force microscopy (PiFM), peak-force
infrared microscopy (PFIR), and photothermal-induced resonance (PTIR),
have been widely used to detect optical spectroscopic data from absorbing
samples.[1−4] Each technique provides a measure of the local sample absorbance,
but other properties that might also contribute to image contrast
are not fully understood. In particular, AFM-IR is an imaging modality
that uses AFM to measure the PTIR signal produced by a pulsed IR laser[5,6] with theorized resolutions significantly below the diffraction limits
of far-field IR microscopy.[7] In response
to an IR laser with a slow repetition rate (∼1 kHz), the approach
records data by exciting cantilever oscillation at resonant modes
to produce a ringdown signal with an amplitude proportional to the
local sample absorbance.[4,8−12] Newer adaptations of this technique operate at higher frequencies
and incorporate lock-in detection of the cantilever deflection signal,
demonstrating improvements in the signal-to-noise ratio (SNR) and
data acquisition speed. The AFM-IR technique has been shown to closely
resemble far-field FTIR transmission spectra.[13,14] At present, however, this imaging modality suffers from signal fluctuations
resulting from probe–sample mechanical interactions.[15] These fluctuations can have little or no correlation
to the local sample expansion (or spectral contrast). It has been
shown that these fluctuations can be mitigated by tracking a cantilever-resonance
peak[16] during data acquisition (hereafter
referred to as resonance-enhanced operation) or by using IR-peak ratios[15] for analysis postacquisition. These methods,
however, restrict the available data and are not always effective.
Improved optomechanical probes can be designed to be less sensitive
to mechanical-property variations.[17] To
date, however, imaging sample expansion free of probe–sample
mechanical interactions has not been demonstrated.Resonance-enhanced
AFM-IR outperforms scattering-based techniques
with greatly improved detection sensitivity[18−21] and has been successfully demonstrated
for thin, weakly absorbing samples across many fields of study.[8,22−25] However, for thick samples at wavelengths corresponding to mid-IR
fundamental modes (best for molecular-spectral analysis), absorption
is strong and results in a large sample expansion. The sensitivity
improvement on resonance is not necessarily realized in these cases,
as the laser intensity needs to be reduced (sometimes less than 1%
of the full power) to avoid signal saturation or sample melting.[26] Moreover, both the amplitude and frequency of
resonance peaks are functions of the local mechanical properties of
the sample.[16] This results in an undesirable
outcome in some cases where the variation in the resonance amplitude
becomes dominant, especially for high-frequency resonance modes. As
a result, resonance tracking is typically restricted to the low-frequency-cantilever-resonance
modes, which have higher levels of noise. Thus, this current-state-of-the-art
approach can result in lower sensitivity from the lower-illumination
signal and higher noise from operating at lower-resonance modes. The
performance ceiling is seemingly limited without an alternate approach.
Here, we propose that an explicit analytical understanding of the
fundamental imaging process and its dependence on experimental parameters
can prevent artifacts in PTIR-signal acquisition and raise the limits
of sensitivity and accuracy of AFM-IR imaging. In this report, we
first describe the AFM-IR-image-formation process theoretically and
then use the insight obtained to develop techniques for improving
the accuracy and repeatability of AFM-IR imaging.
Experimental
Section
Instrumentation Design and Implementation
The quantum-cascade
laser (QCL) and piezo signals were generated using two trigger outputs
from a commercially available Nano-IR2 from Anasys Instruments Corporation
with a standard Anasys contact-mode probe (PN PR-EX-nIR2-10). The
first trigger output was a 100 μs Transistor-Transistor Logic
(TTL) pulse that occurred at the start of every trace and retrace
scan. Using a National Instruments Data Acquisition (DAQ) device (USB-6009)
and lab view, the falling edge of this trigger signal was used to
generate two TTL output signals. These output signals switched between
high and low voltages at the start of every alternate trace scan,
so if one signal was high during the scan, the other signal was low.
The second trigger output from the instrument was a TTL pulse train
with the repetition rate and pulse width set in the analysis-studio
software from Anasys. The two output signals from the DAQ and this
trigger signal were fed to a logic circuit to create two TTL pulse-train
signals that switched on and off at alternating trace scans. These
two signals were fed to the QCL and piezo trigger inputs, respectively,
resulting in the desired, interlaced image.
Data Collection and Processing
Measured transfer-function
curves were collected using a commercial Nano-IR2 instrument from
Anasys Instruments Corporation. The curves were measured by pulsing
the QCL laser at a 1 kHz repetition rate with a 300 ns pulse width,
averaging the 2048 time-series ringdown profiles, multiplying the
time-series ringdown data with a triangle curve, and then applying
a Fourier transform. The ringdown measurement was repeated up to 1000
times and averaged in the time domain to further reduce noise for
some of the curves shown.For the equipment used here, the frequency
range used for curve fitting was 250 kHz to 2 MHz. Curve fitting was
conducted using lsqcurvefit in Matlab with the equations first defined
symbolically and then converted to functions using matlabFunction().
We fit an array of n parameters, x(1:n), which had the functional form {m, Γ, ...} = {e, e, ...}, in relation to the unknown parameters.
This was done to constrain the parameters {m, Γ,
...} to be positive. The desired parameters {m, Γ,
...} and 95% confidence intervals were then determined from the array
of fit parameters, x(1:n). The 95%
confidence intervals were computed using the outputs from lsqcurvefit
as inputs for nlparci functions in Matlab.A standard protocol
was followed for optimally focusing the QCL
laser spot to the sample under the AFM tip. The QCL spot position
was swept through the area using the Analysis Studio spot-optimization
software from Anasys Corporation while pulsing at the third cantilever-resonance
mode (∼390 kHz) to reduce the influence of the cantilever heating.
This provided sufficient QCL focus optimization for all samples tested
in the paper.All other data and images shown were collected
using the operations
described in the responsivity-correction-methodology section with
the QCL laser-pulse width set to 500 ns and the lock-in time constant
set according to the scan rate of the collected data (unless otherwise
specified). For example, the polystyrene–polybutadiene–polystyrenepolymer images were collected at a 0.5 Hz scan rate (trace and retrace)
with 1000 × 1000 pixels, resulting in a lock-in time constant
of 1 ms. Resonance tracking was performed using the built-in procedure
for the Nano-IR2 with frequency-threshold values of approximately
±20 kHz around the desired resonant frequency. All data sets
were collected using nominally identical probes.
Polymer-Test-Sample
Preparation
PMMA films were fabricated
by spinning 950PMMA A2 photoresist from MicroChem Corporation to 100
nm thickness. The gold mirrors used were economy gold mirrors from
Thor Laboratories (PN ME05S-M01), and the silicon wafer was from University
Wafers (ID 453). The films were spun at 3000 rpm for 60 s using a
headway spinner and then heated to 180 °C for 5 min. The 1951
United States Air Force (USAF) target was fabricated using a Raith
Eline (electron-beam-lithography system) at a voltage of 10 kV, a
working distance of 10 mm, an area dose of 100 mC/cm2,
and a line dose of 300 PC/cm to generate the USAF pattern. The targets
were then developed in a 1:3 MIBK–IPA solution and heated again
above 125 °C to reflow the polymer to produce smooth features.Polystyrene polybutadiene polymer films were prepared using 0.983
g of a polystyrene–polybutadiene–polystyrene triblock
copolymer from Sigma-Aldrich (PN 432490-250G) mixed with 23 mL of
toluene and spun at 3000 rpm on a low-emissivity (low-E) slide. Films
were scratched to allow for determining the absolute height of the
sample and then heated overnight between 60 to 90 °C to allow
for phase separation of the two polymers. Overnight, the final film
appeared slightly brown and showed phase-separated domains observable
using a visible microscope. The phase separation was also apparent
when observed using FTIR. The FTIR data is provided in Supplemental Section S4.
Cell Culture and Sample
Preparation
MCF 10A (breast
epithelial cells) were grown in Dulbecco’s modified Eagle’s
medium (DMEM) supplemented with horse serum, hydrocortisone, cholera
toxin, epidermal growth factor, insulin, and penicillin–streptomycin.
The cells were grown on sterilized low-E glass until 60–70%
confluence. Finally, the cells were incubated with a 4% paraformaldehyde
solution followed by three PBS washes, quenching with 0.15 M glycine,
two PBS washes, and two sterile-water washes. These fixed cells were
dried overnight for subsequent imaging.
Results and Discussion
Theoretical
Description of the Cantilever-Transfer Function
Here, we
sought to quantify the dependence of the recorded signal
as a function of the actual sample perturbation and the response of
the cantilever. The response of a cantilever to an IR-absorbing sample
has been studied previously.[4,27] We undertook the development
of an analytical model, described in detail in Supplemental Sections S1–S3. A summary of this analysis
as well as specific extensions we make for studying nanoscale IR responses
are explained below. Considering the free-body diagrams shown in Figure a,[28] the position of the cantilever can be described as follows:
Figure 1
Transfer-function validation.
(a) Free-body diagram of the AFM
cantilever beam. The deflection signal (D) is defined
as the slope of the cantilever at xo and
is proportional to the sample expansion (ϵ) via the cantilever-transfer
function (Hc). (b,c) Transfer-function
comparison of the theoretical fit to measured data for a PMMA polymer
film and a gold-coated mirror, respectively. (d) Sample-independent
curve-fit parameters for the fit data shown in (b,c). (e,f) Sample-dependent
fit parameters for the fit data from (b,c).
Transfer-function validation.
(a) Free-body diagram of the AFM
cantilever beam. The deflection signal (D) is defined
as the slope of the cantilever at xo and
is proportional to the sample expansion (ϵ) via the cantilever-transfer
function (Hc). (b,c) Transfer-function
comparison of the theoretical fit to measured data for a PMMApolymer
film and a gold-coated mirror, respectively. (d) Sample-independent
curve-fit parameters for the fit data shown in (b,c). (e,f) Sample-dependent
fit parameters for the fit data from (b,c).Equation is
a normalized
form of Euler–Bernoulli beam theory with a set of boundary
conditions specific to this analysis. Here, m is
the mass of the cantilever, kc is the
cantilever spring constant, Γ is the viscous dampening of the
cantilever, L is the length of the cantilever, Ltip is the length of the cantilever tip, and mtip is the additional tip mass. The properties
that depend on the sample are the expansion signal, ϵ; the lateral
spring and damper parameters, km and km, and the vertical spring and damper
parameters, kf and kf, respectively. These parameters are depicted
in Figure a.Our approach is comparable to expressions from previous theories
with two major differences: there is additional mass at the tip to
account for the tip geometry, and the source that generates the deflection
signal is an out-of-plane sample expansion, ϵ, instead of a
harmonic point force.[4,27] These additions are both rigorous
and necessary for accurately relating the cantilever response to an
out-of-plane sample expansion. One relatively straightforward solution
to this system is, by means of a transfer function,[29] defined by the following:Here, xo is the position of the
deflection
laser on the cantilever and Hc(s) is the cantilever-transfer function. The deflection-laser-position
parameter, xo, is depicted in Figure a. Equation describes the Laplace domain
representation of the input–output response of the cantilever
deflection, D(s), to an out-of-plane,
free-surface sample-expansion signal, ϵ(s).
The expansion signal, ϵ(s), can be considered
the expansion of the surface without the presence of the cantilever
tip, or a stress-free surface expansion. This follows from concepts
in contact mechanics and has been described in previous work.[16,27] Here, we assume the expansion is out-of-plane; however, the deflection
signal could theoretically be influenced by lateral sample motion
as well. The preferential direction of the motion of the sample is
normal to the surface because of the low mechanical impedance of air
(i.e., vertical). Special consideration should be taken for samples
that are mechanically isolated from neighboring material, such as
beads, which would expand isotopically. The vertical-expansion assumption
has proven reliable for all samples considered here.The transfer
function from eq can
also be considered the cantilever’s responsivity.[30] Unlike typical photon detectors, however, the
cantilever’s responsivity is influenced by the sample mechanical
properties, which mask the desired expansion signal. We propose variations
in the cantilever responsivity provide an analytical formulation that
explains the previously reported probe–sample-interaction effect.
The general solution for the transfer function can be determined by
solving the system shown here:The analytical
solution of the transfer function is determined
by performing a matrix inversion of eq and then applying the solution to eq . For clarity and ease of calculation,
the mechanical properties in our formalism have been grouped into
four K values. These K values are
frequency-dependent stiffness functions that are defined depending
on the choice of the tip–sample-stiffness model. For the tip–sample
spring-damper model depicted in Figure a, the K values are defined here:The four K values
shown in eq can be
determined by taking
the Laplace transform of eq . Kc4, Kf, Km, and Ke arise from the resistance
to motion of the cantilever, the tip translation, the tip rotation,
and the sample motion, respectively. Aspects of the responsivity behavior,
such as resonance-frequency shifts, have been demonstrated previously.[4,16] More generally, the definition of the transfer function discussed
here reveals all the intricate changes to the cantilever responsivity
due to the sample mechanical properties.Figure b,c shows
a comparison of the theoretical transfer function and the experimentally
measured data using a standard commercial contact-mode probe for a
100 nm poly(methyl methacrylate) (PMMA) polymer film and a gold substrate,
respectively. Only frequencies between 250 kHz and 2 MHz were chosen
for curve fitting because of noise and discrepancies between the model
and the measured data of Figure (see the Experimental Section for details). We believe there is a behavioral change in the cantilever
response for low frequencies that is not accounted for by the model;
however, for frequencies above 250 kHz, this model provides a theoretical
understanding for improving the accuracy of AFM-IR imaging. Possible
sources of this behavior are discussed in a later section.A
list of curve-fit values with 95% confidence intervals is provided
in Figure d–f.
Because there are only eight unique parameters, approximate values
for kc, Ltip, and L based on supplier data were used. These
three values were assumed to be 0.2 N/m, 10 μm, and 450 μm
respectively. All sample-stiffness values show relatively accurate
trends, and the added tip mass is about 7% of the total mass of the
cantilever. Assuming Hertz contact behavior,[31] we can approximate the stiffness values as the product of the local
effective Youngs modulus and the tip contact-area radius. Assuming
a contact radius of 20 nm, the effective Youngs modulus for PMMA and
gold are 28 and 65 GPa, respectively. These values are largely dependent
on the tip geometry, AFM engagement settings, and film thickness.
Regardless, the values presented here are the correct order of magnitude
and provide accurate relative values. The mass values equate to a
10 μm radius ball of silicon at the end of a silicon cylindrical
beam with a radius of 6 μm and a length of 450 μm. The
addition of this tip mass was essential for accurately describing
the unique shape of the transfer function. This theory could be adapted
to improve the accuracy of measuring the mechanical properties of
samples. The idealized spring model from Figure a depends on the sample mechanical properties
local to the AFM tip (on the order of the tip radius).[32] Stiffness measurements of layered samples (like
the PMMA film here) would have localized depth dependence and could
offer a means to detect surface mechanical properties. For purposes
of this paper, the transfer function is used to provide understanding
of the responsivity variation present in AFM-IR images.
Cantilever
Frequency–Response Investigation
A detailed investigation
of the mathematical nature of this transfer
function leads to two major conclusions: the deflection signal responds
linearly to any out-of-plane sample motion, and the responsivity of
the cantilever is dependent on the mechanical properties of the sample
local to the cantilever tip. To test the transfer function dependence
on the sample mechanical properties, the transfer function was measured
by pulsing a quantum cascade laser (QCL) at 1 kHz on both a gold mirror
and a 100 nm thick PMMA photoresist film. The resulting ringdown was
used to produce the frequency–response curves shown in Figure a. These are the
same curves from Figure b,c, now normalized by the expansion amplitude of PMMA and gold,
respectively, to isolate the probe–sample mechanical interaction.
Figure 2
Piezo
and QCL frequency responses. (a) Measured transfer functions
on 100 nm PMMA film and gold-mirror surface normalized to the scaling
factors determined from their respective curve fits in Figure . The green arrow indicates
a large amplitude change on the fourth resonance mode, and the two
blue arrows indicate locations where the transfer functions overlap,
suggesting little change in responsivity effect for 225 and 420 kHz.
(b) Table of percent relative differences (PRDs) between normalized
measured transfer functions on PMMA and gold for ∼390 (third
resonance), 420, 485, ∼665 (fourth resonance), and 840 kHz.
(c) Ratio of measured gold and PMMA transfer functions using QCL and
piezo.
Piezo
and QCL frequency responses. (a) Measured transfer functions
on 100 nm PMMA film and gold-mirror surface normalized to the scaling
factors determined from their respective curve fits in Figure . The green arrow indicates
a large amplitude change on the fourth resonance mode, and the two
blue arrows indicate locations where the transfer functions overlap,
suggesting little change in responsivity effect for 225 and 420 kHz.
(b) Table of percent relative differences (PRDs) between normalized
measured transfer functions on PMMA and gold for ∼390 (third
resonance), 420, 485, ∼665 (fourth resonance), and 840 kHz.
(c) Ratio of measured gold and PMMA transfer functions using QCL and
piezo.After normalization, the two response
curves overlap at the two
locations indicated by the blue arrows at ∼225 and ∼420
kHz. This overlap indicates two fixed frequencies that are unaffected
by changes in probe–sample mechanical interactions for this
setup. It is important to note that these overlap points are specific
to the cantilever and instrumentation tested and would vary for different
equipment. Interestingly, the amplitudes of the resonance peaks show
significant variations between polymer and substrate. An example is
the amplitude of the fourth resonance mode (665 kHz), indicated by
the green arrow in Figure a. In this example, tracking this resonance peak over a heterogeneous
sample would produce significant signal fluctuations as a result of
the sample mechanical properties. The fifth and sixth cantilever-resonance
modes would be entirely impossible to track as they vanish completely
upon transition between PMMA and the gold substrate. Because the transfer
function is multiplicative, influence of the probe–sample effect
can be quantified as the percent relative difference (PRD) between
the normalized response curves of any two points P1 and P2 on a given
sample, defined as follows:The PRD values for two points located on PMMA
and gold for select
frequencies are shown in Figure b. Low PRD values imply a smaller contribution from
the sample mechanical properties in the PTIR signal. The resonance
modes do not appear to exhibit any unique isolation from mechanical
variations indicated by large PRD values for this sample. In fact,
pulsing at fixed 420 kHz appears to be the best candidate for detecting
the pure sample expansion signal for the equipment tested here. In
general, for an arbitrary sample, measuring the PRD at multiple points
could reveal an optimum fixed pulsing frequency for minimizing responsivity
effects in AFM-IR images.The above studies clearly point to
the role of cantilever responsivity
in both the magnitude and quality of recorded data as well as in the
difficulty in conducting resonance-mode experiments. We hypothesize
that real-time detection of changes in the cantilever responsivity
could greatly improve the fidelity of chemical imaging at nanoscale
resolutions. Atomic force acoustic microscopy (AFAM) is one technique
that uses out-of-plane vibrations generated by a piezo below the sample
for determining the sample mechanical properties.[33,34] Alternative methods exist for determining the mechanical properties
of the sample by vibrating the cantilever (known as force-modulation
mode);[35] however, we propose that out-of-plane
sample vibrations more accurately replicate the photoinduced thermal
expansion.[36,37] Hence, we hypothesize that measuring
the cantilever-response variations in AFM-IR images with a subsample
piezo as used in AFAM measurements can provide an accurate measure
of the transfer-function variation present in the PTIR signal. To
test this idea, the curves in Figure a were remeasured using an out-of-plane vibration generated
by a piezo actuator placed under the sample. Unlike the curves generated
by the QCL alone, the piezo used here has additional acoustic behavior
that makes a direct comparison of QCL and piezo signals impossible;
however, it is only required that the ratio of the two sample locations
have similar frequency responses for proper correction of the responsivity
effect. Figure c shows
the ratio of the measured transfer function on gold and PMMA for both
piezo and QCL with good agreement between 250 kHz and 2 MHz. The bandwidth
of the piezo used throughout this paper is limited to about 2 MHz;
thus, the piezo data becomes increasingly noisy above 1.25 MHz. Additionally,
it is currently unclear as to why the behavior deviates for low frequencies.
For frequencies above 250 kHz, the piezo-signal response to a stiffness
change matches the QCL signal. This data suggests that the piezo signal
can be used at fixed frequencies to completely remove cantilever-responsivity
variations due to the sample mechanical properties, allowing for an
accurate measure of the local sample expansion induced by the absorption
of a pulsed infrared laser. Moreover, this technique allows for accurate
detection of any thermal-expansion signal and could have potential
applications in measuring nanoscale heat transfer as well.[38,39]
Responsivity Correction in Nanoscale Chemical Imaging
We
modified a commercial nano-IR2 system with the addition of a piezo
under the sample. The standard instrument operates by pulsing a QCL
while the AFM scans the sample in the standard AFM trace and retrace
pattern. The deflection signal is then filtered and fed to a lock-in
amplifier to extract the harmonic amplitude of the expansion signal
induced by the QCL absorption. The addition of a piezo under the sample
allows for the generation of a constant out-of-plane mechanical vibration
at the same spatial location and pulsing frequency as the QCL signal
to uniquely determine the cantilever responsivity. Real-time detection
of two harmonic signals with the same frequency, however, is not possible,
so the signals must be separated in either time or frequency space.
The best way to do this would be to scan the same line twice, once
for the QCL signal and again for the piezo. Another path involves
corecording by interlacing the piezo and QCL signals in the same image
with a small enough step size to allow for approximate overlap of
the two signals. This limits the step size to either the smallest
mechanical feature of the sample or the cantilever-tip radius to ensure
accurate overlap and requires minimal changes to the commercial instrument.
A full description of signal processing is provided in the Experimental Section. Figure shows the four-dimensional data set of interlaced
images. After collection, the interlaced lock-in amplitude images
can be separated into the two unique data sets and divided to isolate
the sample-expansion signal. This is the procedure used for the data
presented in this paper. More generally, this process could be extended
to measure complex amplitudes of the expansion signal by processing
the lock-in phase data as well.
Figure 3
Responsivity-correction methodology: four-dimensional
data set
of interlaced QCL and piezo amplitude data showing the cantilever-responsivity-correction
operation. The interlaced images are separated into the raw QCL and
piezo signals and then divided to produce the corrected images.
Responsivity-correction methodology: four-dimensional
data set
of interlaced QCL and piezo amplitude data showing the cantilever-responsivity-correction
operation. The interlaced images are separated into the raw QCL and
piezo signals and then divided to produce the corrected images.
Responsivity and IR-Ratio-Correction
Methods on Polymer Samples
The responsivity effect produces
a multiplicative error that is
constant for different wavenumbers but changes with pulsing frequency.
As a result, the ratio of recorded absorbance at two wavenumbers postacquisition
is a common method for obtaining chemical images.[15,40] The use of two wavenumbers reduces the effectiveness of using a
discrete-frequency imaging approach and has increased susceptibility
to system drift due to sequential image collection. Hence, we do not
recommend the use of this common approach. Moreover, we propose that
identifying the contrast of a single wavenumber without responsivity
variations is only possible with responsivity-correction techniques.
To illustrate the recommendation, we collected PTIR images of a polystyrenepolybutadienepolymer film. Figure a shows the absolute height image near the edge of
the film, and Figure b shows the responsivity-corrected 1485 cm–1 image
of the same region. Figure c shows point spectra taken at the orange and blue points
in Figure b. The 1309
and 1485 cm–1 peaks were selected as characteristic
polybutadiene and polystyrene frequencies, respectively, consistent
with FTIR-spectroscopy data. Additional details are provided in Supplemental Section S4.
Figure 4
Responsivity correction
on polymer samples. (a) AFM height image
flattened and offset relative to the substrate (blue region). Scale
bar is 10 μm. Scan rate is 0.5 Hz. (b) Responsivity-corrected
PTIR image (1485 cm–1, 420 kHz) of the same region
as that in (a). (c) PTIR point spectra taken at the orange and blue
points in (b). (d) Red ROI from (b) showing the raw and responsivity-corrected
PTIR images for 1309 and 1485 cm–1. Blue arrows
indicate regions with high responsivity variation resulting from local
sample mechanical variations. Scale bar is 1 μm. Scan rate is
1 Hz. (e) Raw piezo signal for each pulsing frequency of the same
region as that in (d). (f) Ratio images of 1485 cm–1 divided by 1309 cm–1 using the 485 kHz pulsing
frequency. The left is the ratio using the raw QCL data, and the right
is the ratio using the corrected data.
Responsivity correction
on polymer samples. (a) AFM height image
flattened and offset relative to the substrate (blue region). Scale
bar is 10 μm. Scan rate is 0.5 Hz. (b) Responsivity-corrected
PTIR image (1485 cm–1, 420 kHz) of the same region
as that in (a). (c) PTIR point spectra taken at the orange and blue
points in (b). (d) Red ROI from (b) showing the raw and responsivity-corrected
PTIR images for 1309 and 1485 cm–1. Blue arrows
indicate regions with high responsivity variation resulting from local
sample mechanical variations. Scale bar is 1 μm. Scan rate is
1 Hz. (e) Raw piezo signal for each pulsing frequency of the same
region as that in (d). (f) Ratio images of 1485 cm–1 divided by 1309 cm–1 using the 485 kHz pulsing
frequency. The left is the ratio using the raw QCL data, and the right
is the ratio using the corrected data.To avoid aliasing any small features, a 6 μm region
was selected
and imaged at these wavenumbers for pulsing frequencies 300, 420,
and 485 kHz. The pulsing frequencies were chosen to sample the available
modulation range: the first harmonic of the laser is limited to 500
kHz, and responsivity correction provides high quality correction
above 250 kHz. Moreover, each of these frequencies reveals a significantly
different contrast (as a result of their location on the cantilever-transfer-function
curve). Figure d shows
both the raw and responsivity-corrected PTIR images at these pulsing
frequencies. The raw images of the beadlike feature for 1309 cm–1 show enhanced contrast near the interface indicated
by the blue arrows in Figure d. Without knowledge of the cantilever-responsivity effect,
any one of these images would incorrectly suggest a unique chemical
feature at the interface of this bead domain. This behavior appears
to change with different pulsing frequencies and is equally present
in the raw piezo signal also indicated by blue arrows in Figure e. After dividing
the raw PTIR images with the piezo data, the corrected images for
all pulsing frequencies produce comparable contrast, and the interface
variation becomes completely absent. This suggests the variation at
the interface was the result of contrast due to variations in cantilever
responsivity at different pulsing frequencies. This effect is equally
present in the 1485 cm–1 images as well as in most
images collected to date using this technique to varying degrees.Peak ratios have also been used to remove responsivity variations;[15] however, here we show that measuring the responsivity
variations in real time ensures reliable IR ratio data. Figure f shows 1485–1309 cm–1 infrared peak-ratio images for a 485 kHz pulsing
frequency. The left image is the ratio using the conventionally recorded
data, and the right is the ratio using the corrected images proposed
here. An inconsistency in the ratio images using current-state-of-the-art
methods is shown by the green arrow. This artifact is not present
in the ratio data for the other pulsing frequencies using either technique.
The wavenumber images were collected 15 min apart, over which time
the cantilever response changed slightly as a result of system drift.
Redundancies, such as repeated measurements or hyperspectral imaging,
could rule out such artifacts, but that reduces the effectiveness
of discrete-frequency imaging. Ratio images can be supplemented with
local spectra to provide a better understanding in studies. However,
the decision to scan the spectrum often relies on a few wavenumber
images susceptible to the effects of the sample mechanical properties
and their heterogeneity. Here, real-time detection using the piezo
signal allowed for proper correction of cantilever-responsivity effects
when the ratio method failed.
Responsivity Variations
with Resonance-Tracking Techniques
IR chemical imaging of
biological samples has been widely attempted
with AFM-IR.[8,22,25,41] Resonance-enhanced operation is the current
gold standard for minimizing responsivity effects; however, we have
found that this is not the case for many samples in biology.[16] Here, we demonstrate improvements in the chemical
specificity for AFM-IR imaging compared with the standard resonance-enhanced
operation using MCF-10A wild type mammary epithelial cell samples. Figure a,b shows the raw
PTIR images for a 5 × 5 μm region on a cell sample using
fixed 420 kHz and resonance-enhanced operation at the second resonance
mode, respectively. The image domain and scan speed here were restricted
to ensure accurate tracking for resonance-enhanced operation and sufficient
sampling of all sample features. Tracking accuracy was confirmed by
comparing the trace and retrace signals for consistency. Further details
are provided in Supplemental Figure S4.
Comparing the images from Figure a,b reveals that operating at different laser-repetition
rates can have a significant effect on the contrast because of the
dependence on the sample mechanical properties. We have also confirmed
that the responsivity-corrected 420 kHz image reveals little change
from the raw data, suggesting the difference in contrast shown is
due to responsivity effects present in the resonance-tracking image
(see Supplemental Figure S4 for responsivity-corrected
images). Figure c
shows the second-resonance-mode peak-frequency image for the same
region, which is commonly used to indicate mechanical contrast (resulting
from responsivity variations). Here, the resonance-peak frequency
and amplitude images show a clear correlation, suggesting a strong
influence of responsivity variations in the PTIR signal while tracking
the second cantilever-resonance mode. The green, red, and blue boxed
regions of Figure a–c are enlarged for a clear comparison. Unlike the polymer
sample of Figure ,
the PTIR-signal variations here are largely a result of the surface
topography and illustrate the challenge of imaging samples that are
not prepared with controlled surface characteristics. Tracking a cantilever-resonance
mode is insufficient for imaging the pure-sample expansion isolated
from responsivity variations for heterogeneous samples.
Figure 5
Responsivity
variations on resonance. (a) PTIR image (1236 cm–1) of a 5 × 5 μm region of an MCF-10A wild-type
mammary epithelial cell using a fixed 420 kHz repetition rate. (b)
PTIR image (1236 cm–1) collected by tracking the
second harmonic of the cantilever for the same region as that in (a).
(c) Peak-frequency image of the second resonant mode collected simultaneously
with (b). Green, red, and blue zoomed sections at the bottom compare
the regions of interest indicated in (a–c), respectively. (d)
PTIR line profiles for 1525 cm–1 resonance-tracking
operation using the second cantilever-resonance mode with a scan rate
of 0.05 Hz. Plot shows one scan and the average of 50 scans for calculating
SNR values. (e) Same line profile as in d for the responsivity-corrected
420 kHz fixed-frequency operation for a scan rate of 0.25 Hz. (f)
Comparison of scan rates (SRs), signal-to-noise ratios (SNRs), and
normalized pixel rates (NPRs) for resonance-enhanced operation (using
the second resonance mode) as well as 420 kHz and 485 kHz responsivity-correction
operation using the line profiles from (d,e).
Responsivity
variations on resonance. (a) PTIR image (1236 cm–1) of a 5 × 5 μm region of an MCF-10A wild-type
mammary epithelial cell using a fixed 420 kHz repetition rate. (b)
PTIR image (1236 cm–1) collected by tracking the
second harmonic of the cantilever for the same region as that in (a).
(c) Peak-frequency image of the second resonant mode collected simultaneously
with (b). Green, red, and blue zoomed sections at the bottom compare
the regions of interest indicated in (a–c), respectively. (d)
PTIR line profiles for 1525 cm–1 resonance-tracking
operation using the second cantilever-resonance mode with a scan rate
of 0.05 Hz. Plot shows one scan and the average of 50 scans for calculating
SNR values. (e) Same line profile as in d for the responsivity-corrected
420 kHz fixed-frequency operation for a scan rate of 0.25 Hz. (f)
Comparison of scan rates (SRs), signal-to-noise ratios (SNRs), and
normalized pixel rates (NPRs) for resonance-enhanced operation (using
the second resonance mode) as well as 420 kHz and 485 kHz responsivity-correction
operation using the line profiles from (d,e).In addition to responsivity effects on resonance, we sought
to
quantify the benefits of the theory and the subsequent approach developed
here. Figure d shows
the PTIR signal of a 5 μm line profile taken at the edge of
a breast epithelial cell for 1525 cm–1 using resonance-enhanced
operation at the second resonance mode of the cantilever and a scan
rate of 0.05 Hz. The plot shows a representative single scan as well
as the average of 50 consecutive scans. Figure e shows responsivity-corrected PTIR line
profiles for a pulsing frequency of 420 kHz at a scan rate of 0.25
Hz. For equal comparison, the scan rate and lock-in time constant
were adjusted such that all the data sets had 1000 samples for every
trace scan. Because the resonance-tracking method used requires testing
multiple frequencies for locating the resonance peak, the scan rate
is 5 times slower when compared with that of a fixed-frequency operation. Figure f shows the dark-field-corrected
signal-to-noise-ratio (SNR) calculation for each of these profiles.
The signal and noise measurements were taken from region 1 of Figure d, and the dark-field
signal was taken from the substrate section of region 2 for each scan.
SNR and scan rates do not provide a good metric for directly comparing
imaging techniques. A better way to compare these modalities is via
the normalized pixel rate (NPR) defined by[42]The NPR is proportional to the number of pixels (n) and the well-known scaling between the acquisition time (t) and the resulting SNR for random white noise (SNR ∼ t1/2). Operating at 420 kHz with responsivity
correction is nearly 30 times faster than resonance-enhanced operation
using the second resonance mode. Operation at 485 kHz is still faster
but only by a factor of ∼5 times compared with that of the
resonance-enhanced measurement. This reduction of speed at 485 kHz
is due to an increase in responsivity variations combined with the
repeatability of the instrument reducing the SNR by a factor of 2.
Raw line profiles as well as repeated measurements using smooth samples
(SU8 polymer films) are provided in Supplemental Figure S5. Responsivity correction at frequencies with minimal
responsivity variations allow for rapid imaging of heterogeneous
samples. We emphasize that this demonstration is simply a first example
of the implementation of the theoretical insight; better controls
and hardware could further improve these figures of merit.
Improving
AFM-IR Accuracy and Sensitivity
The sensitivity
of resonance-enhanced AFM-IR has been demonstrated previously;[18] here, we apply responsivity correction to demonstrate
further improvements in sensitivity and accuracy. Figure a shows an AFM height plot
of a 1951 USAF resolution target on a silicon wafer fabricated using
100 nm PMMA e-beam photoresist. Resonance-enhanced-operation PTIR
line profiles were collected along the magenta line in Figure a. Figure b shows the PTIR signal for resonance-tracking
operation using the first, second, and third cantilever-resonance
modes (i–iii, respectively). The magenta- and black-line plots
were collected using different QCL laser-focus positions (R1 and R2,
respectively). The QCL laser-focus-optimization plots are shown to
the right of their respective line profiles. The focus-optimization
plots indicate the QCL focus position that produces the highest deflection
response. The maximum signal is expected to occur when the laser is
optimally focused on the sample under the AFM tip. For the first and
second resonance modes, the highest signal in the focus-optimization
plots occurs in region R2; however, the PTIR data produced with this
laser-focus position shows little contrast. This is true for all three
black-line profiles, and most likely suggests that the laser is focused
directly to the cantilever beam. Operation using the third resonance
mode shows the highest signal with the QCL laser focused to region
R1, and the respective magenta-line profile shows reduced noise and
improved contrast (resembling the sample height profile) when compared
with that of the other two resonance modes. This data suggests the
QCL focus position R1 is optimal for maximizing the signal produced
by the sample expansion and minimizing the effects of cantilever heating
on the recorded data. Moreover, this result demonstrates an increased
susceptibility to cantilever heating for the lower cantilever-resonance
modes, which is the most likely culprit for the deviation of the measured
data and the transfer-function fit observed previously. Top-side QCL
illumination, as opposed to the earlier practice of evanescent heating
in AFM-IR, will likely heat the cantilever more, which, for very weakly
absorbing samples (where sensitivity is crucial), will render the
first- and second-harmonic modes less effective. One solution is to
operate at higher repetition rates (above the third harmonic, as the
data suggests). Figure c shows the normalized transfer function on PMMA and silicon for
this setup. From lessons learned modeling the cantilever, the presence
of the additional mass in the cantilever tip results in large amplitude
variations near the fourth (∼610 kHz) and fifth (∼940
kHz) cantilever-resonance modes, making tracking methods virtually
impossible in this frequency regime or, at the very least, highly
susceptible to the sample mechanical properties. Thus, responsivity
correction at fixed frequencies above the third harmonic is one practical
solution for avoiding the influence of cantilever heating on recorded
data and removing the sample mechanical variations. These factors
illustrate the practical challenges in the present state-of-the-art
of nanoscale IR measurements and often necessitate careful sample
preparation or experimental design by experts. Our proposed approach
implements data collection in a manner that will enable this technology
to be broadly used by those unfamiliar with the intricate details
of potential artifacts.
Figure 6
Accurate detection of polymer thermal expansion.
(a) Height image
of group 6 element 4 of 100 nm thick PMMA 1951 USAF target. (b) PTIR-signal
line profiles collected at the magenta line in (a). (i–iii)
Line profiles for the first-, second-, and third-harmonic resonance-tracking
operation, respectively. The frequencies for the first, second, and
third resonance modes are approximately 66, 185, and 390 kHz, respectively.
The magenta and black plots were taken with the QCL laser focused
to regions R1 and R2, respectively. The magenta plot was normalized
with its maximum value, and the black-plot mean value was set to 1.5
for clarity. (iv–vi) PTIR-signal intensities with the QCL focus
scanned across the sample for first-, second-, and third-harmonics,
respectively. Regions R1 and R2 light up because of the focus on the
sample and cantilever beam, respectively. The line profiles reveal
the most accurate signal when the QCL is focused to region R1 and
show decreased noise as the frequency increases. (c) Transfer function
for the black and orange points in (a). These plots are normalized
according to the procedure shown in Figure . Tracking the fifth-harmonic (940 kHz) resonance
mode failed because of large variations in peak amplitude. (d) Piezo,
QCL, and corrected signal for fixed 940 kHz pulsing frequency at the
magenta line in (a). The magenta lines are the actual signals, and
the black lines are the predicted signals using only the measured-height
profile.
Accurate detection of polymer thermal expansion.
(a) Height image
of group 6 element 4 of 100 nm thick PMMA 1951 USAF target. (b) PTIR-signal
line profiles collected at the magenta line in (a). (i–iii)
Line profiles for the first-, second-, and third-harmonic resonance-tracking
operation, respectively. The frequencies for the first, second, and
third resonance modes are approximately 66, 185, and 390 kHz, respectively.
The magenta and black plots were taken with the QCL laser focused
to regions R1 and R2, respectively. The magenta plot was normalized
with its maximum value, and the black-plot mean value was set to 1.5
for clarity. (iv–vi) PTIR-signal intensities with the QCL focus
scanned across the sample for first-, second-, and third-harmonics,
respectively. Regions R1 and R2 light up because of the focus on the
sample and cantilever beam, respectively. The line profiles reveal
the most accurate signal when the QCL is focused to region R1 and
show decreased noise as the frequency increases. (c) Transfer function
for the black and orange points in (a). These plots are normalized
according to the procedure shown in Figure . Tracking the fifth-harmonic (940 kHz) resonance
mode failed because of large variations in peak amplitude. (d) Piezo,
QCL, and corrected signal for fixed 940 kHz pulsing frequency at the
magenta line in (a). The magenta lines are the actual signals, and
the black lines are the predicted signals using only the measured-height
profile.To demonstrate our approach, instead
of tracking the fifth resonance
peak, we collected profiles using a fixed 940 kHz pulsing frequency
and applied responsivity correction. In addition, we illustrate that
the signal produced by the subsample piezo represents the local cantilever-responsivity
contrast by predicting the raw QCL and piezo profiles using only the
silicon and PMMA transfer-function-curve fits and the AFM height profile.
Because the laser used here is limited to a 500 kHz repetition rate,
we measured the laser second harmonic produced by pulsing at 470 kHz
with a 300 ns pulse width. The magenta plots of Figure d show raw piezo, raw QCL, and responsivity-corrected
line profiles for the magenta line in Figure a. Using the curve fits from the transfer
functions on PMMA and silicon, we can accurately predict the piezo
signal profile; however, the transfer functions only represent two
locations on the sample. Because the local sample mechanical properties
are dependent on film thickness, we apply a standard correction proposed
by Doerner and Nix.[32] The predicted piezo
signal using the transfer-function-fit data suggests that the measured
contrast results from changes in the cantilever responsivity at 940
kHz. The presence of this effect is also apparent in the raw QCL signal.
To predict the QCL data, the local sample expansion is also required.
The sample chosen for this experiment is sufficiently smooth and thin,
which allows for approximating the local sample expansion as a one-dimensional
thin-film expansion (i.e., linear with thickness). Thus, the expansion
of this sample, to the first approximation, should be proportional
to the local thickness of the film with an additional offset to account
for substrate heating. Details regarding the one-dimensional-expansion
model are provided in Supplemental Section S2.The measured height data and transfer-function fits allow
for predicting
the raw piezo and QCL signals. More sophisticated analytical models
could be incorporated to better understand the thermal-expansion behavior
of samples with well-defined geometries and relate the responsivity-corrected
PTIR data to sample expansion. This could provide a heightened understanding
of the governing thermoelastic behavior of these materials. For more
complex geometry, predicting these signals would require more information
and numerical methods for determining the sample expansion. Regardless,
this demonstrates that correcting AFM-IR with the signal generated
by the piezo enables accurate, model-free detection of the PTIR signal
free of responsivity effects. This contrast is theoretically proportional
to the desired sample expansion, which more closely resembles the
desired spectral information. Additionally, responsivity correction
can be performed at any pulsing frequency and is not restricted to
low-frequency resonance modes. This allows for lower noise, higher
sensitivity, and more accurate imaging than that of resonance-enhanced
operation.
Conclusion
Detection of photoinduced
thermal expansion with atomic force microscopy
(AFM) offers highly sensitive, nanoscale correlated chemical imaging.
However, variations in probe–sample mechanical interactions
corrupt the underlying chemical contrast. These variations are a direct
result of changes in the cantilever response to sample expansion.
Here, we developed an analytical understanding of the process and
provided practical paths to realizing its advantages. Using a mechanically
induced out-of-plane vibration, we show that the responsivity variations
can be measured and removed from the AFM-IR signal to isolate the
sample expansion. Removing responsivity variations in this way allows
for fixed-pulsing-frequency operation, which was shown to improve
signal sensitivity by operating outside the noise bandwidth of the
system where resonance tracking fails. The methods proposed here demonstrate
a more robust chemical-imaging modality with improved accuracy and
repeatability when compared with those of the present state-of-the-art
(i.e., resonance-enhanced operation). Better piezo controls and hardware
as well as higher frequencies offer untapped potential in terms of
sensitivity and accuracy, which resonance-enhanced operation alone
will never achieve. This work should lead to practical achievements
of high-fidelity, robust, lower-noise, and faster nanoscale IR imaging.
Moreover, by eliminating the need for detailed knowledge of artifacts
and pitfalls to avoid in acquiring accurate data by means of a theoretical
understanding, it paves the way for this emerging technique to be
widely used by nanoscale researchers with confidence.
Authors: Jungseok Chae; Sangmin An; Georg Ramer; Vitalie Stavila; Glenn Holland; Yohan Yoon; A Alec Talin; Mark Allendorf; Vladimir A Aksyuk; Andrea Centrone Journal: Nano Lett Date: 2017-08-08 Impact factor: 11.189
Authors: Matthew R Rosenberger; Michael Cai Wang; Xu Xie; John A Rogers; SungWoo Nam; William P King Journal: Nanotechnology Date: 2017-06-28 Impact factor: 3.874
Authors: L Baldassarre; V Giliberti; A Rosa; M Ortolani; A Bonamore; P Baiocco; K Kjoller; P Calvani; A Nucara Journal: Nanotechnology Date: 2016-01-18 Impact factor: 3.874