Literature DB >> 29876514

Knowledge of endoscopic ultrasound-delivered fiducial composition and dimension necessary when planning proton beam radiotherapy.

Ferga C Gleeson1, Erik J Tryggestad2, Nicholas B Remmes2, Chris J Beltran2, Jon J Kruse2, Michael G Haddock2, Christopher L Hallemeier2, Andrew C Storm1, Michael J Levy1.   

Abstract

BACKGROUND AND STUDY AIMS: Little consideration has been given to selection of endoscopic ultrasound-guided fiducials for proton radiotherapy and the resulting perturbations in the therapy dose and pattern. Our aim was to assess the impact of perturbations caused by six fiducials of different composition and dimensions in a phantom gel model.
MATERIALS AND METHODS: The phantom was submerged in a water bath and irradiated with a uniform 10 cm × 10 cm field of 119.7 MeV monoenergetic spot scanning protons delivered through a 45 mm range shifter. The proton "Bragg Peak" was evaluated.
RESULTS: Dose perturbations manifesting as dose reductions up to 30 % were observed. A carbon composite (1 × 5 mm) and gold (0.4 × 10 mm) fiducial with backload potential rather than dedicated EUS pre-loaded gold fiducial needles had the best performance in terms of minimizing the dose perturbation.
CONCLUSIONS: Our data demonstrate that a carbon composite fiducial has a less untoward effect on proton therapy dose distribution than dedicated EUS pre-loaded gold fiducial needles. Such information is important to consider when selecting fiducials specifically for proton therapy.

Entities:  

Year:  2018        PMID: 29876514      PMCID: PMC5988542          DOI: 10.1055/a-0588-4800

Source DB:  PubMed          Journal:  Endosc Int Open        ISSN: 2196-9736


Introduction

Proton radiotherapy (RT) is a form of external beam RT (EBRT) that is of increasing interest, as it is associated with a reduction in normal tissue radiation exposure when compared with conventional photon RT. EBRT in soft tissue regions may be guided radiographically using implanted fiducial markers composed of a high atomic number (Z) material such as gold or other metallic alloys. Ideal markers whether placed by interventional radiology or by endoscopic ultrasound (EUS) should have good radiographic visibility, not migrate, display minimal artifacts on computed tomography or magnetic resonance imaging used for treatment planning and, in particular, not distort the delivered dose of the treatment beam. Conventional photon beams are relatively insensitive to presence of fiducial markers, but the same does not apply to proton beams; the perturbation in dose distribution for proton RT is strongly influenced by fiducial composition 1 2 3 . As the role and utility of proton RT is clinically increasing for gastrointestinal malignancies, interest in EUS-guided fiducial placement for pancreas ductal adenocarcinoma proton RT is gaining momentum. Currently available options include back loading of individual fiducials into an EUS needle and sealing with bone wax or use of newly introduced EUS pre-loaded devices that contain several gold fiducials 4 5 . In the current study, we sought to assess the “worst case” dosimetric impact of perturbations caused by six commercially available carbon composite or gold fiducial markers of different composition and dimensions in a phantom gel model, as potential radiopaque markers for use in proton RT.

Materials and methods

The fiducials with back load potential, were carbon composite or gold measuring 1 × 5 mm (#1), 0.4 × 10 mm (#2) and 0.28 × 10 mm in a clustered arrangement (#3), respectively: all with 22G needle potential. The EUS pre-loaded gold fiducial needles (#4 – #6) were either 0.43 × 5 mm (#4 and #5) via a 22G needle (EchoTip Ultra Fiducial Needle, Cook Medical or Beacon FNF Pre-loaded Needle, Medtronic) or 0.75 × 5 mm (#6) via a 19G needle (Beacon FNF Pre-loaded Needle, Medtronic). A gel phantom containing the aforementioned fiducials placed in both side-on and edge-on orientation relative to the beam axis for all samples except #2 and #3 was submerged in a water bath and irradiated with a uniform 10 cm × 10 cm field of 119.7 MeV monoenergetic spot scanning protons delivered through a 45-mm range shifter. Radiation-sensitive, self-developing Gafchromic EBT3 films (Ashland Advanced Materials, Bridgewater, New Jersey, United States) were placed in a sandwich-stack arrangement incorporating paper spacers of known proton stopping power with approximate water-equivalent film spacing of 1.1 mm and positioned immediately downstream of the water bath. The bath depth was adjusted such that the film stack sampled the region of maximum dose deposition known as the proton “Bragg Peak.”

Results

Dose perturbations manifesting as dose reductions up to 30 % were observed. The magnitude and size of the observed dose reductions are likely clinically relevant. Fig.1 represents both a qualitative and quantitative summary of our findings. We noted an apparent correlation with both fiducial composition and diameter. Edge-on configurations resulted in generally smaller dose perturbations than side-on configurations. As demonstrated by Fig.1e , the side-on 0.75 × 5 mm fiducial (#6) gave the largest perturbation in terms of both magnitude and size. Fiducials #1 and #2 had the best performance in terms of minimizing the dose perturbation. The clustered fiducial (#3) configuration did not compare favorably to #1 and #2. The results for samples #4 and #5 were generally similar in magnitude to #3.
Fig. 1 a

Schematic of experimental setup: a gel phantom containing fiducial samples was submerged in a water bath and irradiated with a uniform 10 × 10 cm 2 field of 119.7 MeV protons using discrete spot scanning. Exiting the water bath, the protons stopped in a paper stack arrangement interleaved with four EBT3 (Gafchromic) films spaced approximately 1.1 mm apart (water-equivalent spacing). b Digitized EBT3 film (16.3 mm behind fiducial plane, closest to the Bragg Peak) identifying the set of fiducial samples included in the present study. Note that an edge-on example for Sample 2 was not incorporated in the gel phantom. The side-on example for Sample 2 is difficult to visualize and therefore black arrows are included as a guide-to-eye (in b , c , and d ). c Dose in cGy for a set of EBT3 films exposed to the end-of-range region, with water-equivalent depths behind the fiducial plane as shown. The small ROIs indicate the regions used for nominal (or background) dose sampling (for normalization) and the large ROIs designate the cropping regions for d . d Dose difference (from background) in % for corresponding cropped film regions from c. e Dose differences observed nearest the Bragg Peak (16.3 mm film) for zoomed (6.4 × 6.4 mm 2 ) ROIs affected by the five given side-on samples (#1, #2, #4 – 6).

Schematic of experimental setup: a gel phantom containing fiducial samples was submerged in a water bath and irradiated with a uniform 10 × 10 cm 2 field of 119.7 MeV protons using discrete spot scanning. Exiting the water bath, the protons stopped in a paper stack arrangement interleaved with four EBT3 (Gafchromic) films spaced approximately 1.1 mm apart (water-equivalent spacing). b Digitized EBT3 film (16.3 mm behind fiducial plane, closest to the Bragg Peak) identifying the set of fiducial samples included in the present study. Note that an edge-on example for Sample 2 was not incorporated in the gel phantom. The side-on example for Sample 2 is difficult to visualize and therefore black arrows are included as a guide-to-eye (in b , c , and d ). c Dose in cGy for a set of EBT3 films exposed to the end-of-range region, with water-equivalent depths behind the fiducial plane as shown. The small ROIs indicate the regions used for nominal (or background) dose sampling (for normalization) and the large ROIs designate the cropping regions for d . d Dose difference (from background) in % for corresponding cropped film regions from c. e Dose differences observed nearest the Bragg Peak (16.3 mm film) for zoomed (6.4 × 6.4 mm 2 ) ROIs affected by the five given side-on samples (#1, #2, #4 – 6).

Discussion

The dosimetric effects of polymer-based or biodegradable esophageal stents is less than that of self-expanding metal stents and attributed mainly to the stent mesh density 6 . Dose perturbation changes have also been demonstrated for fiducials composed of carbon, plastic-coated stainless steel and conventional gold fiducials for patients with prostate cancer undergoing proton RT, favoring carbon composite markers 7 . For endosonographers, there is growing call to place fiducial markers to guide treatment planning and delivery of proton RT. However, there has been little consideration regarding the composition and diameters of fiducials and the resulting perturbations in the dose and pattern of therapy. Resulting decrements in radiation dosage result in a less effective tumor cell death, and conversely secondary dose escalations risk injury to healthy tissues. Our data demonstrate that carbon composite (1 × 5 mm) and gold (0.4 × 10 mm) fiducials, both with needle back loading capabilities, have less untoward effects on proton RT dose distribution than the dedicated EUS pre-loaded gold fiducial needles. Such information is important to consider when working with our radiation oncology partners and selecting fiducials to guide RT and for industry when considering product development.
  7 in total

1.  Dose perturbations and image artifacts caused by carbon-coated ceramic and stainless steel fiducials used in proton therapy for prostate cancer.

Authors:  Joey Cheung; Rajat J Kudchadker; X Ronald Zhu; Andrew K Lee; Wayne D Newhauser
Journal:  Phys Med Biol       Date:  2010-11-12       Impact factor: 3.609

2.  Monte Carlo simulations of the dosimetric impact of radiopaque fiducial markers for proton radiotherapy of the prostate.

Authors:  Wayne Newhauser; Jonas Fontenot; Nicholas Koch; Lei Dong; Andrew Lee; Yuanshui Zheng; Laurie Waters; Radhe Mohan
Journal:  Phys Med Biol       Date:  2007-05-02       Impact factor: 3.609

3.  Microscopic gold particle-based fiducial markers for proton therapy of prostate cancer.

Authors:  Young Kyung Lim; Jungwon Kwak; Dong Wook Kim; Dongho Shin; Myonggeun Yoon; Soah Park; Jin Sung Kim; Sung Hwan Ahn; Jungwook Shin; Se Byeong Lee; Sung Yong Park; Hong Ryeol Pyo; Dae Yong Kim; Kwan Ho Cho
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-08-01       Impact factor: 7.038

4.  Esophageal self-expandable stent material and mesh grid density are the major determining factors of external beam radiation dose perturbation: results from a phantom model.

Authors:  B K Abu Dayyeh; J J Vandamme; R C Miller; T H Baron
Journal:  Endoscopy       Date:  2012-12-19       Impact factor: 10.093

5.  EUS-guided gold fiducial insertion for image-guided radiation therapy of pancreatic cancer: 50 successful cases without fluoroscopy.

Authors:  Walter G Park; Brian M Yan; Devin Schellenberg; Jeff Kim; Daniel T Chang; Albert Koong; Cheryl Patalano; Jacques Van Dam
Journal:  Gastrointest Endosc       Date:  2010-03       Impact factor: 9.427

6.  Dosimetric impact of fiducial markers in patients undergoing photon beam radiation therapy.

Authors:  Oleg N Vassiliev; Rajat J Kudchadker; Deborah A Kuban; Steven J Frank; Seungtaek Choi; Quynh Nguyen; Andrew K Lee
Journal:  Phys Med       Date:  2011-08-27       Impact factor: 2.685

7.  Comparative analysis of traditional and coiled fiducials implanted during EUS for pancreatic cancer patients receiving stereotactic body radiation therapy.

Authors:  Mouen A Khashab; Katherine J Kim; Erik J Tryggestad; Aaron T Wild; Teboh Roland; Vikesh K Singh; Anne Marie Lennon; Eun Ji Shin; Mark A Ziegler; Reem Z Sharaiha; Marcia Irene Canto; Joseph M Herman
Journal:  Gastrointest Endosc       Date:  2012-11       Impact factor: 9.427

  7 in total
  2 in total

1.  Endoscopic Ultrasound-Guided Fiducial Placement for Stereotactic Body Radiation Therapy in Pancreatic Malignancy.

Authors:  Seong-Hun Kim; Eun Ji Shin
Journal:  Clin Endosc       Date:  2021-05-28

2.  Experimental Comparison of Fiducial Markers Used in Proton Therapy: Study of Different Imaging Modalities and Proton Fluence Perturbations Measured With CMOS Pixel Sensors.

Authors:  Claire-Anne Reidel; Felix Horst; Christoph Schuy; Oliver Jäkel; Swantje Ecker; Katrin Henkner; Stephan Brons; Marco Durante; Uli Weber
Journal:  Front Oncol       Date:  2022-03-25       Impact factor: 6.244

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.