Literature DB >> 29793851

Evaluation of Examiner Performance Using a Duplex Ultrasound Simulator. Flow Velocity Measurements in Dialysis Access Fistula Models.

Daniel F Leotta1, R Eugene Zierler2, Kurt Sansom3, Alberto Aliseda3, Mark D Anderson4, Florence H Sheehan4.   

Abstract

We developed a duplex ultrasound simulator for training and assessment of scanning skills. We used the simulator to test examiner performance in the measurement of flow velocities in dialysis access fistulas. Test cases were created from 3-D ultrasound scans of two dialysis access fistulas by reconstructing 3-D blood vessel models and simulating blood flow velocity fields within the lumens. The simulator displays a 2-D B-mode or color Doppler image corresponding to transducer position on a mannequin; a spectral waveform is generated according to Doppler sample volume location and system settings. Examiner performance was assessed by comparing the measured peak systolic velocity (PSV) with the true PSV provided by the computational flow model. The PSV measured by four expert examiners deviated from the true value by 7.8 ± 6.1%. The results indicate the ability of the simulator to objectively assess an examiner's measurement accuracy in complex vascular targets.
Copyright © 2018 World Federation for Ultrasound in Medicine and Biology. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Color Doppler; Dialysis access fistula; Doppler spectral waveform; Duplex ultrasound; Medical simulation; Peak systolic velocity

Mesh:

Year:  2018        PMID: 29793851      PMCID: PMC6026548          DOI: 10.1016/j.ultrasmedbio.2018.04.012

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  20 in total

1.  Serial measurement of cross-sectional area in peripheral vein grafts using three-dimensional ultrasound.

Authors:  D F Leotta; J F Primozich; K W Beach; R O Bergelin; D E Strandness
Journal:  Ultrasound Med Biol       Date:  2001-01       Impact factor: 2.998

2.  Carotid Doppler velocity measurements and anatomic stenosis: correlation is futile.

Authors:  Kirk W Beach; Daniel F Leotta; R Eugene Zierler
Journal:  Vasc Endovascular Surg       Date:  2012-07-10       Impact factor: 1.089

3.  Training and simulation for patient safety.

Authors:  Rajesh Aggarwal; Oliver T Mytton; Milliard Derbrew; David Hananel; Mark Heydenburg; Barry Issenberg; Catherine MacAulay; Mary Elizabeth Mancini; Takeshi Morimoto; Nathaniel Soper; Amitai Ziv; Richard Reznick
Journal:  Qual Saf Health Care       Date:  2010-08

Review 4.  Technology-enhanced simulation to assess health professionals: a systematic review of validity evidence, research methods, and reporting quality.

Authors:  David A Cook; Ryan Brydges; Benjamin Zendejas; Stanley J Hamstra; Rose Hatala
Journal:  Acad Med       Date:  2013-06       Impact factor: 6.893

5.  Measurement of abdominal aortic aneurysms with three-dimensional ultrasound imaging: preliminary report.

Authors:  D F Leotta; M Paun; K W Beach; T R Kohler; R E Zierler; D E Strandness
Journal:  J Vasc Surg       Date:  2001-04       Impact factor: 4.268

6.  A longitudinal study of remodeling in a revised peripheral artery bypass graft using 3D ultrasound imaging and computational hemodynamics.

Authors:  Patrick M McGah; Daniel F Leotta; Kirk W Beach; James J Riley; Alberto Aliseda
Journal:  J Biomech Eng       Date:  2011-04       Impact factor: 2.097

7.  Standardized ultrasound evaluation of carotid stenosis for clinical trials: University of Washington Ultrasound Reading Center.

Authors:  Kirk W Beach; Robert O Bergelin; Daniel F Leotta; Jean F Primozich; P Max Sevareid; Edward T Stutzman; R Eugene Zierler
Journal:  Cardiovasc Ultrasound       Date:  2010-09-07       Impact factor: 2.062

8.  Development of a Duplex Ultrasound Simulator and Preliminary Validation of Velocity Measurements in Carotid Artery Models.

Authors:  R Eugene Zierler; Daniel F Leotta; Kurt Sansom; Alberto Aliseda; Mark D Anderson; Florence H Sheehan
Journal:  Vasc Endovascular Surg       Date:  2016-05-11       Impact factor: 1.089

9.  Incomplete restoration of homeostatic shear stress within arteriovenous fistulae.

Authors:  Patrick M McGah; Daniel F Leotta; Kirk W Beach; R Eugene Zierler; Alberto Aliseda
Journal:  J Biomech Eng       Date:  2013-01       Impact factor: 2.097

10.  Disaster response team FAST skills training with a portable ultrasound simulator compared to traditional training: pilot study.

Authors:  Michael T Paddock; John Bailitz; Russ Horowitz; Basem Khishfe; Karen Cosby; Michelle J Sergel
Journal:  West J Emerg Med       Date:  2015-03-06
View more
  3 in total

1.  Robust and durable aberrative and absorptive phantom for therapeutic ultrasound applications.

Authors:  Alex T Peek; Gilles P L Thomas; Daniel F Leotta; Petr V Yuldashev; Vera A Khokhlova; Tatiana D Khokhlova
Journal:  J Acoust Soc Am       Date:  2022-05       Impact factor: 2.482

2.  Treating Porcine Abscesses with Histotripsy: A Pilot Study.

Authors:  Thomas J Matula; Yak-Nam Wang; Tatiana Khokhlova; Daniel F Leotta; John Kucewicz; Andrew A Brayman; Matthew Bruce; Adam D Maxwell; Brian E MacConaghy; Gilles Thomas; Valery P Chernikov; Sergey V Buravkov; Vera A Khokhlova; Keith Richmond; Keith Chan; Wayne Monsky
Journal:  Ultrasound Med Biol       Date:  2020-11-26       Impact factor: 2.998

3.  Sonographic Features of Abscess Maturation in a Porcine Model.

Authors:  Daniel F Leotta; Matthew Bruce; Yak-Nam Wang; John Kucewicz; Tatiana Khokhlova; Keith Chan; Wayne Monsky; Thomas J Matula
Journal:  Ultrasound Med Biol       Date:  2021-04-24       Impact factor: 3.694

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.