Evaluation of a Relative Afferent Pupillary Defect using the RAPDx® Device Before and After Treatment in Patients with Optic Nerve Disease.

Tsukasa Satou1, Hitoshi Ishikawa2,3, Toshiaki Goseki2, Ken Asakawa3.   

Abstract

We evaluated the amplitude and latency scores in the RAPDx® device together with other ophthalmic examinations, before and after treatment in four patients with optic nerve disease. In all patients, the visual acuity (VA) and visual field (VF) after treatment was resolved. Both scores after treatment were lower, with reduced laterality-based differences in VA and critical flicker fusion frequency (CFF). Even after treatment, 3 patients had laterality-based differences in circumpapillary retinal nerve fiber layer thickness (cpRFNLT). Both scores for evaluation of RAPD by RAPDx® correlated with subjective examinations and were useful for evaluation of the efficacy of treatment.

Entities:  

Keywords:  Optic nerve disease; relative afferent pupillary defect; treatment

Year:  2017        PMID: 29796046      PMCID: PMC5958953          DOI: 10.1080/01658107.2017.1380050

Source DB:  PubMed          Journal:  Neuroophthalmology        ISSN: 0165-8107


  5 in total

1.  Evaluation of Relative Afferent Pupillary Defect Using RAPDx Device in Patients with Optic Nerve Disease.

Authors:  Tsukasa Satou; Hitoshi Ishikawa; Ken Asakawa; Toshiaki Goseki; Takahiro Niida; Kimiya Shimizu
Journal:  Neuroophthalmology       Date:  2016-04-14

2.  How to measure the relative afferent pupillary defect.

Authors:  H S Thompson; J J Corbett; T A Cox
Journal:  Surv Ophthalmol       Date:  1981 Jul-Aug       Impact factor: 6.048

3.  Retinal nerve fiber structure versus visual field function in patients with ischemic optic neuropathy. A test of a linear model.

Authors:  Donald C Hood; Susan Anderson; Jacinthe Rouleau; Adam S Wenick; Larissa K Grover; Myles M Behrens; Jeffrey G Odel; Andrew G Lee; Randy H Kardon
Journal:  Ophthalmology       Date:  2007-09-17       Impact factor: 12.079

4.  Symmetry of the pupillary light reflex and its relationship to retinal nerve fiber layer thickness and visual field defect.

Authors:  Dolly S Chang; Michael V Boland; Karun S Arora; Wasu Supakontanasan; Bei Bei Chen; David S Friedman
Journal:  Invest Ophthalmol Vis Sci       Date:  2013-08-19       Impact factor: 4.799

5.  Effects of Age and Sex on Values Obtained by RAPDx® Pupillometer, and Determined the Standard Values for Detecting Relative Afferent Pupillary Defect.

Authors:  Tsukasa Satou; Toshiaki Goseki; Ken Asakawa; Hitoshi Ishikawa; Kimiya Shimizu
Journal:  Transl Vis Sci Technol       Date:  2016-04-28       Impact factor: 3.283