Alarming increase in HbA1c and near misdiagnosis of diabetes mellitus resulting from a clinical laboratory instrument upgrade and haemoglobin variant.

Steven D Chessler1, Donald E Lee1.   

Abstract

A 29-year-old woman was referred for new-onset diabetes mellitus after her glycosylated haemoglobin (HbA1c) was found to be 10.2%. Three years earlier, the patient's HbA1c-measured by the same clinical laboratory-had been 5.5%. The newer HbA1c was discordant with fasting glucose levels and a lack of diabetes-associated symptoms. The laboratory reported that their assay methodology remained unchanged and also that no haemoglobin variants were detected. Further investigation, however, revealed, first, that the patient carried a haemoglobin alpha chain mutation (Hb Wayne) that can sometimes cause assay interference and, second, that although the laboratory's assay methodology had not changed, their assay instrument had. Depending on assay methodology, haemoglobin variants can cause HbA1c assay interference and the presence of these variants may not be detected by the performing laboratory. Interference may not only be dependent on assay methodology but also on the assay instrument used. © BMJ Publishing Group Ltd (unless otherwise stated in the text of the article) 2018. All rights reserved. No commercial use is permitted unless otherwise expressly granted.

Entities:  

Keywords:  diabetes; endocrinology; haematology (incl blood transfusion)

Mesh:

Substances:

Year:  2018        PMID: 29903779      PMCID: PMC6011579          DOI: 10.1136/bcr-2018-225358

Source DB:  PubMed          Journal:  BMJ Case Rep        ISSN: 1757-790X


  7 in total

Review 1.  Pathology consultation on HbA(1c) methods and interferences.

Authors:  Jeanne M Rhea; Ross Molinaro
Journal:  Am J Clin Pathol       Date:  2014-01       Impact factor: 2.493

Review 2.  Effects of hemoglobin variants and chemically modified derivatives on assays for glycohemoglobin.

Authors:  L Bry; P C Chen; D B Sacks
Journal:  Clin Chem       Date:  2001-02       Impact factor: 8.327

3.  Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus.

Authors:  David B Sacks; Mark Arnold; George L Bakris; David E Bruns; Andrea Rita Horvath; M Sue Kirkman; Ake Lernmark; Boyd E Metzger; David M Nathan
Journal:  Diabetes Care       Date:  2011-06       Impact factor: 19.112

4.  Hemoglobin variants detected by hemoglobin A1c (HbA1c) analysis and the effects on HbA1c measurements.

Authors:  Nadzimah Mohd Nasir; M Thevarajah; Chew Yee Yean
Journal:  Int J Diabetes Dev Ctries       Date:  2010-04

5.  Structural and functional studies of hemoglobin Wayne: an elongated alpha-chain variant.

Authors:  W F Moo-Penn; D L Jue; M H Johnson; M J McDonald; S M Turci; T B Shih; R T Jones; B L Therrell; A Arnone
Journal:  J Mol Biol       Date:  1984-12-25       Impact factor: 5.469

6.  Hemoglobin Wayne: a frameshift mutation detected in human hemoglobin alpha chains.

Authors:  M Seid-Akhavan; W P Winter; R K Abramson; D L Rucknagel
Journal:  Proc Natl Acad Sci U S A       Date:  1976-03       Impact factor: 11.205

7.  Identification of Hb Wayne and its effects on HbA1c measurement by 5 methods.

Authors:  Karina Rodríguez-Capote; Mathew P Estey; Vilte E Barakauskas; Teralee Burton; Deborah Holmes; Richard Krause; Trefor N Higgins
Journal:  Clin Biochem       Date:  2015-08-04       Impact factor: 3.281

  7 in total
  1 in total

1.  Hemoglobin Wayne causing a falsely elevated hemoglobin A1c.

Authors:  Alexis Bejcek; David Wenkert
Journal:  Proc (Bayl Univ Med Cent)       Date:  2021-10-15
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.