| Literature DB >> 29732223 |
Norihide Murayama1, Kikuno Murayama1.
Abstract
BACKGROUND: Ultrasonic, jet, and mesh nebulizers have all been used in the treatment for asthma. Mesh nebulizers reportedly offer the best inhalation efficiency.Entities:
Mesh:
Substances:
Year: 2018 PMID: 29732223 PMCID: PMC5872591 DOI: 10.1155/2018/1648652
Source DB: PubMed Journal: Pulm Med ISSN: 2090-1844
Clinical examinations before and after treatment for children ≥ 2 years old.
| >2 years | Jet type | Mesh type | ||
|---|---|---|---|---|
| Before | After | Before | After | |
|
| 28 | 23 | ||
| Age, years | 3.55 (1.17) | 3.44 (0.90) | ||
| Inhalation time, s | 377 (58.7) | 248 (31.4) | ||
| Heart rate, beats/min | 129 (22.4) | 136 (21.3) | 135 (16.5) | 145 (17.6) |
| SpO2, % | 97.2 (1.57) | 97.4 (1.59) | 97.2 (1.53) | 97.0 (1.59) |
| Symptom score | 1.14 (0.71) | 0.25 (0.52) | 1.04 (0.61) | 0.29 (0.45) |
Mean (SD), p < 0.001, p < 0.0001, paired t-test, and independent t-test.
Clinical examinations before and after treatment for children < 2 years old.
| <2 years | Jet type | Mesh type | ||
|---|---|---|---|---|
| Before | After | Before | After | |
|
| 17 | 20 | ||
| Age, years | 1.30 (0.35) | 1.38 (0.40) | ||
| Inhalation time, s | 364 (53.9) | 236 (22.5) | ||
| Heart rate, beats/min | 141 (21.4) | 145 (18.3) | 141 (22.6) | 147 (18.4) |
| SaO2, % | 97.7 (1.44) | 97.9 (1.73) | 97.6 (1.52) | 98.0 (1.73) |
| Symptom score | 1.27 (0.57) | 0.38 (0.60) | 1.47 (0.81) | 0.38 (0.66) |
Mean (SD), p < 0.0001, paired t-test, and independent t-test.
Figure 1Changes in children ≥ 2 years old. (a) Changes in heart rate. (b) Changes in SpO2. (c) Changes in symptom score.