M T Bahreyni Toossi1, B Khajetash1, M Ghorbani1.
Abstract
BACKGROUND: One of the main causes of induction of secondary cancer in radiation therapy is neutron contamination received by patients during treatment. Objective: In the present study the impact of wedge and block on neutron contamination production is investigated. The evaluations are conducted for a 15 MV Siemens Primus linear accelerator.Entities:
Keywords: Cerrobend block ; Monte Carlo simulation; Siemens Primus linac ; Wedge ; Neutron contamination
Year: 2018 PMID: 29732336 PMCID: PMC5928309
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
Compositions (weight fraction) of cerrobend obtained by atomic absorption spectrophotometry method
| Element | Percentage (%) |
|---|---|
| Tin (Sn) | 21.31 |
| Lead (Pb) | 26.34 |
| Bismuth (Bi) | 52.35 |
Relative neutron dose values (mSv/mSv) for 30˚, 45˚ and 60˚ wedges and cerrobend block at 0.5, 2, 3 and 4 cm depths in 10 × 10 cm2 field size without taking wedge/tray factor into account
| Depth (cm) | 30˚ Wedge | 45˚ Wedge | 60˚ Wedge | Block |
|---|---|---|---|---|
| 0.5 | 0.70 | 0.57 | 0.57 | 0.50 |
| 2.0 | 0.65 | 0.55 | 0.54 | 0.48 |
| 3.0 | 0.67 | 0.54 | 0.55 | 0.45 |
| 4.0 | 0.62 | 0.49 | 0.50 | 0.49 |
Wedge factors related to 30˚, 45˚, 60˚ wedges and block tray for 10 × 10 cm2 field size of 15 MV photon beam of siemens primus linac
| Wedge/Tray Factor | |
|---|---|
|
| 0.60 |
|
| 0.40 |
|
| 0.43 |
|
| 0.96 |
Relative neutron dose values (mSv/mSv) for 30˚, 45˚ and 60˚ wedges and cerrobend block at depths of 0.5, 2, 3 and 4 cm in 10 × 10 cm2 field size with taking wedge/tray factor into account
| Depth (cm) | 30˚ Wedge | 45˚ Wedge | 60˚ Wedge | Block |
|---|---|---|---|---|
| 0.5 | 1.16 | 1.42 | 1.33 | 0.51 |
| 2.0 | 1.09 | 1.37 | 1.25 | 0.49 |
| 3.0 | 1.12 | 1.35 | 1.29 | 0.47 |
| 4.0 | 1.04 | 1.22 | 1.17 | 0.51 |
Figure1Relative neutron dose versus depth in water phantom without taking wedge/tray factor into account
Figure2Relative neutron dose versus depth in water phantom without taking wedge/tray factor into account
Values of relative neutron dose by F4 and F6 tallies at different depths for 30˚, 45˚ and 60˚ wedges and cerrobend block in water phantom with and without taking wedge/tray factor into account
| Without taking wedge/tray factor into account | ||||||||
|---|---|---|---|---|---|---|---|---|
| F4 Tally (mSv/mSv) | F6 Tally (cGy/cGy) | |||||||
| Depth (cm) | 30˚ wedge | 45˚ wedge | 60˚ wedge | Block | 30˚ wedge | 45˚ wedge | 60˚ wedge | Block |
| 0.5 | 0.70 | 0.57 | 0.57 | 0.50 | 0.70 | 0.56 | 0.57 | 0.51 |
| 2.0 | 0.65 | 0.55 | 0.54 | 0.48 | 0.63 | 0.53 | 0.53 | 0.52 |
| 3.0 | 0.67 | 0.54 | 0.55 | 0.45 | 0.65 | 0.53 | 0.55 | 0.49 |
| 4.0 | 0.62 | 0.49 | 0.50 | 0.49 | 0.59 | 0.47 | 0.49 | 0.51 |
|
| ||||||||
| 0.5 | 1.16 | 1.42 | 1.33 | 0.51 | 1.16 | 1.41 | 1.31 | 0.53 |
| 2.0 | 1.09 | 1.37 | 1.25 | 0.49 | 1.04 | 1.31 | 1.22 | 0.54 |
| 3.0 | 1.12 | 1.35 | 1.29 | 0.47 | 1.08 | 1.31 | 1.29 | 0.51 |
| 4.0 | 1.04 | 1.22 | 1.17 | 0.51 | 0.98 | 1.16 | 1.13 | 0.53 |
Monte Carlo type A statistical uncertainty (%) for the F4 and F6 tallies in 0.5, 2.0, 3.0 and 4.0 cm depths in water phantom
| Depth (cm) | Open field | 30˚ wedge | 45˚ wedge | 60˚ wedge | Block | |||||
|---|---|---|---|---|---|---|---|---|---|---|
| F4 | F6 | F4 | F6 | F4 | F6 | F4 | F6 | F4 | F6 | |
|
| 2.08 | 2.11 | 2.52 | 2.57 | 2.78 | 2.82 | 2.76 | 2.80 | 1.90 | 1.89 |
|
| 5.23 | 5.57 | 6.07 | 6.30 | 6.62 | 6.97 | 6.78 | 7.24 | 5.00 | 4.85 |
|
| 6.37 | 6.88 | 7.36 | 7.86 | 8.09 | 8.74 | 8.38 | 9.13 | 6.87 | 6.51 |
|
| 7.58 | 8.02 | 7.90 | 8.37 | 8.80 | 9.32 | 9.02 | 9.60 | 8.72 | 8.49 |