Performance of the Oxylog® 1000 portable ventilator in a hyperbaric environment.

Lie Sui An1,2, Loy Shun Ting2, Lee Cheng Chuan3, Kim Soo Joang3, Soh Chai Rick3.   

Abstract

INTRODUCTION: The management of mechanically ventilated patients in the hyperbaric environment requires knowledge of how the physical properties of gases change under pressure and how this affects the operation of the ventilator. The primary objective of this study was to test the performance of the Dräger Oxylog 1000® ventilator in a hyperbaric environment.
METHODS: Each of two ventilators was connected to a mechanical test lung system with an in-built pressure gauge. We used a Wright's respirometer to measure the tidal volumes. The same ventilator settings were tested under varying environmental pressures from ambient (101.3 kPa) to 18 meters' sea water (284 kPa) in a multiplace hyperbaric chamber.
RESULTS: A decrease was found in tidal volume, decrease in airway pressure and increase in respiratory rate delivered by the Dräger Oxylog 1000 portable ventilator with increasing pressures to 284 kPa. DISCUSSION: These findings can be explained by the operating principles of the Oxylog 1000, which is a time-controlled, constant-volume ventilator that functions as a flow chopper. Even between the two Oxylog 1000 ventilators tested there were different absolute changes in tidal volume, airway pressures and respiratory rates at various depths. Hence, the trend of changes in these variables is probably more important than absolute values.
CONCLUSION: In summary, understanding the trend of changes in ventilator variables will allow clinicians to make appropriate corrections in ventilator settings and carefully monitor adequacy of ventilation to prevent adverse ventilator-associated events. The Dräger Oxylog 1000 portable ventilator is an adequate back-up ventilator for use with straight-forward, ventilator-dependent patients undergoing hyperbaric treatment. Copyright: This article is the copyright of the authors who grant Diving and Hyperbaric Medicine a non-exclusive licence to publish the article in electronic and other forms.

Entities:  

Keywords:  Equipment; Hyperbaric research; Performance; Pressure; Safety

Mesh:

Year:  2018        PMID: 29888382      PMCID: PMC6156829          DOI: 10.28920/dhm48.2.102-106

Source DB:  PubMed          Journal:  Diving Hyperb Med        ISSN: 1833-3516            Impact factor:   0.887


  8 in total

1.  ANAESTHESIA IN A PRESSURE CHAMBER.

Authors:  D G MCDOWALL
Journal:  Anaesthesia       Date:  1964-07       Impact factor: 6.955

2.  The performance of Dräger Oxylog ventilators at simulated altitude.

Authors:  J G Flynn; B Singh
Journal:  Anaesth Intensive Care       Date:  2008-07       Impact factor: 1.669

3.  The pneuPAC hyperbaric variant HB: a ventilator suitable for use within a one-man hyperbaric chamber.

Authors:  M J Spittal; S J Hunter; L Jones
Journal:  Br J Anaesth       Date:  1991-10       Impact factor: 9.166

4.  Flow resistance of airways under hyperbaric conditions.

Authors:  E Hrncír
Journal:  Physiol Res       Date:  1996       Impact factor: 1.881

Review 5.  Hyperbaric intensive care technology and equipment.

Authors:  Ian L Millar
Journal:  Diving Hyperb Med       Date:  2015-03       Impact factor: 0.887

6.  Evalution of mechanical ventilators in a hyperbaric environment.

Authors:  T J Gallagher; R A Smith; G C Bell
Journal:  Aviat Space Environ Med       Date:  1978-02

7.  A European code of good practice for hyperbaric oxygen therapy.

Authors:  Jacek Kot; Jordi Desola; Antonio Gata Simao; Roly Gough-Allen; Robert Houman; Jean-Luis Meliet; Francois Galland; Christian Mortensen; Peter H J Mueller; Seppo Sippinen
Journal:  Int Marit Health       Date:  2004

8.  Emergency mechanical ventilation at moderate altitude.

Authors:  M Roeggla; G Roeggla; A Wagner; B Eder; A N Laggner
Journal:  Wilderness Environ Med       Date:  1995-08       Impact factor: 1.518