A novel cause of rebreathing carbon dioxide related to removed CLIC-seal on the Dräger Apollo© anesthesia machine.

B Niknam1, Z Bebic2, A Roseman2.   

Abstract

We present a case report involving two sequential, surgically uneventful, laparoscopic cholecystectomies using the same anesthesia machine (Drager Apollo©) for which the level of inspired carbon dioxide was noted to be elevated following various diagnostic interventions including replacing the sodalime, increasing fresh gas flows, and a full inspection of equipment for malfunction. Eventually it was discovered that a rubber ring seal connecting the Dragersorb CLIC system© to the sodalime canister was inadvertently removed during the initial canister exchange resulting in an apparent bypassing of the absorbent and thus an inability of the exhaled gas to contact the sodalime. To our knowledge this is the first such description of this potential cause of elevated inspired carbon dioxide and should warrant consideration when other conventional interventions have failed.

Entities:  

Keywords:  Dragersorb CLIC; Inspired carbon dioxide; Rebreathing; Sodalime

Mesh:

Substances:

Year:  2018        PMID: 29804264     DOI: 10.1007/s10877-018-0161-0

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  1 in total

1.  Carbon dioxide rebreathing caused by deformed silicon leaflet in the expiratory unidirectional valve.

Authors:  Arumugam Vasudevan; Satyen Parida; Lenin Babu Elakkumanan; Sandeep Mishra
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2013-01
  1 in total