| Literature DB >> 29761017 |
Elke Pucks-Faes1, Gabriel Hitzenberger1, Heinrich Matzak1, Elena Fava1, Giulio Verrienti1, Ilse Laimer2, Josef Fritz3, Leopold Saltuari1,4.
Abstract
Objective: Treatment with intrathecal baclofen (ITB) is commonly used in patients with severe spasticity. However, complications may occur after implantation of the ITB-device, albeit mainly procedure- and device-related problems. The aim of the study was to assess surgical- as well as catheter- and pump-related complications and define their risk factors.Entities:
Keywords: autonomic nervous system diseases; baclofen; brain injuries; sympathetic nervous system; traumatic
Mesh:
Substances:
Year: 2018 PMID: 29761017 PMCID: PMC5943769 DOI: 10.1002/brb3.965
Source DB: PubMed Journal: Brain Behav Impact factor: 2.708
Demographic patients’ characteristics and further clinical information
| Patient characteristics | |
| Age at implantation (mean ± standard deviation; years) | 39 ± 16.7 |
| Sex (women/men; | 39/77 |
| Pathologic disorders | |
| Supraspinal [ | |
| Traumatic brain injury | 34/29% |
| Cerebral hypoxia | 16/14% |
| Cerebral palsy | 7/6% |
| Intracerebral hemorrhage | 6/5% |
| Ischemic stroke | 4/3% |
| Subarachnoidal hemorrhage | 3/3% |
| Stiff person syndrome | 1/1% |
| Extra‐/pontine myelinolysis | 1/1% |
| Unknown neurodegenerative disease | 1/1% |
| Spinal [ | |
| Traumatic spinal injury | 18/15% |
| Multiple sclerosis | 14/12% |
| Spastic paralysis | 7/6% |
| Disc herniatio | 1/1% |
| Ischemic lesion | 1/1% |
| Myelopathy due to extradural granuloma | 1/1% |
| Spastic paresis of unclear etiology | 1/1% |
| Investigated parameters | |
| Median modified Rankin Scale (IQR) | 5 (4‐5) |
| Median Barthel Index (IQR) | 0 (0‐54) |
| Median ITB‐dosage (IQR) [μg/day] | 200 (129‐320) |
| Median ITB‐concentration (IQR) [μg/ml] | 2,000 (500‐2,000) |
| Median flow rate (IQR) [μl/hr] | 6.3 (3.5‐10.4) |
| Median Ashworth Scale (IQR) | 4 (3‐4) |
| Catheter type [Ascenda®, Model 8,781/other types; | 39/77 |
IQR, interquartile range; ITB, intrathecal baclofen.
The medians of the investigated parameter were calculated on basis of the patient population.
Clinical data and information on investigated risk factors of the 29 patients with 32 complications
| No | Pathology | Complication | Time to event | Imaging | Procedure | Median modified Rankin Scale | Median barthel index | Median flow rate [μl/hr] | Median ashworth scale | Catheter |
|---|---|---|---|---|---|---|---|---|---|---|
| 1 | SP | Catheter dislocation of distal segment | 64 | XR: dislocation | Exchange of entire catheter | 2 | 95 | 9.2 | 4 | 8731SC |
| 2 | TBI | Catheter dislocation of distal segment | 914 | CT: dislocation | Exchange of entire catheter | 5 | 0 | 5.8 | 4 | 8731SC |
| 3 | TSI | CSF leakage (early) | 24 | No imaging | Patching | 3 | 80 | 25.4 | 4 | 8731SC |
| 4 | Hypoxia | CSF leakage (late) | 1,822 | No imaging | Exchange of entire system | 4 | 60 | 18.6 | 4 | 8709/8596 |
| 5 | Disc herniation | Catheter disconnection | 737 | No imaging | Exchange of entire catheter | 4 | 65 | 10.8 | 0 | 8709/8596 |
| 6 | Disc herniation | Pump site infection | 884 | No imaging | Exchange of pump | 4 | 65 | 5.1 | 0 | 8709/8596 |
| 7 | ICH | Pump site infection | 168 | No imaging | Explantation of pump | 4 | 45 | 6.7 | 4 | 8731SC |
| 8 | TBI | Catheter kinking | 19 | XR+CA: no signs | No revision | 4 | 0 | 4.6 | 4 | 8731SC |
| 9 | Spinal ischemia | Unknown catheter dysfunction | 2,471 | Fluoroscopy: No signs | Exchange of entire system | 4 | 30 | 7.1 | 4 | 8731SC |
| 10 | TBI | Catheter kinking | 115 | XR+CA: no signs | Exchange of distal segment | 5 | 0 | 22.9 | 5 | 8711 |
| 11 | TSI | Catheter dislocation of distal segment | 32 | No imaging | Exchange of entire catheter | 4 | 75 | 19.2 | 3 | 8731SC |
| 12 | TSI | Catheter dislocation of distal segment | 28 | XR: no signs | Exchange of entire catheter | 4 | 75 | 19.2 | 3 | 8731SC |
| 13 | MS | Catheter kinking | 364 | XR & CT+CA: no signs | Exchange of entire catheter | 5 | 20 | 2.7 | 4 | 8781 |
| 14 | TSI | Catheter dislocation of distal segment | 657 |
XR: catheter at L3 | Exchange of distal segment | 4 | 75 | 2.7 | 1 | 8731SC |
| 15 | CP | Pump site infection | 26 | No imaging | Exchange of entire system | 5 | 0 | 4.2 | 4 | 8781 |
| 16 | Hypoxia | Pump site infection | 368 | No imaging | Exchange of entire system | 5 | 0 | 8.3 | 5 | 8731SC |
| 17 | SAH | Catheter dislocation of distal segment | 25 |
CT+CA: dislocation | Exchange of distal segment | 5 | 0 | 3.1 | 4 | 8711 |
| 18 | ICH | Catheter disconnection | 2,470 | XR: disconnection | Exchange of pump and proximal segment | 4 | 40 | 2.3 | 4 | 8709/8596 |
| 19 | TBI | Catheter dislocation of distal segment | 838 | Fluoroscopy: no signs | Exchange of entire catheter | 5 | 0 | 2.3 | 5 | 8711 |
| 20 | SP | Unknown catheter dysfunction | 1,174 | MRI: no signs | Exchange of distal segment | 1 | 100 | 22.5 | 5 | 8781 |
| 21 | Hypoxia | Pump site infection | 23 | No imaging | Exchange of entire system | 5 | 0 | 39.2 | 5 | 8781 |
| 22 | MS | Catheter disconnection of proximal segment | 33 | CT: no signs | Exchange of proximal segment | 5 | 5 | 7.5 | 5 | 8781 |
| 23 | Hypoxia | CSF leakage (early) | 16 | No imaging | Extracorporal compression | 5 | 0 | 6.3 | 4 | 8731SC |
| 24 | TSI | Catheter disconnection | 1,864 | XR & CT: no signs | Exchange of entire catheter | 4 | 75 | 11.9 | 0 | 8711 |
| 25 | SP | Catheter break | 1,570 | CT+CA: no intradural contrast | Exchange of entire system | 2 | 85 | 14.6 | 4 | 8731SC |
| 26 | SP | CSF leakage (early) | 44 | MRI: leakage | Patching | 1 | 100 | 2.7 | 3 | 8781 |
| 27 | TBI | Catheter dislocation of distal segment | 2,058 | CT+CA: no signs | Exchange of entire system | 5 | 0 | 7.2 | 5 | 8731SC |
| 28 | SP | Catheter disconnection | 2,179 |
XR: disconnection | Exchange of entire system | 2 | 100 | 11.0 | 3 | 8731SC |
| 29 | TSI | CSF leakage (late) | 2,200 |
XR + CA: no signs | Exchange of entire system | 4 | 60 | 5.2 | 0 | 8731SC |
| 30 | TSI | Pump hypermobility | 1,683 | No imaging | Pump exchange | 4 | 67.5 | 10 | 0 | 8731SC |
| 31 | TSI | Catheter dislocation of distal segment | 12 | No imaging | Exchange of entire catheter | 4 | 80 | 11.2 | 0 | 8731SC |
| 32 | SAH | CSF leakage (late) | 2,270 | XR: no signs | Exchange of pump and proximal segment | 4 | 10 | 8.3 | 3 | 8731SC |
CA, contrasting agent; CP, cerebral palsy; CSF, cerebrospinal fluid; CT, computer tomography; MRI, magnet resonance imaging; MS, multiple sclerosis; No, number of complication; SAH, subarachnoidal hemorrhage; SP, spastic paralysis; TBI, traumatic brain injury; TSI, traumatic spinal injury; XR, conventional x‐ray.
2 complications in 3 patients each.
Time to event was calculated from the last surgical procedure.
First implantation or revision surgery.
Scheduled device swap due to end of battery life to the date of a consecutive intervention (surgical or non‐surgical interventions).
Figure 1Temporal distribution of occurring complications within the first 7 years after the last intrathecal baclofen (ITB)‐related surgical procedure (first implantation surgery, replacement surgery due to end of battery life or due to a revision after a complication)
Figure 2Temporal distribution of occurring complications within the first 12 months after the last surgical procedure (first implantation surgery, replacement surgery due to end of battery life or due to a revision after a complication)