Literature DB >> 29888971

A comparison of bladder volumes based on treatment planning CT and BladderScan® BVI 6100 ultrasound device in a prostate radiation therapy population.

Laura Mullaney1,2, Evelyn O'Shea2,3, Mary T Dunne2,4, Pierre G Thirion2,5, John G Armstrong2,5.   

Abstract

OBJECTIVE: : The aim of this study is to investigate if a handheld ultrasound device (BladderScan® BVI 6100) can accurately measure bladder volumes in prostate radiotherapy (RT) patients.
METHODS: : A comparison was made of contoured bladder volumes based on treatment planning CT (TPCT) and BladderScan® BVI 6100 ultrasound device in a large prostate RT population. Three bladder volume (BV) measurements were taken using the bladder volume instrument (BVI) device on prostate RT patients immediately prior to TPCT (n = 190). The CT delineation bladder volumes were also recorded. The mean of the three BVI readings (BVImean) and the maximum (BVImax) of the readings were considered for a comparative analysis.
RESULTS: : There was a strong positive correlation between the BVI and CT delineated bladder volumes (BVImean r = 0.825; BVImax r = 0.830). The mean difference [± standard deviation (SD)] was an underestimation of BV for both BVImean and BVImax (44.8 ± 88.2 ml and 32.9 ± 87.5 ml, respectively).
CONCLUSION: : This is the largest study to date (n = 190), assessing the accuracy of the BladderScan® BVI 6100 in the prostate RT population. The BVI 6100 provides an acceptable indication of BV for use in prostate RT patients for the purposes of monitoring BV. ADVANCES IN KNOWLEDGE:: The BladderScan® BVI 6100 provides a convenient and non-irradiating method of indicating BV for use in prostate RT patients.

Entities:  

Mesh:

Year:  2018        PMID: 29888971      PMCID: PMC6475940          DOI: 10.1259/bjr.20180160

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  10 in total

1.  A practical method to achieve prostate gland immobilization and target verification for daily treatment.

Authors:  A V D'Amico; J Manola; M Loffredo; L Lopes; K Nissen; D A O'Farrell; L Gordon; C M Tempany; R A Cormack
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-12-01       Impact factor: 7.038

2.  Clinical agreement between automated and calculated ultrasound measurements of bladder volume.

Authors:  N J Dudley; M Kirkland; J Lovett; A R Watson
Journal:  Br J Radiol       Date:  2003-11       Impact factor: 3.039

3.  Assessing the daily consistency of bladder filling using an ultrasonic Bladderscan device in men receiving radical conformal radiotherapy for prostate cancer.

Authors:  S Hynds; C K McGarry; D M Mitchell; S Early; L Shum; D P Stewart; J A Harney; C R Cardwell; J M O'Sullivan
Journal:  Br J Radiol       Date:  2010-12-15       Impact factor: 3.039

4.  Bladder filling variation during radiation treatment of prostate cancer: can the use of a bladder ultrasound scanner and biofeedback optimize bladder filling?

Authors:  Marcel R Stam; Emile N J Th van Lin; Lisette P van der Vight; Johannes H A M Kaanders; Andries G Visser
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-03-15       Impact factor: 7.038

5.  Three-dimensional ultrasound device for rapid determination of bladder volume.

Authors:  L S Marks; F J Dorey; M L Macairan; C Park; J B deKernion
Journal:  Urology       Date:  1997-09       Impact factor: 2.649

6.  Physical and clinical implications of radiotherapy treatment of prostate cancer using a full bladder protocol.

Authors:  Raffaella Cambria; Barbara A Jereczek-Fossa; Dario Zerini; Federica Cattani; Flavia Serafini; Rosa Luraschi; Guido Pedroli; Roberto Orecchia
Journal:  Strahlenther Onkol       Date:  2011-11-25       Impact factor: 3.621

7.  The BladderScan BVI 6200® is not accurate enough for use in a bladder retraining program.

Authors:  Goedele M Beckers; Henricus J R van der Horst; Jos Frantzen; Martijn W Heymans
Journal:  J Pediatr Urol       Date:  2013-01-29       Impact factor: 1.830

8.  Inter-fraction bladder filling variations and time trends for cervical cancer patients assessed with a portable 3-dimensional ultrasound bladder scanner.

Authors:  Rozilawati Ahmad; Mischa S Hoogeman; Sandra Quint; Jan Willem Mens; Ilse de Pree; Ben J M Heijmen
Journal:  Radiother Oncol       Date:  2008-08-12       Impact factor: 6.280

9.  A randomized trial comparing bladder volume consistency during fractionated prostate radiation therapy.

Authors:  Laura M Mullaney; Evelyn O'Shea; Mary T Dunne; Marie A Finn; Pierre G Thirion; Lesley Ann Cleary; Maeve McGarry; Louise O'Neill; John G Armstrong
Journal:  Pract Radiat Oncol       Date:  2014-01-08

10.  Variability of bladder filling in patients receiving radical radiotherapy to the prostate.

Authors:  Una M O'Doherty; Helen A McNair; Andrew R Norman; Elizabeth Miles; Simon Hooper; Mark Davies; Naomi Lincoln; Jan Balyckyi; Peter Childs; David P Dearnaley; Robert A Huddart
Journal:  Radiother Oncol       Date:  2006-06-16       Impact factor: 6.280

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.