| Literature DB >> 1931410 |
M J Spittal1, S J Hunter, L Jones.
Abstract
We describe the performance of the pneuPAC hyperbaric variant HB, a ventilator designed for use in a one-man hyperbaric chamber. The ventilator delivered minute volumes of 11-23 litre at 1 atm abs to 7.6-16 litre at 2.5 atm abs. The delivered minute volume may be controlled easily from outside the chamber by manipulation of the ventilator rate, 11.5-31 b.p.m.Entities:
Mesh:
Year: 1991 PMID: 1931410 DOI: 10.1093/bja/67.4.488