Dapeng Wang1, Fuqin Bao1, Qiang Li2, Yugang Teng2, Jianjun Li3.
Abstract
OBJECTIVE: To investigate the effect of additional semiautomatic intermittent pneumatic compression device (IPCD) in the prevention of deep vein thrombosis (DVT) of lower extremity in these patients undergoing major orthopedic surgery, when compared with the conventional graduated compression stockings alone.Entities:
Keywords: Deep vein thrombosis; Graduated compression stockings; Intermittent pneumatic compression device; Orthopedic surgery
Mesh:
Year: 2018 PMID: 29903003 PMCID: PMC6002995 DOI: 10.1186/s12938-018-0513-5
Source DB: PubMed Journal: Biomed Eng Online ISSN: 1475-925X Impact factor: 2.819
Baseline characteristics of included patients
| Experimental group (N = 51) | Control group (N = 61) | |
|---|---|---|
| Gender (male/female) | 29/22 | 36/25 |
| Age (x ± s, years old) | 55.6 ± 14.2 | 50.6 ± 19.2 |
| Body weight (x ± s, Kg) | 61.2 ± 9.5 | 61.8 ± 8.5 |
| Primary diseases | ||
| Femoral neck fracture | 14 | 15 |
| Ischemic necrosis of femoral head | 6 | 9 |
| Hip osteoarthritis | 7 | 9 |
| Knee joint osteoarthritis | 11 | 15 |
| Femoral dysplasia | 13 | 13 |
| Anesthesia method | ||
| General anesthesia in the trachea | 45 | 51 |
| Continuous epidural anesthesia | 6 | 10 |
| Prosthesis fixation | ||
| Non-cement fixation | 7 | 9 |
| Cement fixation | 32 | 36 |
| Mixed fixation | 12 | 16 |
Fig. 1Comparison of primary diseases between graduated compression stockings alone and IPCD in addition
Fig. 2Comparison of anesthesia method and prosthesis fixation between graduated compression stockings alone and IPCD in addition
Fig. 3Comparison of DVT incidence between graduated compression stockings alone and IPCD in addition
Comparison of blood flow in femoral vein between graduated compression stockings alone and IPCD in addition
| Group | Number | Prior treatment | Posttreatment |
|---|---|---|---|
| Experimental group | 51 | 114.23 ± 23.69 | 149.56 ± 26.35 |
| Control group | 61 | 118.12 ± 22.11 | 130.15 ± 22.56 |
| t | 0.017 | 12.011 | |
| P | 0.954 | 0.000 |
Fig. 4Comparison of blood flow in femoral vein between graduated compression stockings alone and IPCD in addition
Perioperative blood loss and bleeding signs between graduated compression stockings alone and IPCD in addition
| Group | Number | Bleeding signs | Perioperative blood loss (ml) |
|---|---|---|---|
| Experimental group | 51 | 0 | 800.5 ± 320.7 |
| Control group | 61 | 1 | 950.1 ± 305.9 |
Fig. 5Comparison of perioperative blood loss and bleeding signs between graduated compression stockings alone and IPCD in addition