Arumugam Vasudevan1, Satyen Parida, Lenin Babu Elakkumanan, Sandeep Mishra.
Abstract
Rebreathing of carbon dioxide caused by incompetent 'cage and disc' unidirectional valves has been reported earlier. Some manufacturers have changed the design of unidirectional valves to 'flexible leaflets'. We report a series of cases where a deformed membrane leaflet in expiratory unidirectional valves led to rebreathing of carbon dioxide.Entities:
Keywords: Anesthesia machine; circle breathing system; incompetent unidirectional valves
Year: 2013 PMID: 23493853 PMCID: PMC3590513 DOI: 10.4103/0970-9185.105816
Source DB: PubMed Journal: J Anaesthesiol Clin Pharmacol ISSN: 0970-9185
Figure 1Capnogram showing elevated baseline and higher end-tidal carbon dioxide
Figure 2Deformed ‘flexible leaflet’ of expiratory unidirectional valve (shown after removing the covering dome)
Figure 3Ventilator display showing the abnormal reverse flow through expiratory unidirectional valve during inspiratory phase (I) and the normal flow during expiration (E)