Comparison between an Ascenda and a silicone catheter in intrathecal baclofen therapy in pediatric patients: analysis of complications.

Francesco Motta1, Clara Eleonora Antonello1.   

Abstract

OBJECTIVE In this single-center study the authors investigated the complications occurring before and after the introduction of the new Ascenda intrathecal catheter (Medtronic Inc.) in pediatric patients treated with intrathecal baclofen therapy (ITB) for spasticity and/or dystonia. METHODS This was a retrospective review of 508 children who had received ITB, 416 with silicone catheters in the 13 years between September 1998 and September 2011 and 92 with Ascenda catheters in the 3 years between September 2011 and August 2014. The authors evaluated major complications such as infections, CSF leaks treated, and problems related to the catheter or pump, and they compared the 2 groups of patients who had received either a silicone catheter or an Ascenda catheter implant. RESULTS One hundred twenty patients in the silicone group (29%) and 1 patient in the Ascenda group (1.1%; p < 0.001) had a major complication. In the silicone group 23 patients (5.5%) were affected by CSF leakage and 75 patients (18%) experienced 82 catheter-related events, such as occlusion, dislodgment, disconnection, or breakage, which required catheter replacement. In the Ascenda group, only 1 patient (1.1%) was affected by CSF leakage. CONCLUSIONS To the authors' knowledge, this study is the first in the literature to compare the performance of the new Ascenda catheter, introduced in 2011, with the traditional silicone catheter for intrathecal drug infusion. In their analysis, the authors found that the Ascenda catheter can reduce major complications related to the catheter after ITB pump implantation. Further investigation is necessary to expand on and confirm their results.

Entities:  

Keywords:  Ascenda catheter; CP = cerebral palsy; CSF = cerebrospinal fluid; GMFCS = Gross Motor Function Classification System; ITB = intrathecal baclofen therapy; PET = polyethylene terephthalate; complications; functional neurosurgery; intrathecal baclofen; pediatric patient

Mesh:

Substances:

Year:  2016        PMID: 27341610     DOI: 10.3171/2016.4.PEDS15646

Source DB:  PubMed          Journal:  J Neurosurg Pediatr        ISSN: 1933-0707            Impact factor:   2.375


  9 in total

Review 1.  Complications associated with intrathecal drug delivery systems.

Authors:  E M Delhaas; F J P M Huygen
Journal:  BJA Educ       Date:  2019-12-12

Review 2.  Safety of intrathecal route: focus to methylprednisolone acetate (Depo-Medrol) use.

Authors:  Joël Schlatter; David Nguyen; Michèle Zamy; Sofiane Kabiche; Jean-Eudes Fontan; Salvatore Cisternino
Journal:  Eur Spine J       Date:  2017-11-18       Impact factor: 3.134

3.  Plain radiography in patients treated with intrathecal drug delivery using an implantable pump device.

Authors:  Elmar M Delhaas; Biswadjiet S Harhangi; Sander P G Frankema; Frank J P M Huygen; Aad van der Lugt
Journal:  Insights Imaging       Date:  2017-08-24

4.  A case of neglected silicone tube in lacrimal duct for 20 years.

Authors:  Masashi Mimura; Hidehiro Oku; Mari Ueki; Bunpei Sato; Tsunehiko Ikeda
Journal:  Am J Ophthalmol Case Rep       Date:  2018-05-22

5.  Pre- and Post-Interventional Changes in Physiological Profiles in a Patient Presenting With Opioid Withdrawal After Intrathecal Drug Delivery System Failure Related to Assumed Catheter Microfracture.

Authors:  Reza Ehsanian; Eugene Koshkin; Aleyah E Goins; Marena A Montera; Sascha Alles
Journal:  Cureus       Date:  2021-05-04

6.  Eleven years' experience with Intrathecal Baclofen - Complications, risk factors.

Authors:  Elke Pucks-Faes; Gabriel Hitzenberger; Heinrich Matzak; Elena Fava; Giulio Verrienti; Ilse Laimer; Josef Fritz; Leopold Saltuari
Journal:  Brain Behav       Date:  2018-03-30       Impact factor: 2.708

7.  Occlusion of the Ascenda Catheter in a Patient with Pump Twiddler's Sydrome: A Case Report.

Authors:  Jianning Shao; Leonardo Frizon; Andre G Machado; Keith McKee; Francois Bethoux; Jennifer Hartman; Sean J Nagel
Journal:  Anesth Pain Med       Date:  2018-04-28

8.  Catheter Access Port (Computed Tomography) Myelography in Intrathecal Drug Delivery Troubleshooting: A Case Series of 70 Procedures.

Authors:  Elmar M Delhaas; Biswadjiet S Harhangi; Sander P G Frankema; Frank J P M Huygen; Aad van der Lugt
Journal:  Neuromodulation       Date:  2020-04-08

9.  European expert consensus on improving patient selection for the management of disabling spasticity with intrathecal baclofen and/or botulinum toxin type A.

Authors:  Bo Biering-Soerensen; Valerie Stevenson; Djamel Bensmail; Klemen Grabljevec; Mercedes Martínez Moreno; Elke Pucks-Faes; Joerg Wissel; Mauro Zampolini
Journal:  J Rehabil Med       Date:  2022-01-03       Impact factor: 2.912

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.