A Novel Microflow Phantom Dedicated to Ultrasound Microvascular Measurements.

Virginie Grand-Perret1, Jean-René Jacquet2, Ingrid Leguerney1,3, Baya Benatsou1,3, Jean-Marc Grégoire2, Georges Willoquet1, Ayache Bouakaz2, Nathalie Lassau1,3, Stephanie Pitre-Champagnat1.   

Abstract

Tumor microvascularization is a biomarker of response to antiangiogenic treatments and is accurately assessed by ultrasound imaging. Imaging modes used to visualize slow flows include Power Doppler imaging, dynamic contrast-enhanced ultrasonography, and more recently, microvascular Doppler. Flow phantoms are used to evaluate the performance of Doppler imaging techniques, but they do not have a steady flow and sufficiently small channels. We report a novel device for robust and stable microflow measurements and the study of the microvascularization. Based on microfluidics technology, the prototype features wall-less cylindrical channels of diameters ranging from as small as 147 up to 436 µm, cast in a soft silicone polymer and perfused via a microfluidic flow pressure controller. The device was assessed using flow rates from 49 to 146 µL/min, with less than 1% coefficient of variation over three minutes, corresponding to velocities of 6 to 142 mm/s. This enabled us to evaluate and confirm the reliability of the Superb Microvascular Imaging Doppler mode compared with the Power Doppler mode at these flow rates in the presence of vibrations mimicking physiological motion.

Entities:  

Keywords:  flow imaging; flow phantom; microfluidics; microvascular Doppler; power Doppler imaging; superb microvascular imaging; ultrasound microvasculature

Mesh:

Year:  2018        PMID: 29923458     DOI: 10.1177/0161734618783975

Source DB:  PubMed          Journal:  Ultrason Imaging        ISSN: 0161-7346            Impact factor:   1.578


  3 in total

1.  Non-invasive Small Vessel Imaging of Human Thyroid Using Motion-Corrected Spatiotemporal Clutter Filtering.

Authors:  Rohit Nayak; Viksit Kumar; Jeremy Webb; Mostafa Fatemi; Azra Alizad
Journal:  Ultrasound Med Biol       Date:  2019-02-02       Impact factor: 2.998

2.  Compatibility and Validation of a Recent Developed Artificial Blood through the Vascular Phantom Using Doppler Ultrasound Color- and Motion-mode Techniques.

Authors:  Marwan Alshipli; Mohannad Adel Sayah; Ammar A Oglat
Journal:  J Med Ultrasound       Date:  2020-04-23

Review 3.  A Brief Review for Common Doppler Ultrasound Flow Phantoms.

Authors:  Sabri M Shalbi; Ammar A Oglat; Balid Albarbar; Fuzi Elkut; M A Qaeed; Ali Abu Arra
Journal:  J Med Ultrasound       Date:  2020-06-02
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.