Simple and ultra-fast recognition and quantitation of compounded monoclonal antibodies: Application to flow injection analysis combined to UV spectroscopy and matching method.

E Jaccoulet1, A Schweitzer-Chaput2, B Toussaint2, P Prognon3, E Caudron3.   

Abstract

Compounding of monoclonal antibody (mAbs) constantly increases in hospital. Quality control (QC) of the compounded mAbs based on quantification and identification is required to prevent potential errors and fast method is needed to manage outpatient chemotherapy administration. A simple and ultra-fast (less than 30 s) method using flow injection analysis associated to least square matching method issued from the analyzer software was performed and evaluated for the routine hospital QC of three compounded mAbs: bevacizumab, infliximab and rituximab. The method was evaluated through qualitative and quantitative parameters. Preliminary analysis of the UV absorption and second derivative spectra of the mAbs allowed us to adapt analytical conditions according to the therapeutic range of the mAbs. In terms of quantitative QC, linearity, accuracy and precision were assessed as specified in ICH guidelines. Very satisfactory recovery was achieved and the RSD (%) of the intermediate precision were less than 1.1%. Qualitative analytical parameters were also evaluated in terms of specificity, sensitivity and global precision through a matrix of confusion. Results showed to be concentration and mAbs dependant and excellent (100%) specificity and sensitivity were reached within specific concentration range. Finally, routine application on "real life" samples (n = 209) from different batch of the three mAbs complied with the specifications of the quality control i.e. excellent identification (100%) and ± 15% of targeting concentration belonging to the calibration range. The successful use of the combination of second derivative spectroscopy and partial least square matching method demonstrated the interest of FIA for the ultra-fast QC of mAbs after compounding using matching method.
Copyright © 2018 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Flow injection analysis; Matching method; Monoclonal antibody; Qualitative control; Quantitative control; Second derivative UV spectroscopy

Mesh:

Substances:

Year:  2018        PMID: 29853048     DOI: 10.1016/j.talanta.2018.05.042

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  2 in total

1.  Quality control of cytostatic drug preparations-comparison of workflow and performance of Raman/UV and high-performance liquid chromatography coupled with diode array detection (HPLC-DAD).

Authors:  Lars M H Reinders; Martin D Klassen; Claudia Vom Eyser; Thorsten Teutenberg; Martin Jaeger; Torsten C Schmidt; Jochen Tuerk
Journal:  Anal Bioanal Chem       Date:  2021-02-23       Impact factor: 4.142

2.  Comprehensive Analysis of Nivolumab, A Therapeutic Anti-Pd-1 Monoclonal Antibody: Impact of Handling and Stress.

Authors:  Anabel Torrente-López; Jesús Hermosilla; Antonio Salmerón-García; José Cabeza; Natalia Navas
Journal:  Pharmaceutics       Date:  2022-03-23       Impact factor: 6.525

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.