Qualitative study of mechanical parameters of conventional diagnostic X-ray machines in Mizoram.

Jonathan Lalrinmawia1, Kham Suan Pau2, Ramesh Chandra Tiwari3.   

Abstract

The present study examined the mechanical attributes of 135 conventional diagnostic X-ray machines in Mizoram, India. The purpose of studying the X-ray mechanical parameters, such as congruency, perpendicularity of the central beam, and half-value layer, was to improve the quality of the diagnostic image and reduce the patient dose. A battery-operated portable dosimeter was used to measure output radiation of the X-ray machine. The half-value layer was measured at a constant accelerating potential of 70 kVp and tube load. To measure the congruency and beam alignment perpendicularity, a congruence and alignment tool was used. The survey data were collected between June 2015 and June 2016. The authors followed international standard test procedures, and the results were compared to national and international standards. SPSS Statistics for Windows, Version 17 was used to calculate the mean, range, and standard deviation. The half-value layer ranged from 0.45 to 3.00 mm; the mean half-value layer was 1.60 ± 0.51 SD mm. In comparison with national and international standards, only 27.83% (national) and 15.64% (international) of the machines' filtration were found to be within acceptable limits. The congruence misalignment of the x-axis varied between 0.50% and 15.30% of the source-to-image distance; for the y-axis, it ranged from 0.50 to 10.90%. When the congruence between the radiation beam and optical field was tested, 80.85% of diagnostic X-ray machines did not meet the prescribed acceptance parameters. When the perpendicularity between the central beam and the image receptor was tested, 69.81% did not meet safety standards.

Entities:  

Keywords:  Central beam perpendicularity; Congruence between radiation and optical field; Conventional diagnostic X-ray; Half-value layer; Radiation safety

Mesh:

Year:  2018        PMID: 29777432     DOI: 10.1007/s12194-018-0464-3

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  8 in total

Review 1.  Interaction of ionizing radiation with matter.

Authors:  James E Turner
Journal:  Health Phys       Date:  2005-06       Impact factor: 1.316

2.  Measuring scatter radiation in diagnostic X rays for radiation protection purposes.

Authors:  Ioannis Vlachos; Xenophon Tsantilas; Nektarios Kalyvas; Harry Delis; Ioannis Kandarakis; George Panayiotakis
Journal:  Radiat Prot Dosimetry       Date:  2015-04-24       Impact factor: 0.972

Review 3.  Radiation risks: what is to be done?

Authors:  Walter Huda
Journal:  AJR Am J Roentgenol       Date:  2015-01       Impact factor: 3.959

4.  Radiological safety status and quality assurance audit of medical X-ray diagnostic installations in India.

Authors:  A U Sonawane; Meghraj Singh; J V K Sunil Kumar; Arti Kulkarni; V K Shirva; A S Pradhan
Journal:  J Med Phys       Date:  2010-10

5.  Patient dose measurement in common medical X-ray examinations in Iran.

Authors:  Behrouz Rasuli; Ali Mahmoud-Pashazadeh; Mohammad Ghorbani; Raheleh Tabari Juybari; Mozafar Naserpour
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

6.  Assessment of radiation protection awareness and knowledge about radiological examination doses among Italian radiographers.

Authors:  F Paolicchi; F Miniati; L Bastiani; L Faggioni; A Ciaramella; I Creonti; C Sottocornola; C Dionisi; D Caramella
Journal:  Insights Imaging       Date:  2015-11-23

7.  Diagnostic X-ray facilities as per quality control performances in Tanzania.

Authors:  Yesaya Yohana Sungita
Journal:  J Appl Clin Med Phys       Date:  2006-11-28       Impact factor: 2.102

8.  Evolution of X-ray machine quality control acceptance indices.

Authors:  Maria Lucia Nana I Ebisawa; Maria de Fatima A Magon; Yvone M Mascarenhas
Journal:  J Appl Clin Med Phys       Date:  2009-10-07       Impact factor: 2.102

  8 in total
  1 in total

1.  Evaluation of radiation doses at diagnostic X-ray control panels and outside patient entrance doors in Aizawl district, India.

Authors:  Jonathan Lalrinmawia; Kham Suan Pau; Ramesh Chandra Tiwari
Journal:  Radiol Phys Technol       Date:  2019-08-06
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.