Abstract

A monolithic measurement platform was implemented to enable label-free in-vitro electrical impedance spectroscopy measurements of cells on multi-functional CMOS microelectrode array. The array includes 59,760 platinum microelectrodes, densely packed within a 4.5 mm × 2.5 mm sensing region at a pitch of 13.5 μm. The 32 on-chip lock-in amplifiers can be used to measure the impedance of any arbitrarily chosen electrodes on the array by applying a sinusoidal voltage, generated by an on-chip waveform generator with a frequency range from 1 Hz to 1 MHz, and measuring the respective current. Proof-of-concept measurements of impedance sensing and imaging are shown in this paper. Correlations between cell detection through optical microscopy and electrochemical impedance scanning were established.

Entities:  

Keywords:  CMOS Microelectrode Array (MEA); Electrochemical Impedance Spectroscopy; Electrophysiology; Lock-in Amplifier

Year:  2017        PMID: 29780437      PMCID: PMC5955208          DOI: 10.1109/ICSENS.2016.7808761

Source DB:  PubMed          Journal:  Proc IEEE Sens        ISSN: 1930-0395


  4 in total

1.  A Multi-Modality CMOS Sensor Array for Cell-Based Assay and Drug Screening.

Authors:  Taiyun Chi; Jong Seok Park; Jessica C Butts; Tracy A Hookway; Amy Su; Chengjie Zhu; Mark P Styczynski; Todd C McDevitt; Hua Wang
Journal:  IEEE Trans Biomed Circuits Syst       Date:  2016-01-22       Impact factor: 3.833

2.  A CMOS Electrochemical Impedance Spectroscopy (EIS) Biosensor Array.

Authors:  Arun Manickam; Aaron Chevalier; Mark McDermott; Andrew D Ellington; Arjang Hassibi
Journal:  IEEE Trans Biomed Circuits Syst       Date:  2010-12       Impact factor: 3.833

3.  Label-free impedimetric immunosensor for ultrasensitive detection of cancer marker Murine double minute 2 in brain tissue.

Authors:  Reda Elshafey; Chaker Tlili; Abedelnasser Abulrob; Ana C Tavares; Mohammed Zourob
Journal:  Biosens Bioelectron       Date:  2012-08-03       Impact factor: 10.618

4.  A 200-Channel Area-Power-Efficient Chemical and Electrical Dual-Mode Acquisition IC for the Study of Neurodegenerative Diseases.

Authors:  Jing Guo; Waichiu Ng; Jie Yuan; Suwen Li; Mansun Chan
Journal:  IEEE Trans Biomed Circuits Syst       Date:  2015-10-28       Impact factor: 3.833

  4 in total
  1 in total

1.  Impedance Spectroscopy and Electrophysiological Imaging of Cells With a High-Density CMOS Microelectrode Array System.

Authors:  Vijay Viswam; Raziyeh Bounik; Amir Shadmani; Jelena Dragas; Cedar Urwyler; Julia Alicia Boos; Marie Engelene J Obien; Jan Muller; Yihui Chen; Andreas Hierlemann
Journal:  IEEE Trans Biomed Circuits Syst       Date:  2018-11-12       Impact factor: 3.833

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.