4D modeling in a gimbaled linear accelerator by using gold anchor markers.

Hideharu Miura1,2, Shuichi Ozawa1,2, Takaaki Matsuura1, Atsushi Kawakubo1, Fumika Hosono1, Kiyoshi Yamada1, Yasushi Nagata1,2.   

Abstract

PURPOSE: The purpose of this study was to verify whether the dynamic tumor tracking (DTT) feature of a Vero4DRT system performs with 10-mm-long and 0.28 mm diameter gold anchor markers.
METHODS: Gold anchor markers with a length of 10 mm and a diameter of 0.28 mm were used. Gold anchor markers were injected with short and long types into bolus material. These markers were sandwiched by a Tough Water (TW) phantom in the bolus material. For the investigation of 4-dimensional (4D) modeling feasibility under various phantom thicknesses, the TW phantom was added at 2 cm intervals (in upper and lower each by 1 cm). A programmable respiratory motion table was used to simulate breathing-induced organ motion, with an amplitude of 30 mm and a breathing cycle of 3 s. X-ray imaging parameters of 80 kV and 125 kV (320 mA and 5 ms) were used. The least detection error of the fiducial marker was defined as the 4D-modeling limitation.
RESULTS: The 4D modeling process was attempted using short and long marker types and its limitation with the short and long types was with phantom thicknesses of 6 and 10 cm at 80 kV and 125 kV, respectively. However, the loss in detectability of the gold anchor because of 4D-modeling errors was found to be approximately 6% (2/31) with a phantom thickness of 2 cm under 125 kV. 4D-modeling could be performed except under the described conditions.
CONCLUSIONS: This work showed that a 10-mm-long gold anchor marker in short and long types can be used with DTT for short water equivalent path length site, such as lung cancer patients, in the Vero4DRT system.

Entities:  

Keywords:  4D-modeling; Gold anchor marker; IGRT; Image quality; Vero4DRT

Year:  2018        PMID: 29760592      PMCID: PMC5948321          DOI: 10.1016/j.rpor.2018.02.008

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  14 in total

1.  Dose measurements near a non-radioactive gold seed using radiographic film.

Authors:  James C L Chow; Grigor N Grigorov
Journal:  Phys Med Biol       Date:  2005-08-31       Impact factor: 3.609

2.  Optimization of the x-ray monitoring angle for creating a correlation model between internal and external respiratory signals.

Authors:  Mami Akimoto; Mitsuhiro Nakamura; Nobutaka Mukumoto; Masahiro Yamada; Nami Ueki; Yukinori Matsuo; Akira Sawada; Takashi Mizowaki; Masaki Kokubo; Masahiro Hiraoka
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

3.  Accuracy verification of infrared marker-based dynamic tumor-tracking irradiation using the gimbaled x-ray head of the Vero4DRT (MHI-TM2000).

Authors:  Nobutaka Mukumoto; Mitsuhiro Nakamura; Akira Sawada; Yasunobu Suzuki; Kunio Takahashi; Yuki Miyabe; Shuji Kaneko; Takashi Mizowaki; Masaki Kokubo; Masahiro Hiraoka
Journal:  Med Phys       Date:  2013-04       Impact factor: 4.071

4.  A multi-centre analysis of treatment procedures and error components in dynamic tumour tracking radiotherapy.

Authors:  Yukinori Matsuo; Dirk Verellen; Kenneth Poels; Nobutaka Mukumoto; Tom Depuydt; Mami Akimoto; Mitsuhiro Nakamura; Nami Ueki; Benedikt Engels; Christine Collen; Masaki Kokubo; Masahiro Hiraoka; Mark de Ridder
Journal:  Radiother Oncol       Date:  2015-05-18       Impact factor: 6.280

5.  A finite state model for respiratory motion analysis in image guided radiation therapy.

Authors:  Huanmei Wu; Gregory C Sharp; Betty Salzberg; David Kaeli; Hiroki Shirato; Steve B Jiang
Journal:  Phys Med Biol       Date:  2004-12-07       Impact factor: 3.609

6.  Qualitative evaluation of fiducial markers for radiotherapy imaging.

Authors:  Maria F Chan; Gil'ad N Cohen; Joseph O Deasy
Journal:  Technol Cancer Res Treat       Date:  2014-09-16

7.  Treating patients with real-time tumor tracking using the Vero gimbaled linac system: implementation and first review.

Authors:  Tom Depuydt; Kenneth Poels; Dirk Verellen; Benedikt Engels; Christine Collen; Manuela Buleteanu; Robbe Van den Begin; Marlies Boussaer; Michael Duchateau; Thierry Gevaert; Guy Storme; Mark De Ridder
Journal:  Radiother Oncol       Date:  2014-07-18       Impact factor: 6.280

8.  The management of respiratory motion in radiation oncology report of AAPM Task Group 76.

Authors:  Paul J Keall; Gig S Mageras; James M Balter; Richard S Emery; Kenneth M Forster; Steve B Jiang; Jeffrey M Kapatoes; Daniel A Low; Martin J Murphy; Brad R Murray; Chester R Ramsey; Marcel B Van Herk; S Sastry Vedam; John W Wong; Ellen Yorke
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

9.  Development of a video image-based QA system for the positional accuracy of dynamic tumor tracking irradiation in the Vero4DRT system.

Authors:  Kazuyu Ebe; Satoru Sugimoto; Satoru Utsunomiya; Hiroshi Kagamu; Hidefumi Aoyama; Laurence Court; Katsuichi Tokuyama; Ryuta Baba; Yoshisada Ogihara; Kosuke Ichikawa; Joji Toyama
Journal:  Med Phys       Date:  2015-08       Impact factor: 4.071

10.  Evaluation of different fiducial markers for image-guided radiotherapy and particle therapy.

Authors:  Daniel Habermehl; Katrin Henkner; Swantje Ecker; Oliver Jäkel; Jürgen Debus; Stephanie E Combs
Journal:  J Radiat Res       Date:  2013-07       Impact factor: 2.724

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.