Effect of bearing friction torques on the primary stability of press-fit acetabular cups: A novel in vitro method.

Alexander Jahnke1, Stefan Schroeder1, Carlos A Fonseca Ulloa1, Gafar A Ahmed1,2, Bernd A Ishaque1,2, Markus Rickert1,2.   

Abstract

Aseptic loosening is the main reason for revision of total hip arthroplasty, and relative micromotions between cementless acetabular cups and bone play an important role regarding their comparatively high loosening rate. Therefore, the aim of the present study was to analyze the influence of resulting frictional torques on the primary stability of press-fit acetabular cups subjected to two different bearing partners. A cementless press-fit cup was implanted in bone-like foam. Primary stability of the cup was analyzed by determining spatial total, translational, and rotational interface micromotions by means of an eddy current sensor measuring system. Torque transmission into the cup was realized by three synchronous servomotors considering resultant friction torques based on constant friction for ceramic-on-ceramic (CoC: μ = 0.044; max. resultant torque: 1.5 Nm) and for ceramic-on-polyethylene (CoP: μ = 0.063; max. resultant torque: 1.9 Nm) bearing partners. Rotational micromotion of CoC was 8.99 ± 0.85 µm and of CoP 13.39 ± 1.43 µm. Translational micromotion of CoC was 29.93 ± 1.44 μm and of CoP 39.91 ± 2.25 μm. Maximum total relative micromotions were 37.10 ± 1.07 μm for CoC and 51.64 ± 2.18 μm for CoP. Micromotions resulting from CoC were statistically lower than those resulting from CoP (p < 0.05). The described 3D-measuring set-up offers a novel in vitro method of measuring primary stability of acetabular cups. We can therefore conclude, that primary stability of acetabular cup systems can be observed using either the lower friction curve (CoC) or the higher friction curve (CoP). In future studies different cup designs or cup fixation mechanisms may be tested and compared in vitro and assessed prior to implantation.
© 2018 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 36:2745-2753, 2018. © 2018 Orthopaedic Research Society. Published by Wiley Periodicals, Inc.

Entities:  

Keywords:  acetabular component; cementless; friction torques; hip arthroplasty; primary stability

Mesh:

Year:  2018        PMID: 29744927     DOI: 10.1002/jor.24041

Source DB:  PubMed          Journal:  J Orthop Res        ISSN: 0736-0266            Impact factor:   3.494


  3 in total

1.  How screw connections influence the primary stability of acetabular cups under consideration of different bone models.

Authors:  Gafar Adam Ahmed; Linda Auge; Jessica Loke; Carlos Alfonso Fonseca Ulloa; Christian Fölsch; Markus Rickert; Bernd Alexander Ishaque; Alexander Jahnke
Journal:  J Orthop       Date:  2020-06-06

2.  Clinical evaluation of a novel press-fit acetabular cup using "Ein-Bild-Roentgen-Analysis" (EBRA): A positive short-term prognosis.

Authors:  B A Ishaque; J Wieczorek; C A Fonseca Ulloa; J B Seeger; G A Ahmed; M Rickert; A Jahnke
Journal:  J Orthop       Date:  2020-03-30

3.  The Use of Porous Titanium Coating and the Largest Possible Head Do Not Affect Early Cup Fixation: A 2-Year Report from a Randomized Controlled Trial.

Authors:  Georgios Tsikandylakis; Kristian R L Mortensen; Kirill Gromov; Anders Troelsen; Henrik Malchau; Maziar Mohaddes
Journal:  JB JS Open Access       Date:  2020-12-15
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.