Literature DB >> 29730456

Outcome and safety analysis of 3D-printed patient-specific pedicle screw jigs for complex spinal deformities: a comparative study.

Bhavuk Garg1, Manish Gupta2, Menaka Singh2, Dinesh Kalyanasundaram3.   

Abstract

BACKGROUND CONTEXT: Spinal deformities are very challenging to treat and have a great risk of neurologic complications because of hardware placement during corrective surgery. Various techniques have been introduced to ensure safe and accurate placement of pedicle screws. Patient-specific screw guides with predrawn and prevalidated trajectory seem to be an attractive option.
PURPOSE: We have focused on developing three-dimensional (3D) printing technique for complex spinal deformities in India. This study also aimed to compare the placement of pedicle screw with 3D printing and freehand technique. STUDY DESIGN/SETTINGS: This is a retrospective comparative clinical study in an academic institutional setting. PATIENT SAMPLE: A total of 20 patients were enrolled during the study: 10 were operated on with the help of 3D printing (Group 1) and 10 were operated on with freehand technique (Group 2). Group 1 included six patients with congenital scoliosis, three patients with adolescent idiopathic scoliosis (AIS), and one patient with post-tubercular kyphosis, and Group 2 included five patients with congenital scoliosis, four patients with AIS, and one patient with post-tubercular kyphosis. OUTCOME MEASURES: Primary outcomes were measured in terms of screw violation, and secondary outcomes were measured in terms of surgical time, blood loss, radiation exposure (number of shoots required), and complications.
MATERIALS AND METHODS: MIMICS Base v18.0 software was used for 3D reconstruction from computed tomography scan images of all the patients. 3-Matic software was used to create a drill guide. A 3D printer from Stratasys Mojo with ABS P430 model material cartilage (a thermoplastic material) was used for the printing of the vertebra model and jigs. A two-sample test of proportion was used to compare correctly and wrongly placed pedicle screws with 3D printing and freehand technique. t Test with equal variance was used for operating surgical time and blood loss.
RESULTS: No superior or inferior screw violation was observed in any of our patients in either group. We found a significant difference (p=.03) between the two groups regarding perfect screw placement in favor of 3D printing. There were 13 Grade 2 medial perforations in the freehand group and 3 in the 3D printing group. There was no Grade 3 medial perforation in either group. Six Grade 2 lateral perforations in the freehand group and seven in the 3D printing group were observed. Three Grade 3 lateral perforations in the freehand group and two in 3D printing group were observed. Analysis showed a statistically significant (p=.005) medial violation in the freehand group. Surgical time was significantly less (p=.03) in the 3D printing group compared with the freehand group. Mean blood loss was higher in the freehand group but was not statistically significant (p=.3) in the 3D printing group. Fluoroscopic shots required were less in number in the 3D printing group compared with the freehand group. There was no neurologic deficit in any of the patients in the two groups.
CONCLUSIONS: In our study, focusing on spinal deformities with statistically significant higher rates of accurate screw positioning and higher numbers of inserted screws with 3D printing was possible because of enhanced safety, particularly at apical levels. As such, spinal deformities are difficult to treat worldwide. In India, these deformities are often neglected and present at a very late and a much more deformed state when their treatment becomes even more challenging. Developing these patient-specific drill templates will enable an average spine surgeon to treat these patients with much ease and safety.
Copyright © 2018 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  3D printing; Pedicle screw; Scoliosis; Screw guide; Spinal deformity; Spine surgery

Year:  2018        PMID: 29730456     DOI: 10.1016/j.spinee.2018.05.001

Source DB:  PubMed          Journal:  Spine J        ISSN: 1529-9430            Impact factor:   4.166


  24 in total

1.  CORR Insights®: Does Three-dimensional Printing Plus Pedicle Guider Technology in Severe Congenital Scoliosis Facilitate Accurate and Efficient Pedicle Screw Placement?

Authors:  George H Thompson
Journal:  Clin Orthop Relat Res       Date:  2019-08       Impact factor: 4.176

Review 2.  Current status of 3D printing in spine surgery.

Authors:  Bhavuk Garg; Nishank Mehta
Journal:  J Clin Orthop Trauma       Date:  2018-08-07

Review 3.  3D printing in spine surgery.

Authors:  Evan D Sheha; Sapan D Gandhi; Matthew W Colman
Journal:  Ann Transl Med       Date:  2019-09

Review 4.  [3D printing in spinal surgery-Update].

Authors:  S Roth; S Sehmisch; S Decker
Journal:  Unfallchirurg       Date:  2022-02-17       Impact factor: 1.000

Review 5.  [Update on 3D printing in the surgery of musculoskeletal tumors].

Authors:  Mohamed Omar; Martin Schulze; Nico Bruns; Daniel Kotrych; Georg Gosheger; Max Ettinger
Journal:  Unfallchirurg       Date:  2022-03-21       Impact factor: 1.000

Review 6.  3D printing- creating a blueprint for the future of orthopedics: Current concept review and the road ahead!

Authors:  Vaibhav Bagaria; Rakesh Bhansali; Prashant Pawar
Journal:  J Clin Orthop Trauma       Date:  2018-07-23

7.  Occipitocervical instrumented fixation utilising patient-specific C2 3D-printed spinal screw trajectory guides in complex paediatric skeletal dysplasia.

Authors:  Vejay N Vakharia; Luke Smith; Zubair Tahir; Rachel Sparks; Sebastien Ourselin; Stewart Tucker; Dominic Thompson
Journal:  Childs Nerv Syst       Date:  2021-06-19       Impact factor: 1.475

Review 8.  3D printing and its applications in orthopaedic trauma: A technological marvel.

Authors:  Hitesh Lal; Mohit Kumar Patralekh
Journal:  J Clin Orthop Trauma       Date:  2018-08-03

Review 9.  Publication trends and knowledge mapping in 3D printing in orthopaedics.

Authors:  Raju Vaishya; Mohit Kumar Patralekh; Abhishek Vaish; Amit Kumar Agarwal; Vipul Vijay
Journal:  J Clin Orthop Trauma       Date:  2018-07-30

10.  Three-Dimensional Printing for Preoperative Planning and Pedicle Screw Placement in Adult Spinal Deformity: A Systematic Review.

Authors:  Cesar D Lopez; Venkat Boddapati; Nathan J Lee; Marc D Dyrszka; Zeeshan M Sardar; Ronald A Lehman; Lawrence G Lenke
Journal:  Global Spine J       Date:  2020-08-07
View more