Novel motorized spiral enteroscopy: first clinical case.

Horst Neuhaus1,2, Torsten Beyna1,2, Markus Schneider1,2, Jacques Devière2.   

Abstract

Entities:  

Year:  2016        PMID: 29905207      PMCID: PMC5989127          DOI: 10.1016/j.vgie.2016.08.005

Source DB:  PubMed          Journal:  VideoGIE        ISSN: 2468-4481


× No keyword cloud information.
A novel reusable endoscope (Olympus Corp, Tokyo, Japan) with an integrated motor (Fig. 1A) was developed for rotating a disposable short spiral overtube mounted on the insertion tube portion. The drive motor located in the endoscope handle is activated by foot pedals and controls the direction and speed of rotation of a coupler located in the middle of the endoscope’s insertion tube (Figs. 1B and C). The rotation coupler is the only portion of the endoscope that rotates; the distal and proximal portions of the insertion tube do not rotate. Other than the presence of the drive motor and the rotation coupler, the design of the endoscope is the same as that of conventional flexible endoscopes (Fig. 1D). The single-use spiral assembly is composed of corrugated tubing with an atraumatic plastic spiral bonded to its exterior. It relies on rotation of the spiral component to “pleat” or “unpleat” the bowel either on or off the insertion tube as the spiral thread rotates in a clockwise or counterclockwise direction, respectively. The operator can monitor how much torque is being applied to the small bowel by observing the system’s visual force gauge. This display provides the operator with a visual indication of the direction of rotation and the force of rotation throughout the examination (Fig. 1E). This clinical case represents the first use of this device in humans. Enteroscopy was indicated for treatment of angiodysplasias in the jejunum identified by small-bowel capsule endoscopy in a 48-year-old patient with iron-deficiency anemia. The novel motorized enteroscope could be smoothly inserted approximately 250 cm distal to the ligament of Treitz within 20 minutes (Fig. 1F). The technique allowed controlled movement of the tip of the endoscope, and it provided excellent visualization of the intubated small bowel. An angiodysplastic lesion was detected (Fig. 1G) and treated with argon plasma coagulation (Fig. 1H; Video 1, available online at www.VideoGIE.org). Careful removal of the endoscope with counterclockwise rotation of the spiral revealed no iatrogenic mucosal trauma. No adverse events were registered.
Figure 1

A, Integrated electric motor of the novel enteroscope. B, Disposable short spiral overtube and rotation coupler of the insertion tube. C, The spiral overtube is connected to the rotation coupler. D, Prototype spiral enteroscope ready for use. E, Spiral rotation force indicator. F, Insertion of spiral portion of the endoscope. G, Detection of arteriovenous malformation of the jejunum during spiral enteroscopy. H, Treatment of arteriovenous malformation by argon plasma coagulation.

A, Integrated electric motor of the novel enteroscope. B, Disposable short spiral overtube and rotation coupler of the insertion tube. C, The spiral overtube is connected to the rotation coupler. D, Prototype spiral enteroscope ready for use. E, Spiral rotation force indicator. F, Insertion of spiral portion of the endoscope. G, Detection of arteriovenous malformation of the jejunum during spiral enteroscopy. H, Treatment of arteriovenous malformation by argon plasma coagulation.

Disclosure

Dr Neuhaus and Dr Devière are consultants for Olympus Corp. All other authors disclosed no financial relationships relevant to this publication.
  9 in total

1.  A spiraling journey into the small bowel: a case series of novel motorized power spiral enteroscopies.

Authors:  Madhura Prasad; V G Mohan Prasad; A Sangameswaran; Sunil Chacko Verghese; Vamsi Murthy; Mithra Prasad; Gokul Kruba Shanker; Santhosh Koppal
Journal:  VideoGIE       Date:  2020-09-04

2.  Systematic review with meta-analysis: efficacy of balloon-assisted enteroscopy for dilation of small bowel Crohn's disease strictures.

Authors:  Dominik Bettenworth; Arne Bokemeyer; Lei Kou; Rocio Lopez; James F Bena; Sara El Ouali; Ren Mao; Satya Kurada; Amit Bhatt; Torsten Beyna; Brendan Halloran; Matthew Reeson; Shuhei Hosomi; Masahiro Kishi; Fumihito Hirai; Naoki Ohmiya; Florian Rieder
Journal:  Aliment Pharmacol Ther       Date:  2020-08-19       Impact factor: 8.171

3.  Detection Rate and Clinical Relevance of Ink Tattooing during Balloon-Assisted Enteroscopy.

Authors:  C Römmele; A Ebigbo; M Schrempf; H Messmann; S K Gölder
Journal:  Gastroenterol Res Pract       Date:  2017-11-05       Impact factor: 2.260

Review 4.  Malignant Melanoma of the Gastrointestinal Tract: Symptoms, Diagnosis, and Current Treatment Options.

Authors:  Darina Kohoutova; Dominic Worku; Hala Aziz; Julian Teare; Justin Weir; James Larkin
Journal:  Cells       Date:  2021-02-05       Impact factor: 6.600

Review 5.  Video Capsule Endoscopy and Device-Assisted Enteroscopy.

Authors:  Mark Hanscom; Courtney Stead; Harris Feldman; Neil B Marya; David Cave
Journal:  Dig Dis Sci       Date:  2021-08-12       Impact factor: 3.199

Review 6.  The Evolution of Device-Assisted Enteroscopy: From Sonde Enteroscopy to Motorized Spiral Enteroscopy.

Authors:  Fredy Nehme; Hemant Goyal; Abhilash Perisetti; Benjamin Tharian; Neil Sharma; Tony C Tham; Rajiv Chhabra
Journal:  Front Med (Lausanne)       Date:  2021-12-23

Review 7.  Recent developments in small bowel endoscopy: the "black box" is now open!

Authors:  Luigina Vanessa Alemanni; Stefano Fabbri; Emanuele Rondonotti; Alessandro Mussetto
Journal:  Clin Endosc       Date:  2022-07-14

8.  Safety and efficacy of the novel motorized power spiral enteroscopy: A single-center experience.

Authors:  Pankaj Singh; Vikas Singla; Sawan Bopanna; Muzaffer Rashid Shawl; Pallavi Garg; Jatin Agrawal; Anupama Arya; Vibhuti Mittal; Richa Bhargava; Kaushal Madan
Journal:  DEN open       Date:  2022-07-11

9.  Endoscopic and percutaneous biliary interventions in patients with altered upper gastrointestinal anatomy-the Munich Multicenter Experience.

Authors:  Simon Nennstiel; Kathrin Freivogel; Alexander Faber; Christoph Schlag; Bernhard Haller; Martin Blöchinger; Markus Dollhopf; Björn Lewerenz; Wolfgang Schepp; Jörg Schirra; Roland M Schmid; Bruno Neu
Journal:  Surg Endosc       Date:  2021-01-04       Impact factor: 4.584

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.