| Literature DB >> 29899470 |
Yanjia Zhu1, Caizheng Geng2, Jia Huang3, Juzhen Liu4, Ning Wu5, Jun Xin6, Hao Xu7, Lijuan Yu8, Jianhua Geng9.
Abstract
To ensure the reliability of the planned multi-center clinical trial, we assessed the consistence and comparability of the quantitative parameters of the eight PET/CT units that will be used in this trial. PET/CT images were scanned using a PET NEMA image quality phantom (Biodex) on the eight units of Discovery PET/CT 690 from GE Healthcare. The scanning parameters were the same with the ones to be used in the planned trial. The 18F-NaF concentration in the background was 5.3 kBq/ml, while the ones in the spheres of diameter 37 mm, 22 mm, 17 mm and 10 mm were 8:1 as to that of the background and the ones in the spheres of diameter 28 mm and 13 mm were 0 kBq/ml. The consistency of hot sphere recovery coefficient (HRC), cold sphere recovery coefficient (CRC), hot sphere contrast (QH) and cold sphere contrast (Qc) among these 8 PET/CTs was analyzed. The variation of the main quantitative parameters of the eight PET/CT systems was within 10%, which is acceptable for the clinical trial.Entities:
Mesh:
Year: 2018 PMID: 29899470 PMCID: PMC5997997 DOI: 10.1038/s41598-018-27143-4
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
HRCmean, HRCmax, and CRCmean in eight PET/CT systems.
| PET/CT | 37 mm | 37 mm | 22 mm | 22 mm | 17 mm | 17 mm | 10 mm | 10 mm | 28 mm | 13 mm |
|---|---|---|---|---|---|---|---|---|---|---|
| number | HRCmean | HRCmax | HRCmean | HRCmax | HRCmean | HRCmax | HRCmean | HRCmax | CRCmean | CRCmean |
| 1 | 0.93 | 1.2 | 0.77 | 1.14 | 0.69 | 1.17 | 0.44 | 0.64 | 0.32 | 0.61 |
| 2 | 0.9 | 1.22 | 0.68 | 1.09 | 0.65 | 1.15 | 0.5 | 0.66 | 0.35 | 0.54 |
| 3 | 0.87 | 1.18 | 0.72 | 1.12 | 0.65 | 1.11 | 0.44 | 0.78 | 0.27 | 0.5 |
| 4 | 0.93 | 1.17 | 0.75 | 1.11 | 0.68 | 1.09 | 0.48 | 0.62 | 0.32 | 0.51 |
| 5 | 0.99 | 1.29 | 0.81 | 1.32 | 0.73 | 1.29 | 0.54 | 0.75 | 0.32 | 0.54 |
| 6 | 0.89 | 1 | 0.73 | 0.92 | 0.67 | 0.97 | 0.46 | 0.51 | 0.34 | 0.44 |
| 7 | 0.85 | 1.11 | 0.75 | 1.12 | 0.63 | 1.06 | 0.45 | 0.6 | 0.34 | 0.55 |
| 8 | 0.89 | 1.19 | 0.73 | 1.15 | 0.65 | 1.15 | 0.41 | 0.57 | 0.28 | 0.48 |
| Mean | 0.91 | 1.17 | 0.74 | 1.12 | 0.67 | 1.12 | 0.46 | 0.64 | 0.32 | 0.52 |
| SD | 0.04 | 0.09 | 0.04 | 0.11 | 0.03 | 0.09 | 0.04 | 0.09 | 0.03 | 0.05 |
| Variability/% | 4.40 | 7.69 | 5.41 | 9.82 | 4.48 | 8.04 | 8.70 | 14.06 | 9.38 | 9.62 |
Figure 1HRCmax distribution with spheres of different sizes in eight PET/CT systems. Notes: number 1 to 8 represents for the serial number of PET/CT systems.
Figure 3CRCmean distribution with spheres of different sizes in eight PET/CT systems. Notes: number 1 to 8 represents for the serial number of PET/CT systems.
Hot sphere contrast QH and cold sphere contrast Qc in eight PET/CT systems.
| PET/CT number | 37 mm | 22 mm | 17 mm | 10 mm | 28 mm | 13 mm |
|---|---|---|---|---|---|---|
| QH | QH | QH | QH | Qc | Qc | |
| 1 | 78.12 | 73.87 | 62.23 | 29.76 | 70.91 | 38.77 |
| 2 | 81.69 | 66.06 | 49.45 | 45.95 | 65.22 | 54.57 |
| 3 | 76.74 | 74.08 | 54.76 | 22.83 | 72.73 | 48 |
| 4 | 79.94 | 71.71 | 53.98 | 27.31 | 63.81 | 43.09 |
| 5 | 87.63 | 68.87 | 59.8 | 44.98 | 71.07 | 48.35 |
| 6 | 76.51 | 64.79 | 52.65 | 26.89 | 66.37 | 55.97 |
| 7 | 71.69 | 72.49 | 52.25 | 31.38 | 66.09 | 45.39 |
| 8 | 80.28 | 60.78 | 55.03 | 29.38 | 73.28 | 52.08 |
| Mean | 79.08 | 69.08 | 55.02 | 32.31 | 68.69 | 48.28 |
| SD | 4.62 | 4.83 | 4.15 | 8.51 | 3.70 | 5.84 |
| Variability/% | 5.85 | 6.98 | 7.54 | 26.33 | 5.39 | 12.10 |
Figure 4QH distribution with spheres of different sizes in eight PET/CT systems. Notes: number 1 to 8 represents for the serial number of PET/CT systems.
Figure 5Qc distribution with spheres of different sizes in eight PET/CT systems. Notes: number 1 to 8 represents for the serial number of PET/CT systems.
Background variability in eight PET/CT systems.
| PET/CT number | 37 mm | 28 mm | 22 mm | 17 mm | 13 mm | 10 mm |
|---|---|---|---|---|---|---|
| N | N | N | N | N | N | |
| 1 | 3.98 | 4.61 | 5.51 | 6.88 | 7.94 | 8.78 |
| 2 | 3.93 | 4.78 | 5.81 | 6.3 | 7.97 | 7.95 |
| 3 | 3.3 | 4.62 | 5.72 | 6.42 | 8.08 | 8.26 |
| 4 | 4.28 | 4.82 | 5.42 | 6.51 | 7.45 | 7.16 |
| 5 | 4 | 4.35 | 4.66 | 6.04 | 6.72 | 9.38 |
| 6 | 4.18 | 4.92 | 5.42 | 6.13 | 7.61 | 7.3 |
| 7 | 3.45 | 4.31 | 4.71 | 4.96 | 5.34 | 7.89 |
| 8 | 3.38 | 4.08 | 4.47 | 6.35 | 6.98 | 6.77 |
| Mean | 3.81 | 4.56 | 5.22 | 6.20 | 7.26 | 7.94 |
| SD | 0.38 | 0.29 | 0.52 | 0.56 | 0.91 | 0.87 |
| variability/% | 9.97 | 6.36 | 9.99 | 9.06 | 12.59 | 10.93 |
Figure 6Background variability distribution with different ROI sizes in eight PET/CT systems. Notes: number 1 to 8 represents for the serial number of PET/CT systems.
Figure 2HRCmean distribution with spheres of different sizes in eight PET/CT systems. Notes: number 1 to 8 represents for the serial number of PET/CT systems.