| Literature DB >> 29780288 |
Chiara Mascoli1, Gianluca Faggioli1, Enrico Gallitto1, Vincenzo Vento1, Giuseppe Indelicato1, Rodolfo Pini1, Andrea Vacirca1, Andrea Stella1, Mauro Gargiulo1.
Abstract
Introduction: Iodinated contrast media completion angiography (ICM-A) may underestimate the presence of type II endoleak (ELII) after endovascular aortic repair (EVAR), particularly if they are at low flow. Contrast-enhanced ultrasound (CEUS) has been proposed as the gold standard in ELII detection during EVAR follow-up. Intraprocedural carbon dioxide (CO2) angiography has been shown to be useful in this setting; however no comparative studies including these three techniques are currently available. Our aim was to investigate the accuracy of a new automated CO2 angiographic (CO2-A) system in the detection of ELII, by comparing it with ICM-A and CEUS.Entities:
Mesh:
Substances:
Year: 2018 PMID: 29780288 PMCID: PMC5892285 DOI: 10.1155/2018/7647165
Source DB: PubMed Journal: Contrast Media Mol Imaging ISSN: 1555-4309 Impact factor: 3.161
Figure 1Angiodroid injection system that shows CO2 injection volume and pressure on the display.
Interpretation for the Cohen k coefficient.
|
| Agreement |
|---|---|
| 0 | Absent |
| 0.01–0.20 | Slight |
| 0.21–0.40 | Poor |
| 0.41–0.60 | Moderate |
| 0.61–0.80 | Substantial |
| 0.81–0.99 | Almost perfect |
| 1 | Perfect |
Demographic and clinical characteristics.
|
| % | |
|---|---|---|
| Demographic | ||
| Age (yrs) | 76.1 | 6.7 |
| Sex (male) | 21 | 100 |
|
| ||
| Clinical | ||
| HPT | 20 | 95.2 |
| COPD | 7 | 33.3 |
| CAD | 4 | 19.0 |
| Atrial fibrillation | 3 | 14.3 |
| Dyslipidemia | 13 | 61.9 |
| CRF (GFR < 60 mL/min/1.73 m2) | 7 | 33.3 |
| Dialysis | 2 | 9.5 |
| Smoke | 15 | 71.4 |
| Obesity | 3 | 14.3 |
| Peripheral artery disease | 2 | 9.5 |
Mean and standard deviation (SD); data are expressed as mean and standard deviation (SD) or number and percentage. HPT: hypertension; COPD: chronic obstructive pulmonary disease; CAD: coronary artery disease; CRF: chronic renal failure expressed as glomerular filtration rate (GFR) < 60 mL/min/1.73 m2.
Intraoperative data.
| Intraoperative data |
| % |
|---|---|---|
| Procedure time (m) | 173.3 | (33.1) |
| Endograft | ||
| Suprarenal fixation | 11 | 52.4 |
| Infrarenal fixation | 10 | 47.6 |
| Fluoroscopy time (m) | 19.3 | (5.3) |
| DAP (mGy | 19.0 | (16.7) |
| General anesthesia (n°) | 8 | 38.1 |
| Spinal anaesthesia (n°) | 13 | 61.9 |
| AAA-sac embolization | 10 | 47.6 |
Mean and standard deviation (DS); DAP: dose-area product (mGy∗m2); AAA-sac embolization: abdominal aortic aneurysm intraoperative sac embolization.
Figure 2(a) Iodinated contrast media completion angiography that shows the good positioning of the infrarenal fixation endograft and the absence of endoleaks. (b) Carbon dioxide completion angiography that shows the good positioning of the infrarenal fixation endograft and the presence of ELII. (c) Magnification of ELII from sacral artery (red arrows indicates ELII coming from sacral artery). (d) Contrast-enhanced ultrasound that shows the presence of ELII with the inflow from sacral artery (red arrow).
Figure 3(a) Iodinate contrast media completion angiography that shows the good positioning of the suprarenal fixation endograft and the absence of endoleaks. (b) Carbon dioxide completion angiography that shows the good positioning of the suprarenal fixation endograft and the presence of ELII (as indicated by the red arrows). (c) Contrast-enhanced ultrasound that shows the presence of ELII with the inflow from inferior mesenteric artery (as indicated by the red arrow).
Type II endoleak detected by CO2-A, ICM-A, and CEUS.
| ID | CO2-A | ICM-A | CEUS |
|---|---|---|---|
| # 5 | ✓ | − | ✓ |
| # 9 | ✓ | − | − |
| # 14 | ✓ | − | ✓ |
| # 16 | ✓ | − | − |
| # 17 | ✓ | ✓ | ✓ |
| # 18 | ✓ | − | − |
| # 21 | ✓ | − | ✓ |
ID: patient series number; ✓: type II endoleak presence; −: type II endoleak absence.
Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of CO2-A for type II endoleak detection during EVAR.
| Endoleak | Sensitivity | Specificity | PPV | NPV |
|---|---|---|---|---|
| Type I | 1 | 1 | 1 | 1 |
| Type II | 1 | 0.82 | 0.57 | 1 |
| Type III | 1 | 1 | 1 | 1 |
Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of ICM-A for type II endoleak detection during EVAR.
| Endoleak | Sensitivity | Specificity | PPV | NPV |
|---|---|---|---|---|
| Type I | 1 | 1 | 1 | 1 |
| Type II | 0.25 | 1 | 1 | 0.85 |
| Type III | 1 | 1 | 1 | 1 |