Literature DB >> 23375614

The BladderScan BVI 6200® is not accurate enough for use in a bladder retraining program.

Goedele M Beckers1, Henricus J R van der Horst, Jos Frantzen, Martijn W Heymans.   

Abstract

OBJECTIVE: Bladder scans are used extensively in adult urology to estimate urinary volumes. For children, smaller devices have been developed. Scarce literature shows conflicting results regarding the accuracy of measurements in children. The BladderScan(®) BVI 6200 has been developed to overcome some of the inaccuracies. We tested this device to see whether it can replace conventional ultrasound in a bladder retraining program. PATIENTS AND METHODS: 84 children (mean age 7.8 years) were included. Measurements with the BVI 6200 were compared to measurements with a graduated cylinder or with conventional ultrasound before and after micturition. All volumes were expressed in milliliters. The measured volumes were compared to each other in a Bland-Altman analysis.
RESULTS: Mean (bias) of the different measurements fell within 10% difference of measurements. The spread of the different measurements around the mean was however very large, with a broad spread of over- and under-estimations of different measured volumes.
CONCLUSION: The BVI 6200 is not reliable enough to replace conventional ultrasound for measurements of bladder volumes. It is not advisable to use it in a bladder retraining program.
Copyright © 2012 Journal of Pediatric Urology Company. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  BVI 6200; Bladder retraining program; BladderScan; Measurement; Voiding disorder

Mesh:

Year:  2013        PMID: 23375614     DOI: 10.1016/j.jpurol.2012.12.013

Source DB:  PubMed          Journal:  J Pediatr Urol        ISSN: 1477-5131            Impact factor:   1.830


  4 in total

1.  Bladder volume correction factors measured with 3D ultrasound and BladderScan.

Authors:  Naomi N Vinod; Anna S Nagle; Hameeda A Naimi; Hiren Kolli; Derek Sheen; Naveen Nandanan; Laura R Carucci; John E Speich; Adam P Klausner
Journal:  Can J Urol       Date:  2019-08       Impact factor: 1.344

2.  A comparison of bladder volumes based on treatment planning CT and BladderScan® BVI 6100 ultrasound device in a prostate radiation therapy population.

Authors:  Laura Mullaney; Evelyn O'Shea; Mary T Dunne; Pierre G Thirion; John G Armstrong
Journal:  Br J Radiol       Date:  2018-06-27       Impact factor: 3.039

3.  The accuracy of three-dimensional bladder ultrasonography in determining the residual urinary volume compared with conventional catheterisation.

Authors:  Imran K Jalbani; M Hammad Ather
Journal:  Arab J Urol       Date:  2014-07-29

4.  Comparison of 2D and 3D ultrasound methods to measure serial bladder volumes during filling: Steps toward development of non-invasive ultrasound urodynamics.

Authors:  Anna S Nagle; Rachel J Bernardo; Jary Varghese; Laura R Carucci; Adam P Klausner; John E Speich
Journal:  Bladder (San Franc)       Date:  2018-01-04
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.