A comprehensive theoretical comparison of proton imaging set-ups in terms of spatial resolution.

N Krah1, F Khellaf, J M Létang, S Rit, I Rinaldi.   

Abstract

We present a comprehensive analytical comparison of four types of proton imaging set-ups and, to this end, develop a mathematical framework to calculate the width of the uncertainty envelope around the most likely proton path depending on set-up geometry, detector properties, and proton beam parameters. As a figure of merit for the spatial resolution achievable with each set-up, we use the frequency [Formula: see text] at which the modular transfer function of a density step decreases below 10%. We verify the analytical results with Monte Carlo simulations. We find that set-ups which track the angle and position of individual protons in front of and behind the phantom would yield an average spatial resolution of 0.3-0.35 lp mm-1 assuming realistic geometric parameters (i.e. 30-40 cm distance between detector and phantom, 15-20 cm phantom thickness). For set-ups combining pencil beam scanning with either a position sensitive detector, e.g. an x-ray flat panel, or with a position insensitive detector, e.g. a range telescope, we find an average spatial resolution of about 0.1 lp mm-1 for an 8 mm FWHM beam spot size. The pixel information improves the spatial resolution by less than 10%. In both set-up types, performance can be significantly improved by reducing the pencil beam size down to 2 mm FWHM. In this case, the achievable spatial resolution reaches about 0.25 lp mm-1. Our results show that imaging set-ups combining double scattering with a pixel detector can provide sufficient spatial resolution only under very stringent conditions and are not ideally suited for computed tomography applications. We further propose a region-of-interest method for set-ups with a pixel detector to filter out protons which have undergone nuclear reactions and discuss the impact of tracker detector uncertainties on the most likely path.

Mesh:

Substances:

Year:  2018        PMID: 29864022     DOI: 10.1088/1361-6560/aaca1f

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  The accuracy of helium ion CT based particle therapy range prediction: an experimental study comparing different particle and x-ray CT modalities.

Authors:  L Volz; C-A Collins-Fekete; E Bär; S Brons; C Graeff; R P Johnson; A Runz; C Sarosiek; R W Schulte; J Seco
Journal:  Phys Med Biol       Date:  2021-11-29       Impact factor: 3.609

Review 2.  Latest developments in in-vivo imaging for proton therapy.

Authors:  Katia Parodi
Journal:  Br J Radiol       Date:  2019-12-12       Impact factor: 3.039

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.