Precontoured Clavicular Locking Plate with Broad Lateral End: A Newly Designed Plate for Lateral Third Clavicle Fractures.

K C Kapil-Mani1, P Acharya1, S Arun1.   

Abstract

Introduction: Various treatment modalities are available but no consensus has been reached for optimal treatment of lateral third clavicle fractures. Precontoured locking plates with broad lateral end for multiple screws fixation is a newly designed plate for lateral third clavicle fractures. The objective of our study was to analyse the functional outcomes as well as complications of this technique in a significant number of cases with long follow-up duration. Materials and
Methods: Forty-six patients with distal third clavicle fractures were treated by precontoured clavicular locking plate with broad lateral end. Functional outcomes were assessed on the basis of Constant-Murley Shoulder Outcome Score and University of California, Los Angeles (UCLA) Shoulder Rating Score, active shoulder range of motion, time for fracture union and coraco-clavicular distance.
Results: The mean Constant-Murley score was 92.56±4.47 (range: 79-98) for injured side and 96.22±2.23 (range: 90-100) for normal side with p-Value 0.56. Mean coraco-clavicular distance at final follow-up was 10.52±1.13 mm (range 9.7 to 11.7 mm) in injured side and 10.25±0.98 mm (range 9.6 to 11.2 mm) in normal side. Mean UCLA Shoulder Rating Score was 32.55±2.12 (range: 27-34) for injured side and 33.46±1.88 (range: 31- 35) on normal side with p value 0.58. No major complications that necessitated revision of surgery occurred in our study.
Conclusion: This newly designed plate seemed extremely useful in successful union of lateral third clavicle fractures, with reduced rate of complications like fixation failures, iatrogenic rotator cuff injury, AC joint osteoarthritis and sub-acromial bursitis, with good functional outcomes.

Entities:  

Keywords:  broad lateral end; complications; functional outcomes; lateral third clavicle fractures; precontoured locking plates

Year:  2018        PMID: 29725507      PMCID: PMC5920253          DOI: 10.5704/MOJ.1803.003

Source DB:  PubMed          Journal:  Malays Orthop J        ISSN: 1985-2533


  17 in total

1.  The treatment of distal clavicle fractures with coracoclavicular ligament disruption: a report of 10 cases.

Authors:  Murat Bezer; Nuri Aydin; Osman Guven
Journal:  J Orthop Trauma       Date:  2005-09       Impact factor: 2.512

2.  A method for internal fixation of unstable distal clavicle fractures: early observations using a new technique.

Authors:  Michael Kalamaras; Ken Cutbush; Mark Robinson
Journal:  J Shoulder Elbow Surg       Date:  2007-11-26       Impact factor: 3.019

3.  Stable fixation of distal clavicle fracture with comminuted superior cortex using oblique T-plate and cerclage wiring.

Authors:  Je Hyun Yoo; Jun Dong Chang; Young Jin Seo; Jae Hyuk Shin
Journal:  Injury       Date:  2008-12-18       Impact factor: 2.586

4.  Surgical treatment of distal clavicle fractures using the clavicular hook plate.

Authors:  Masafumi Kashii; Hiroaki Inui; Kouji Yamamoto
Journal:  Clin Orthop Relat Res       Date:  2006-06       Impact factor: 4.176

5.  Precontoured locking plate fixation for displaced lateral clavicle fractures.

Authors:  Sang Ki Lee; Jae Won Lee; Dae Geon Song; Won Sik Choy
Journal:  Orthopedics       Date:  2013-06       Impact factor: 1.390

6.  Hook-plate fixation of unstable lateral clavicle fractures: a report on 63 patients.

Authors:  Tapio Flinkkilä; Jukka Ristiniemi; Martti Lakovaara; Pekka Hyvönen; Juhana Leppilahti
Journal:  Acta Orthop       Date:  2006-08       Impact factor: 3.717

Review 7.  Fracture of distal end clavicle: A review.

Authors:  Balaji Sambandam; Rajat Gupta; Santosh Kumar; Lalit Maini
Journal:  J Clin Orthop Trauma       Date:  2014-06-19

8.  Simple, minimally invasive surgical technique for treatment of type 2 fractures of the distal clavicle.

Authors:  Ofer Levy
Journal:  J Shoulder Elbow Surg       Date:  2003 Jan-Feb       Impact factor: 3.019

9.  Comparison of the efficacy of hook plate versus tension band wire in the treatment of unstable fractures of the distal clavicle.

Authors:  Yih-Shiunn Lee; Ming-Jye Lau; Ya-Chun Tseng; Wen-Chiao Chen; Hsin-Yi Kao; Jyh-Ding Wei
Journal:  Int Orthop       Date:  2008-12-03       Impact factor: 3.075

10.  Non-union of the clavicle. Associated complications and surgical management.

Authors:  J B Jupiter; R D Leffert
Journal:  J Bone Joint Surg Am       Date:  1987-06       Impact factor: 5.284